贝克曼库尔特离心机S7 系统故障的解决方案:目前,实验室内一台贝克曼库尔特离心机开机后不能启动,屏幕显示的错误代码为:“S7 System Error,Call for Service Please.”,考虑到这台离心机很少使用,有可能是主板上ROM 内存在的信息丢失或错误.上网查找一下解决方案,果然有重置ROM参数的方案如下:,改用:或如果您也遇到类似问题、祝您马到成功、工作顺利。

原文参考如下:Q:At the start of a run there's a S7 system error that appears。

A:The S7 Error is a system error where there may be some memory loss due to a power disruption in the building。

Try pressing the Enter key then 911 then CE to clear the error and try another run。

If problems persist after this then service will be required. I hope this helps。

Michael Girard Product Support Services Beckman Coulter Inc。

1—800—369-0333 x353419(Thank You to Michael from Beckman Coulter for your response)Best Wishes,LabWrench Admin。

Q: I'm sorry,can you help me too?We have J-25 old centrifuge but it doesn't have ENTER key. How can I clear S7 error in this case?A:To clear an S7 on a J25; press Time,Accel, CE. All your calibration factors were set to Default (100)and the elevation is set to 0 meters by the S7。

离心机常见故障及解决方法

离心机常见故障及解决方法

离心机常见故障及解决方法以离心机常见故障及解决方法为标题,下面是一些常见的离心机故障及其解决方法。

一、离心机启动困难或无法启动1. 检查电源是否正常,确保电压稳定。

2. 检查电机接线是否松动或烧坏,修复或更换电机。

3. 检查启动电容器是否损坏,如有损坏需更换新的电容器。

4. 检查启动电路是否故障,修复或更换故障部件。

二、离心机运转不正常或噪音过大1. 检查离心机底座是否稳固,如底座松动需及时固定。

2. 检查离心机转子是否平衡,如不平衡需进行平衡校准。

3. 检查离心机轴承是否磨损,如有磨损需更换新的轴承。

4. 检查离心机传动带是否松弛或磨损,如有问题需及时调整或更换传动带。

5. 检查离心机内部零件是否松动,如有松动需固定或更换相关零件。

三、离心机转速不稳定或无法达到要求1. 检查离心机电机是否故障,如有故障需修复或更换电机。

2. 检查离心机调速器是否正常,如有问题需调整或更换调速器。

3. 检查离心机传动系统是否顺畅,如有问题需修复或更换传动系统的零件。

4. 检查离心机转子是否平衡,如不平衡需进行平衡校准。

四、离心机温度异常高或过热1. 检查离心机冷却系统是否正常,如冷却水流量不足需调整或更换冷却水泵。

2. 检查离心机轴承润滑是否良好,如不良需进行润滑或更换润滑油。

3. 检查离心机电机是否过载,如过载需减少负荷或更换功率更大的电机。

五、离心机显示屏故障或故障信息不准确1. 检查离心机显示屏连接线是否松动,如松动需固定连接线。

2. 检查离心机显示屏是否存在故障,如有故障需更换新的显示屏。

3. 检查离心机传感器是否正常,如有问题需修复或更换传感器。

六、离心机安全保护装置报警或无法启动1. 检查离心机安全保护装置是否正常,如有故障需修复或更换相关部件。

2. 检查离心机安全保护装置是否被启动,如被启动需复位或解除保护状态。

总结:离心机常见故障多为电源、电机、传动系统、冷却系统等方面的问题,解决方法主要是修复或更换故障部件。

(精品文档)Backman贝克曼离心机故障详解

(精品文档)Backman贝克曼离心机故障详解

Backman贝克曼离心机故障详解离心机在高速旋转的过程中,由离心力所导致的运动使悬浮于液体中的固体物质形成沉淀,也就是悬浮体液中质量或体积较大的物体向转头半径最大的方向移动,而质量或体积较小的部分沉积在转头半径较近的地方。

由于转子的半径和样品的质量在运转的时候是不变的,只有转速可以通过控制发生变化,因此我们往往习惯用转速来描述一个离心机,如:高速离心机,超高速离心机。

1 离心机触摸键盘失控故障现象:开机后按任何键均无反应,仪器进入死机状态。

故障分析:造成触摸键盘失控一般有以下两个原因:(1)触摸键盘未接通电源。

触摸键盘粘键。

开机后按任何均无反应,可能是某个粘住键盘使计算机无法正确识别键盘状态而造成死机现象。

故障排除方法:经检查,触摸键盘的电源正常。

切断电源,拆下触摸键盘的螺钉。

将触摸键盘从夹层处轻轻掰开。

沿机内键盘夹层中非键盘按键部分用双面胶贴好后,将触摸键盘复原即可。

应注意双面胶不能垫得太高。

2 电机过热故障现象:电机过热平常有以下几种环境:离心机起动时间为≤60s , r如起动时间长,则大概是抛块磨损,重量过轻,摩擦片外貌有油污以及电机存在质量问题,因此必要变加载的抛块或查抄电机。

起动电流大,离心机的起动电流约为额定电流的2.5倍,负荷起动电流约为额定电流的3倍,要是电流高出该值,则大概是电机存在质量问题,电源线路是否精确,电机轴是否弯曲,离合器是否有卡阻征象,离合器抛块是否超重。

3 振动空运转振动大:离心机的振动紧张因为转鼓质量的不屈衡与转鼓支承间间隙太大,各团结部位的螺栓,螺帽松动造成。

要是利用时间长,腐化紧张,则应思量转鼓焊缝或别的部位开裂的大概性,偶然转鼓内积累物料太多,造成转鼓偏重,也会造成振动,别的如主轴弯曲、轴承破坏等也有大概造成振动。

负荷运转振动大:负荷运转振动除了上述缘故原由外,紧张是物料散布不均造成偏重而造成振动,议决调解加料要领、匀称布料即可处理,其次,大概因为转鼓利用时间较长,或遭受磕碰而造成变形,在利用进程中转鼓内形成的滤饼层的厚度就不行能均布,以是呆板就会出现振动。

BECKMAN离心机使用流程

BECKMAN离心机使用流程

BECKMAN离心机使用流程
1、向负责人领取试验所需型号的转头,工具及润滑油;
2、接通电源,开机,打开仓门,更换转头(在熟悉该仪器操作的人员指
导下进行拆卸和组装);
3、平衡样品、上机;
4、选择转头:ROTOR→▲或▼→ENTER:选择转头型号;
5、调整离心参数:
5.1、 PRM (RCF)→▲或▼→ENTRE:设定转速(离心力);
5.2、TIME→▲或▼→ENTRE:设定时间;
5.3、TEMP→▲或▼→ENTRE:设定温度;
5.4、 ACCEL和DECEL:一般设定为5即可;
6、离心结束后将转头小心拆卸,归还仪器负责人;
7、离心结束后请擦干离心仓内的水珠,注意勿碰到传感器,保持离心机
盖是打开状态。

友善提示:
1、本仪器为超速低温离心机,关盖后自行启动降温功能;如需快速降温,
可将转头放入4°C冰箱预冷;
2、关盖时请确定同时压紧两边的锁扣;开盖时轻压离心机盖;
3、离心机的轴心为精密仪器,价格昂贵,望大家小心使用;
4、要超速离心样品,请确定离心管的承受力,或确保管内液体量小于离
心管容量的1/2;
5、转头若不小心受到剧烈撞击(如跌落地面),请及时联系负责人,需拍
X光片,确定无裂纹方可继续使用。

BECKMAN离心机管理员:
2010年4月20日。

离心机常见故障处理

离心机常见故障处理

常见故障处理李国成19EX系列机组控制系统较为先进,具备全方位的多重保护功能于一身,能保证机组在安全的前提下可靠运行。

但当机组运行工况发生明显变化或当部分检测、控制元器件发生故障时便会产生一些控制系统方面的故障,常见故障的处理方法如下:(一)压缩机高扬程、涌浪保护离心式制冷压缩机其压缩比较低,当运行工况发生严重变化时易发生压缩机喘振现象。

开利19EX系列机组控制中心通过对运行数据的监测,可以有效防止喘振现象的发生,该系列机组除了对冷凝压力进行监测外还会测量冷冻水的进出水温差(当冷冻水流量一定时其温差代表着机组的负荷量,即制冷量),及冷凝压力与蒸发压力差(机组的扬程),机组运行在不同的负荷时应有相应的扬程(冷凝压力与蒸发压力差),实际扬程过高则会发生喘振。

当控制中心显示压缩机高扬程或涌浪保护时应检查以下内容:1.机组冷冻水流量过小或缺少制冷剂,使蒸发压力过低。

2.机组冷却水流量过小、制冷剂过多、冷却水进水温度过高或机组内有不凝性气体存在使冷凝压力过高。

3.冷冻水流量过大或水温传感器损坏使冷冻水进出口温差过小,控制中心允许的扬程过小而使保护程序误动作。

4.控制中心的SERVICE1菜单中△T1、△T2、△P1、△P2设定数据不合理使机组保护程序误动作。

当机组出现喘振现象(电流剧裂波动并伴有强裂气流声)时表明控制中心的保护功能未启作用,应检查以下方面:1.控制中心SERVICE1菜单中△T1、△T2、△P1、△P2设定数据是否合理,该数据在机组出厂调试时均已设定,但在运行、维护及更换PSIO板的过程中大多做过调整。

2.冷冻水流量过小,使冷冻水温差过大控制中心允许的扬程过高。

3.冷冻水温度传感器故障使冷冻水进出口温差过大。

(二)水温传感器故障19EX系列机组的温度传感器为负温度系数的热敏电阻,由于其设计的缺陷使得其接线端易因腐蚀而接触不良(接插处直接受冷冻水温过低而出现冷凝水凝结,其插针选材不合理较易锈断),一般均出现接触电阻过大、温度显示值偏低,用除锈剂清洗接头并紧固改善接触状况后可恢复正常,若插针断裂则必须更换。

加工中心常见报警及解决方法

加工中心常见报警及解决方法

旺磐加工中心的常见报警解决方法序号报警内容含义解决方法<一> plc报警问题1.1 LUB LOW (油量过少)1.11 检查润滑油泵的油位1.12 检查油位传感器是否正常1.13检查油位报警线路电源及输入电路是否正常(号码管为DC24V及LUB LOW)1.2COOLANT OVERLOAD (切削液马达过载)1.21 检查动力线是否有缺,1.22 检查电源电压是否为额定电压1.23 过载保护器的过载系数是否设定过小,正常为2.51.24 马达是否为反转或者有烧毁1.25 将上序问题排除后,将过载保护器上的复位按钮按下,再确定信号线是否有24V电源输入(号码管为COOLANT OVERLOAD)1.3 AXIS NOT HOME (3轴未归零)1.31 在原点复归模式下分别将三轴归零,归完成报警信号即完成零1.32 ATC NOT READY 刀库未准备好1.33 刀库记数信号未到位,检查COUNTER信号1.34 刀杯原位信号错误,检查TOOL CUP UP 信号1.35 刀臂持刀点位置不正确,检查121点信号1.4 THE CLAMP SIGNAL ERROR (夹刀信号错误)1.41 检查夹刀到位信号线是否有异常1.42 检查打刀缸夹刀开关是否正常1.43 检查I/F诊断中X4的信号是否为11.5 AIR PRESSURE LOW (空气压力低)1.51 检查空气压力是否5MP以上1.52 检查空气压力输入信号的线路是否有DC24VV电压1.6 ATC COUNTER SINGAL ERROR (刀库记数信号错误)1.61 检查是否为记数信号接再刀库的144点上。

1.62 检查DC24电源144点与0V点之间电压是否为24V,1.63确定I/F诊断中的X1E点信号是否正常!1.7 THE SP-MOTOR OVERLOAD (主轴马达过载)1.71 主轴马达过载,检查回升电阻AL1与AL2间是否为通路1.72 检查PLC输入信号是否有24V1.8 PLC计时器不能工作参数被锁住1.81将6449BIT0位参数设定为01.9 PLC计数器不能工作参数被锁住1.91将6449BIT1位参数设定为01.10 PLC报警无法显示为用户定义形式参数锁住1.101将6550BIT0位和BIT2位设定为11.11 主轴无法转动(无信号显示)1.111 如果是使用面板启动主轴正转时,则检查是否是在手轮方式,然后检查打刀缸上面的紧刀信号是否到位(即I/F诊断中的点X4是否为1),如果此两点信号全部到位,主轴还是无法转动,则检查报警界面是否有伺服报警。

超离管理

超离管理

Beckman 超高速离心机使用说明一、超速离心机由专人保管专人使用,非专管人员严禁擅自操作。

二、离心管要对称、同量、同液放置,使用水平转子时一定要检查离心管是否挂牢,配平需精确到小数点后三位数。

三、超速离心时,液体一定要加满离心管,因超离需抽真空,只有加满才能避免离心管变形。

四、转头在每次使用前要严格检查孔内是否有异物和污垢,以保持平衡。

五、转头放置要规范、要轻拿轻放,垂直放在转轴上,转子一定要水平,放好后轻转转头,有速度显示才可正式运转。

六、不允许转头使用超过最大速度。

七、在离心过程中如发现异常现象,如不正常噪音及振动,应立即停机检查,未找出原因前不得继续运转。

八、关、开离心机的门要轻,防止仪器移位,影响水平。

九、离心后将离心管、转头、腔体清洗干净,规范存放,并按要求填写仪器使用记录。

十、如有问题,请立即与管理人员联系。

超速离心机运行:(1)运行开始前按键“VACUUM”启动真空系统,真空度小于300方可开启转子。

(2)启动运行时先按“ENTER”键,然后按“START”键,约5秒钟后运行开始。

注:如果用户按“START”而未首先按“ENTER”键,运行不启动。

停止键STOP 任何时候均可停止运行。

(3)离心机停止运转后,先按“VACUUM”去真空后方可打开盖子。

特别强调:对称平衡:当离心转速达1-5万(转/分)时,如对称管相差1克,转头半径5厘米,则此时引起两边不平衡可达6-150(公斤),这对离心机的损伤是很大的,至少将缩短离心机的使用寿命。

如离心管盖子密封性差液体就不能加满(针对高速离心且使用角度头),以防外溢。

外溢后果:①失去平衡;②污染转头和离心腔;③影响感应器正常工作、损坏马达。

超速离心时,液体一定要加满离心管,因超离需抽真空,只有加满才能避免离心管变形。

使用角度头时别忘盖转头盖,如未盖,离心腔内会产生很大的涡流阻力和摩擦升温,这等于给离心机的电机和制冷机增加了额外负担,影响离心机的使用寿命。

离心机常见故障及解决方法

离心机常见故障及解决方法
转矩达到阶段2(90%)
停止离心机
检查2齿轮驱动调节离心机过高
进泥量过大,超出承受范围
进泥浓度过大
检查皮带松紧度
0010 Fail
检查电机断路器
检查是否电机热敏PTC工作
主电机过载
0011 Failure secondary motor
辅电机故障
停止离心机
检查变频器显示报警
检查电机断路器
0009 Emergency stop
急停
停止离心机
检查急停位置
检查电气线路
0029 Failure decanter running without product
离心机无进料运行
停止离心机
离心机在无进料情况下长期运行(1800s)
进泥的信号没有输入到PLC
停止进料
检查电机断路器
检查泥泵是否干运转保护装置投入工作
备注:修改参数密码:4740
0001 Failure broken wire torque
转矩线路破坏
停止离心机
检查变频器和plc间电缆
检查变频器参数
0033 Failure flocculent station加药站故障
报警
检查加药站确定具体故障
0035 Failure solids screw conveyor螺旋输送器故障
0013 Failure differential speed low差速低
停止离心机
检查接近开关的PLC信号输入
检查皮带松紧度
0015 Failure torque stage 1
转矩达到阶段1(80%)
停止进料
进泥量过大,超出承受范围
进泥浓度过大,控制进料量

实验室高速离心机常见故障及处理方法

实验室高速离心机常见故障及处理方法

实验室高速离心机常见故障及处理方法实验室高速离心机是现代生命科学研究中不可缺少的设备,在细胞分离和样品分析中有着广泛的应用。

由于设备的使用频率高,存在较多的故障情况。

本文将介绍实验室高速离心机常见故障,以及解决问题的方法。

故障1:离心机转子快速脱落若实验室高速离心机转子脱落速度快,则需要立即停止使用设备。

通常这种情况是由于装置锁扣未正确锁定导致的,或转子与设备之间的联接处出现问题。

具体解决方法为:1.确认锁扣系统已正确连接2.检查装置和设备之间的联接处,检查摩擦垫和设备表面是否清洁3.如果问题依旧,需要联系生产商或维修人员寻求更好的解决方法。

故障2:离心机操作不平,振动实验室高速离心机应该尽可能保持平衡和稳定,如果设备出现晃动或振动,则说明存在问题。

这种情况的一个常见原因是设备和转子之间存在不平衡问题,导致设备不能平衡地旋转。

解决此类问题的方法包括:1.检查转子是否平整,不平整的转子需要更换2.检查离心机是否使用在平衡的表面上,重新放置设备位置3.清除设备表面上堵塞的物料4.正确装载样品,尽量将样品分配在转子的两侧,避免单侧重复,从而导致不平衡的情况。

故障3:离心机噪音异常大若实验室离心机出现突然增加的噪音,则说明存在问题。

这种情况的一个常见原因是设备内的电动机发生了故障。

解决此类问题的方法包括:1.断开电源和其它连接设备的电缆,判断是否有缠绕的物体杂滞在了设备内2.修理或更换电动机3.确认设备的承重能力,减少设备转速,或更换新的转子抑或更换设备。

故障4:离心失败,需再次操作某些情况下,离心的样品未被足够的分离,需再次操作。

可能的原因包括:1.选用的离心管太小,超出离心机的承载范围2.没有为设备和样品正确定位和排除气泡3.设备转数不够高,需要提高转数或设备更换在操作实验室高速离心机时,注意观察设备的运行情况,并在设备出现问题时采取适当的措施来解决问题。

在使用设备之前,通常应对其进行定期维护,以确保设备安全和稳定。

台式高速离心机常见故障及故障代码与排除处理解决方法

台式高速离心机常见故障及故障代码与排除处理解决方法

台式高速离心机常见故障及故障代码与排除处理解决方法1.电源故障常见故障代码:机器无法启动,显示屏无反应。

解决方法:1)检查电源插头是否松动或损坏,确保电源连接正常。

2)检查电源线路是否断电,尝试更换电源插座或使用电压稳定器。

3)检查设备的保险丝是否熔断,如有需要更换新的保险丝。

2.转速控制故障常见故障代码:设备转速不稳定,转速无法调节。

解决方法:1)检查控制器连接是否松动或损坏,重新连接或更换控制器。

2)检查转速传感器是否损坏,如有需要更换新的传感器。

3)检查转速控制器的参数设置是否正确,调整参数设置使之符合要求。

3.不均衡故障常见故障代码:设备产生剧烈震动或噪音。

解决方法:1)检查离心轨道是否均匀,如不均匀则需重新安装离心轨道。

2)检查离心机转盘上的样品是否均匀放置,如不均匀需重新调整样品的位置。

3)检查设备的均衡装置是否损坏,如有需要修复或更换新的均衡装置。

4.温度过高故障常见故障代码:设备运行时温度过高,显示屏提示。

解决方法:1)检查设备的通风系统是否正常工作,如不正常需清洁或维修通风系统。

2)检查设备的冷却系统是否正常工作,如不正常则需要修复或更换冷却系统的部件。

3)检查设备的传热介质是否充足,如不充足则需添加足够的传热介质。

5.声音异常故障常见故障代码:设备发出异常的噪音。

解决方法:1)检查设备的传动系统是否损坏,如有需要修复或更换传动系统的部件。

2)检查设备的轴承是否磨损或缺乏润滑,如有需要更换新的轴承或添加润滑油。

3)检查设备是否有异物或杂质进入,清理相关部件。

贝克曼库尔特实验室自动化系统常见故障与处理

贝克曼库尔特实验室自动化系统常见故障与处理

贝克曼库尔特实验室自动化系统常见故障与处理姜宇海;余伟;汤玲;石萍;王亚平【期刊名称】《临床检验杂志》【年(卷),期】2012(030)011【总页数】1页(P931)【关键词】实验室自动化系统;报警;处理【作者】姜宇海;余伟;汤玲;石萍;王亚平【作者单位】无锡市第二人民医院检验科,江苏无锡214002;无锡市第二人民医院检验科,江苏无锡214002;无锡市第二人民医院检验科,江苏无锡214002;无锡市第二人民医院检验科,江苏无锡214002;无锡市第二人民医院检验科,江苏无锡214002【正文语种】中文【中图分类】R446.1本院2010年12月引进美国贝克曼库尔特(Beckman Coulter)公司的实验室自动化系统,包括:入口模块、条形码校验模块、离心模块、血清液面检测模块、去盖模块、智能分杯模块、仪器连接与检测模块、加盖模块、样本储存模块和出口模块,其中生化检测模块配置的是Olympus AU5421全自动生化分析仪,是亚太地区第一条配置Olympus AU5421全自动生化分析仪的贝克曼库尔特公司的检验流水线。

经过1年多的使用,获得了一些故障处理的体会,总结如下。

1 流水线常见问题报警及处理一般情况下,当流水线出现问题时相应模块的显示器上都会出现报警(指示灯闪烁和报警音),并显示错误代码。

此时可以根据使用手册上的错误代码表,找出问题原因,对问题进行处理。

1.1 入口模块有样本被转移到Pending架 (1)如果是条形码粘贴或打印出了问题,就重新打印一张贴上;(2)如果是未签收样本,就签收后再放回流水线(因为我科所有样本必须经人工签收后才能上流水线工作);(3)如果是非线上样本,另行处理。

1.2 条形码校验模块报警0_11 可能是条形码粘贴不正确或条形码左边预留空白不足2 mm,必须重新打印一张贴好。

1.3 离心机报警7_22、7_23、7_24、7_25、7_26 表明在离心机抓手上的样本管固定错误(可能散落在离心机平台上或离心机内)。

BECKMANCX9故障报警word文档版

BECKMANCX9故障报警word文档版

BECKMANCX9故障报警word文档版主键盘各键意义STOP : 停止键,平时不按。

如果机器在运行时按此键,机器会马上停下来,浪费试剂。

Start: 开始键,编好样本项目及样本位置(或选定定标项目及定标液位置)后,如果机器在standy 状态,须按此键,机器才会运行。

Paus e: 暂停键,若机器在运行中,按此键,机器会把已吸取试剂的项目做完,然后停下,不浪费试剂。

Sys Home: 系统复位键,由于外力使机器一个或几个机械臂偏离了原来的位置时,可按此键使机器回到原位。

机器在运行中按了STOP 键后,也须按此键使机器复位。

Sys Idle: 系统休眠键,按此键可使机器进入休眠状态,然后可进入关机程序。

Prev Screen: 回上一屏幕键,也称退出键。

按此键可回到上一屏。

select: 选择键。

Clear: 清除键。

开关机(一) 如何关机1 当机器在STANDY状态时按[Sys Idle ] 键2 机器会提问一个问题(是否进入休眠状态),这时按y 键, 回车3 按F4(resume) 键4 机器会提问一个问题,这时按y 键, 然后回车5等待一会儿, 屏幕左上角会出现闪烁的一行"Press to continue ",这时按[Prev Screen] 键.6 按数字键" 1 "7 出现红色提示框, 关闭机器主电源.(二)如何开机如果关机是严格按上述步骤,则只须打开机器主电源,机器约十分钟后会自动回到主屏幕如果是非正常关机:1 打开电源后,过几分钟,机器会报警,并出现红色提示框, 此时按[Prev Screen]2 等一会儿后,屏幕上会出现F1(RESTORE)F2(BACKUP)F3(REBUILD DATABASE)F4(RESUME),选择F4。

3 机器会提问一个问题,这时按y 键, 然后回车4 等待一会儿,屏幕左上角会出现闪烁的一行"Press to continue ",这时按[Prev Screen] 键.5 按数字键" 2 "6 等一会儿后,机器会回到主屏幕.定标(一)操作步骤1 主菜单下按F3 定标2 移动光标并用选择键选上要定标的项目3 按F1 架/杯号设定4 输入定标液所放的架号,并设定杯号5 退回主屏幕6 按开始键(二)定标强行通过(overridden)1 主菜单下按F3 定标2 按F3 定标强行通过3 输入所要强行通过定标的试剂位置(Loc)号,回车(三)取消试验项目(Chemistry bypass)1 主菜单下按F3 定标2 按F4 取消试验项目3 输入所要取消试验项目的试剂位置(Loc)号,回车(四)延长定标时间(extend)1 主菜单下按F3 定标2 按F5 延长定标时间3 输入所要延长定标时间的试剂位置(Loc)号,回车(五)定标选择中各项的内容:1.定标磁盘装载2.查看定标结果的可接受范围3.修改定标液定值的设定点(改过设定点后,仪器会要求重新定标)4.修改斜率/截距5.批内定标(建议所有项目全选上)6.重新打印定标报告7.设定定标液名称样本编程(一)单个编程1 主菜单下按F1 样本编程2 按F1 样本单个编程3 输入样本号,回车4 输入组合号和/或选择项目,如果是急样本则按F4 STAT如果是其它类型样本(如尿液)则:①在选好项目后,回车②光标移动到样本类型选项(sample type) ,③按F1选项④把光标移动到正确的样本类型⑤回车5 按F8 下一样本或按【Prev Screen】键退出(二)批量编程(用于多个样本所做的项目都相同)1 主菜单下按F1 样本编程2 按F2 批量编程3 输入组合号和/或选择项目4 按F7 样本号输入5 依次输入各样本号6 按F8 结束批量编程或按【Prev Screen】键(三)设定样本架号及杯号1 主菜单下按F1 样本编程2 按F6 架/杯号设定3 输入样本所放的架号,并设定杯号4 按F8 下一架或?按【Prev Screen】键退出(四)清除样本方法一:按样本号清除(每次只能清除一个样本)1 主菜单下按F1样本编程2 按F5 样本清除3 选第一项按样本号清除4 输入样本号5 回车方法二:按序列号清除(一次可清除多个样本)1 主菜单下按F1 样本编程2 按F5 样本清除3 选第三项按序列号清除4 输入序列号(可同时输入多个序列号)5 回车(五)清除架号1 主菜单下按F1 样本编程2 按F6 架/杯号设定3 按F1 清除架号4 输入架号(可以同时输入多个架号)5 回车6 按“Y”键确认特殊功能中的常用项(仪器冲洗、结果查询、系统设置)(一)仪器冲洗1 主菜单下按F4 特殊功能2 选第1 项冲洗3 移动光标,用选择键选择要清洗的项目4 按F1 开始5 如果有选CX3 部分的冲洗项目,则还须输入冲洗次数,然后回车(二)结果查询1 主菜单下按F4 特殊功能2 选第2 项结果查询3 选择查询方式①按架号查询②按架号/杯号查询③按样本号查询④按序列号查询4 输入相应数值5 选第1 项在屏幕上显示结果选第2 项在英文打印机打印结果选第3 项把结果送到中文电脑(四)系统设置主菜单下按F4 特殊功能,其中的第4 项为系统设置,共有14 项内容)1 定义试验菜单2 定义条码组合3 条码设置4 定义参考范围5 报告设置(其中的F4打印设置建议选第2 项disable)6 特殊计算(用于定义计算项目的计算公式)7 单位选择(可用选择键改变各项目的单位及精确度)8 自动稀释(建议全部选上)9 设置日期时间及温度10 通讯参数11 试验重复次数12 定义备注13 病人结果设置(建议全部设置为disable)14 定义可报告范围故障说明及解决方法CX4的错误信息1. AIR PRESSURE NOT IN RANGE(AIR PRESSURE IS TOO LOW OR HIGH)气压不在范围(气压太低或太高)a)检查相关管路和瓶盖是否有漏气。

三相卧式离心机常见故障及解决方法

三相卧式离心机常见故障及解决方法
排出的固渣太多,来不及排放;或设定的扭矩值太低
排出的固渣太多,来不及排放;或设定的扭矩值太低 频 繁 启停 电机 ,使 电机 温度 过高 .停 机, 等电 机冷 却 后再 启 动. 电机电流过大引起热保护 空气开关跳闸,检 查电机电流,等 开关冷却后复位.
故障名称
S73 固渣泵变频器故障
S74 固渣泵跳闸
检查供油系统 机电流,等 开关冷却后复位. 频 繁 启停 电机 ,使 电机 温度 过高 .停 机, 等电 机冷 却 后再 启 动. 检查开启方向的限位开关.如果偏掉了,予以调节到位. 检查开关闭向的限位开关.如果偏掉了,予以调节到位. 检查保护气体气源 保护气体流量低,手动调节进气阀到设定流量. 保护气体压力低,手动调节进气阀到设定压力. 检查保护 气体气源 ,使之在 规定的 时间内 达到压力 河流量 的要求.
S45 液压驱动油压超绿色限位/切断进 料
S46 液压驱动油压超红色限位,停机 S49 螺旋输送机变频器故障 S50 螺旋输送机跳闸
S54 油脂泵跳闸 S55 油脂流量低 S56 油脂罐空 S57 进料泵变频器故障 S58 进料泵跳闸
S59 进料泵电机温度高
S60 进料泵管压高
S61 进料泵干运行 S62 进料泵运行但排渣连锁不正确 S65 固渣插板阀开启扭矩高 S66 固渣插板阀关闭扭矩高 S67 固渣插板阀温度保护动作
S4 SIMP 齿轮箱油位超低 S5 螺旋驱动扭矩高 S6 螺旋驱动扭矩超高 S7 SIMP 齿轮箱油温高 S8 转鼓转速超高仪表保护,停机 S9 急停开关动作
故障解决方法
看变频器故障报警信息(可以从手册上查阅解决的方法)
频 繁 启停 电机 ,使 电机 温度 过高 .停 机, 等电 机冷 却 后再 启 动. 根据 FLOTTWEG 的操作手册对齿轮箱加油,并检查齿轮 箱渗漏情况 根据 FLOTTWEG 的操作手册对齿轮箱加油,并检查齿轮 箱渗漏情况 扭矩太高--进料停止.减少进料,增加差转速 扭矩太高--离心机停止工作. 扭矩太高--减少进料,降低进料浓度 离心机由 于速度太 高而停 止(硬件 上就是变 频器频 率>63Hz) 紧停开关被按下,检查按钮

第四节 离心机的使用维护与常见故障处理全套PPT

第四节 离心机的使用维护与常见故障处理全套PPT
检查真空泵表及油压指示值 清理整流子及电刷,使其接触良好,或者更换。
离心过程中不得开启离心室盖,不得用手或异物碰 第四节 离心机的使用、维护与常见故障处理
每次使用前要严格检查转头孔内是否有异物和污垢,以保持平衡。 每次使用后,都必须仔细检查,并用50℃~60℃的温水及中性洗涤剂浸泡清洗或定期用消毒液消毒(每周消毒一次),最后用蒸馏水冲洗
返回知识点
第四节 离心机的使用、维护与常见故障处理
6.低温离心样品时,应先将空的转头在2x103rpm预 冷一定时间,预冷时温度控制在0℃左右,也可将转头 放在冰箱中,预冷数小时备用,离心杯可直接存放在冰 箱中预冷。
返回知识点
第四节 离心机的使用、维护与常见故障处理
7. 每隔三个月应对离心机主机校正一次水平度,每使 用5亿转处理真空泵油一次,每使用1500小时左右,应清洗 驱动部位轴承并加上高速润滑油脂,转轴与转头接合部应 经常涂酯防锈,长期不用时应涂防锈油加油纸包扎。
2.转头孔内有异物,负荷不平衡或使用了不合格的试管 套。 3.转轴上端固定螺帽松动,转轴摩擦或弯曲。
4.电机转子不在磁场中心会产生噪音。 5.机座上减震弹簧的固定螺丝松动或其中一根弹簧断裂。
6.转子本身损伤。
返回知识点
3.离心过程中应随时观察离心机上的仪表是否正常 工作,如有异常应立即停机检查,及时排除故障。未 找出原因前不得继续运转。
返回知识点
第四节 离心机的使用、维护与常见故障处理
4.每次使用前要严格检查转头孔内是否有 异物和污垢,以保持平衡。每次使用后,都必 须仔细检查,并用50℃~60℃的温水及中性洗 涤剂浸泡清洗或定期用消毒液消毒(每周消毒 一次),最后用蒸馏水冲洗,软布擦干后用电 吹风吹干、上蜡、干燥保存。

beckman常见报警

beckman常见报警

CX3与CX4维修常见问题与解决方法44. MOTION ERROR:SAMPLE MIXER UP/DOWN CHECK MIXER 定位--- ALIGNMENT SAMPLE MIXER VERTI CAL & HEIGHTMOTOR & CONNECTORHOME SENSORBELTMSC 2 BOARDSAMPLE MIXER ROTATE CHECK MIXER 定位及是否UP位置--- ALIGNMENT SAMPLE M IXER ROTATIONMOTOR & HARNESS CONNECTORSENSORMSC 2 BOARDSAMPLE MIXER SPIN CHECK SPEED CONTROLLER BOARDMOTOR & CONNECTORBEARINGMSC 2 BOARD45.MOTION ERROR—REAGENT PROBE UP/DOWN CHECK PROBE 定位--- ALIGNMENT REAGENT PROBE VERT ICAL HEIGHTMOTOR (P898头)SENSOR (P896头)BELTREAGENT CARTRIDGE--- 有无气泡MSC 3 BOARDREAGENT PROBE ROTATE CHECK PROBE定位及是否UP位置--- ALIGNMENT REAGENT P ROBE ROTATIONMOTOR (P873头)HOME SENSORMSC 3 BOARD46.MOTION ERROR:REAGENT MIXER ROTATE CHECK MIXER 定位及是否UP位置--- ALIGNMENT REAGENT MIXERMOTORHOME SENSORMSC 3 BOARDREAGENT MIXER UP/DOWN CHECK MIXER 定位--- ALIGNMENT REAGENT MIXER VERTI CAL & VERTICALMOTORBELTUP HOME SENSORREACTION CAROUSEL 位置MSC 3 BOARDREAGENT MIXER SPIN CHECK SPEED CONTROLLER BOARDMOTOR, BEARING & CONNECTORMSC 3 BOARD47. MOTION ERROR:REAGENT CAROUSEL CHECK CAROUSEL 有无松动--- ALIGNMENT有无CARTRIDGE 未推入位置HOME SENSORPOSITION SENSORBELTMOTORHARNESS & CONNECTORMSC 1 BOARDSAMPLE CAROUSEL CHECK 有无异物--- ALIGNMENTMOTOR, HARNESS, CONNECTORPOSITION SENSORMSC 5 BOARD48.MOTION ERROR:AUTOLOADER UP/DOWN ERROR CHECK PRIME PRESSUREMSC 5 BOARDMAC VALVE15PSI REGULATOR & RELIEF VALVE相关的HARNESSUP/DOWN SENSORAUTOLOADER SPIN CHECK AUTOLOADER ASSEMBLYMOTORFLAG & SENSORMSC 5 BOARDWASH STATION CHECK MOTOR & CABDESENSOR--- ADJUST (ENTER DIAGNOSTICS)REACTION CAROUSEL 之CUVETTE 位置--- ALIGNMENT REACTION C AROUSELMOTOR 之导螺杆---有无动并润滑之49. MOTION ERROR:SAMPLE PROBE WASH CHECK 8L SOLENOID VALVE & CABLECHECK MSC 4 BOARDSAMPLE PROBE AIR STRIP CHECK 4A SOLENOID VALVE & CABLECHECK MSC 4 BOARDREAGENT PROBE WASH CHECK 7L SOLENOID VALVE & CABLECHECK MSC 4 BOARDREAGENT PROBE AIR STRIP CHECK 5A SOLENOID VALVE & CABLE CHECK MSC 4 BOARDWASH TO CUVETTE WASH CHECK 9L SOLENOID VALVE & CABLE CHECK MSC 5 BOARDCUVETTE VACUUM CHECK 10V SOLENOID VALVE & CABLECHECK MSC 5 BOARDWASH MIX VALVE CHECK 4L SOLENOID VALVE & CABLECHECK MSC BOARD50.MOTION ERROR:SUMP PUMP CHECK 5L SOLENOID VALVE & CABLECHECK MSC 2 BOARDSIP VALVE CHECK 11V SOLENOID VALVE & CABLECHECK MSC 2 BOARDVACUUM RESERVOIR DUMP CHECK 5L SOLENOID VALVE & CABLE CHECK MSC 5 BOARDDI WATER MIXER VALVE CHECK 3L SOLENOID VALVE & CABLECHECK MSC 4 BOARDWASH TO ISE CHECK 6L SOLENOID VALVE & CABLECHECK MSC 4 BOARDDI WATER TO CUVETTE WASH CHECK 2L SOLENOID VALVE & CABLECHECK MSC 5 BOARD51. MOTION ERROR:MIX BOTTLE 1 CHECK 1A SOLENOID VALVE & CABLEMSC 1 BOARDMIX BOTTLE 2 CHECK 2A SOLENOID VALVE & CABLEMSC 2 BOARDSAMPLE PROBE WASH CHECK 8L SOLENOID VALVE & CABLEMSC 4 BOARDSAMPLE PROBE AIR STRIP CHECK 4A SOLENOID VALVE & CABLEMSC 4 BOARDREAGENT PROBE WASH CHECK 7L SOLENOID VALVE & CABLEMSC 4 BOARDREAGENT PROBE AIR STRIP CHECK 5A SOLENOID VALVE & CABLEMSC 4 BOARD52. MOTION ERROR:REACTION CAROUSEL CHECK MOTOR & CABLECHECK 有无异物---注意WASH CUP 之出气FILTER(在HYDRO 上方) CHECK PROXIMITY SENSOR---ENTER DIAGNOSTIC RUN SENSOR ALIGNMENTCHECK MSC 1 BOARDENTER DIAGNOTIC SYSTEM ( EXTENDED MODE )--- RUN 7. REACTION CAROUSEL SYSTEM1. REACTION CAROUSEL3. STEPPER/SENSOR1. HOMING TEST2. START-UP RELIABILITY TEST3. CORRELATION TEST53. MOTION ERROR:SAMPLE VALVE CHECK MSC 4 BOARDVALVE MOTOR---J838 接头VALVE SENSOR--- J840 接头VALVE SENSOR DISCCABLE BETWEEN MSC 4 BOARD AND PUMP INTERFACE BOARD SAMPLE PUMP LUBRICATION UP/DOWNCHECK MOTOR, 接头J842,CABLEFLAG DISC , HOME SENSORSAMPLE VALVESAMPLE SYRINGE & PROBEREAGENT VALVE CHECK MSC 4 BOARDVALVE MOTOR-- J837 接头VALVE SENSOR-- J839 接头VALVE SENSOR DISCCABLE BETWEEN MSC 4 BOARD AND PUMP INTERFACE BOARD REAGENT PUMP LUBRICATION UP/DOWNCHECK MOTOR, 接头J837,CABLEREAGENT VALVEREAGENT SYRINGE & PROBE54.MOTOR ERROR:REAGENT DOOR IS OPEN CHECK SENSOR---ADJUST SENSOR POSITIONCHECK CABLECHECK INTERFACE BOARDPROBE RINSE 没消耗 INLINE FILTERCHECK BUBBLE CENERATOR--- SOLENOID VALVE (25PSI & RINSE ) 有无动作CHECK 10PSI PRESSURE--- WATER PRESSURE REGULATOR, RELIFE VA LVEFASH BOTTLE 没有水 CHECK SOLENOID VALVE (3L & 4L )CHECK SOLENOID VALVE (1A, 2A )55. AUTOLOADER TIMEOUT CHECK 25PSI PRESSURE15PSI REGULATOR, RELIFE VALVEAUTOLOADER SOLENOID VALVEAUTOLOADER 有无漏气UP/DOWN SENSORMSC VALVE (PTS)INLINE FILTER (PTS )ENTER DIAGNOSTIC---ADJUST UP/DOWN TIMING (PTS)56. MOTION ERROR:BAD MESSAGE INTERRUPT ON MMC CHECK MMC BOARDUNDEFINED MSB ERROR CHECK MSC BOARDNO GOLDEN PHASE ON REAGENT CHECK MSC 2 BOARDNO GOLDEN PHASE ON SAMPLE CHECK MSC 3 BOARDNO GOLDEN PHASE ON REAGENT CAROUSEL CHECK MSC 1 BOARDNO HOME ON REAGENT CHECK REAGENTNO HONE ON SAMPLENO HOME ON REAGENT CAROUSEL57.MOTION ERROR:ERROR IN MMC COMMUNICATION CHECK MMC BOARD--- REBOOTREPLACE MMC BOARDBAD PACKET CHECKSUM CHECK MMC BOARD--- REBOOTMAC PROCESSOR OR EPROM FAILURE CHECK MSC BOARD--- CR24, CR25 是否亮,REP LACE MSC BOARDREBOOTCHECKSUM ERROR IN MSC DOWNLOAD CHECK POWER SUPPLY 3MMC/ MSC BOARD--- REBOOTBAD INIT HANDSHAKE WITH MSC CHECK MSC BOARD--- ROBOOT58. MOTION SYSTEM IS BUSY IDLE SYSTEM THAN REBOOTBAD ALIGNMENT DATA DETECTED RESTORE BACKUP DISKETTEREBOOT59. MOTION ERROR:CX3 PROBE NOT CLEAR HOMECHECK SAMPLE CRANE+13V POWER SUPPLYMDR 1 BOARDMMC BOARD60.AIR PRESSURE OUT OF RANGE STANDBY 25PSI REGULATOR 是否已调到底--- C HECK有无漏气RUNNING 时,气体建立是否快--- CHECK INLINE FILTERPUMP A3-A4 TUBEPILOT OPERATOR SWITCH水压有无上下漂移---CHECK RELIFE VALVECHECK SENSORCHECK COMPRESSOR--- REPLACE REBUILT KIT61. WATER PRESSURE LOW OR HIGHT CHECK 25PSI PRESSURECHECK REGULATOR, RELIFE VALVECHECK有无漏气CHECK WASH BOTTLE & 1A, 2A VALVECHECK SENSOR62. LIQUID TRAP TOO HIGH CHECK PRIME PRESSURE--- 见25PSI PRESSURE LOW3 WAY VALVE (5L) 有无动作, 25PSI AIR进入把WASTE赶入WASTE BOTTLE---REPLACECHECK MSC 5 BOARD--- CONTROL 5A VALVESENSORRESERVOIR --- CLEAN63. VACUUM BOTTLE FULL WATER CHECK 3 WAY VALVE (5L)CHECK 1L, 6L VALVE 与其连接之CHECK VALVECHECK PILOTOPERATOR SWITH 动作LIQUID TRAP 或有积水64. VACUUM PRESSURE LOW 有无漏气TANK有无锁紧CHECK 5L, 10V, 1L CHECK VALVE6L CHECK VALVE出气FILTER有无漏气CHECK SENSOR65. SECONDARY PRESSURE HOMELOW OR HIGH CHECK 3A VALVECHECK REGULATOR P/N 759858CHECK WASH CONCENTRATE 2 BOTTLECHECK SENSOR66. DI WATER LEVEL LOW CHECK 水系统(进水压力要大于30PSI )CHECK WATER PRESSURE--- REGULATER & RELIFE VALVECHECK SENSOR 及接头(J985 & J984 )MPT VALVE是否关住(进水关闭)进水VALVE, 接头P818--- MSC 3 BOARD67. POWER SUPPLY FAILURE CHECK STATUS BOARDCHECK INDICATOR BOARDPOWER SUPPLYPOWER SUPPLY1 V1: +5V LOGIC, DISK DRIVERV2: +24V HEATERS, LAMP SUPPLYV3: +12V LOGIC, DISK DRIVER, COMMUNICATIONV4: -12V COMMUNICATIONPOWER SUPPLY2 V2: +24V REFRIGERATORV3: +15V ANALOGV4: -15V ANALOGPOWER SUPPLY3 V1: +5V MOTOR LOGICV3: +38V MOTORSV4: +12V SOLENOID VALVE68. PRINTER NO FUNCTION CHECK PRINTER CABLERESET PRINTERRESTART PRINTERCHECK ISBX PRINTER INTERFACER BOARD (ONCPU BOARD)RESET INSTRUMENT69. CUVETTE FAILED WATER BLANK TEST CUVETTE---CLEANCUVETTE---CHECK WASH STATIONALIGNMENT WASH STATIONCHECK REACTION CAROUSEL 定位CUVETTE有无淹水--- CHECK WATER PRESSURE & 7L, 8L VALVE VACUUM PRESSURE (5L VALVE)VACUUM/WASH PROBECHECK VALVELAMP 有无FLASH---CHECK REACTION WHEELLAMP POWER SUPPLYMAC BOARDSYNC PROXIMITY SENSOR70. SECTOR CAN’T CLEAR REBOOT71. SECTOR 不RUN SECTOR LABEL, 有问题CHECK HYDRO, SYSTEM 左上方SECTOR READER BOARD之二个L ED在READ SECTOR有无一闪一闪---一为DATA TRACK二为TIMING TRACK72. CABTRIDGE CAN’T AUT OLOAD LABEL有问题CHECK BARCODE READERCHECK INTERFACE BOARD73. WASTE BUCKET FULL DRAIN PUMP 是否工作正常---REPLACE DRAIN PUMP OR CLEANCHECK SSR & CABLEWASH BOTTLE 1 & 2 有无气泡---REPLACE 其上方CHECK VALVECHECK FLOAT SWITCH (WASTER BOTTLES内)有无动作不良,因此造成CHECK FLOAT SWITCH (WASH BOTTLE 1 & 2 内)动作是否正常74. COMPRESSOR没动作 CHECK K1 SSRSTATUS BOARDAC/ DC CONTROL BOARD (在PS8下方)MOTOR & CABLE75. BUBBLE GENERATOR无气泡产生:OLD TYPE 针阀有无动作--- CHECK接头J928POWER +12V POWER SUPPLYSOLENOID VALVE (25 PSI PRESSURE, RIGHT)---PRIME PRESSURE---CHECK PROBE RINSE 限制器& FILTERCHECK RINSE SOLENOID VALVE (LEFT), WATER PRESSUREADJUST 10 PSI REGULATORCHECK BLUE PRESSURECHECK MSC 3 BOARDNEW TYPE CHECK接头J928POWER +12V POWER SUPPLYSOLENOID VALVE---J1: PROBE RINSEJ2: WATER PRESSURE (BLUE TUBE)J3: PROBE RINSE/ WATER (SYRINGE)CHECK PROBE RINSE限制器(二个) & FILTERCHECK MSC 3 BOARD76. REAGENT PROBE 漏水 CHECK BUBBLE GENERATORREAGENT PROBE 位置偏ALIGNMENT REAGENT PROBECHECK 25 PSICHECK WASH CUP INSERT 5A VALVE, MSC 4 BOARDSYRINGE & 3 WAY VALVE77. SAMPLE PROBE 滴水或悬水珠 SAMPLE PROBE定位问题CHECK 25 PSIWASH CUP INSERT 4A VALVE, MSC 4 BOARD78. WASH CUP 有水气喷出CHECK HYDRONMENTIC SYSTEM---WASH CUP 排气FILTER 有无下垂或阻塞79. SAMPLE NOT DETECTED 周期性 CHECK PROBE CABLE LEVEL SENSOR BOARD干扰GROUNDINGSAMPLE CAROUSEL ALIGNMENTSAMPLE PROBE ALIGNMENT80. REAGENT NOT DETECTED AT SAMPLE 周期性发生CHECK SAMPLE PROBE CABLEINJECTION REAGENT STATUSREAGENT PROBEBUBBLE GENERATORREAGENT PUMP---SYRINGE, 3 WAYVALVE看EVENT LOG---那个CUVETTE发生CHECK CUVETTE那个CHEMISTRY REAGENT有气泡REAGENT CARTRIDG破CUVETTE 位置错误 ALIGNMENT REACTION CAROUSELNOISE 干扰 GROUNDING问题81. PELTIER FAILURE CHECK 6个HEATSINK 电压REPLACE HEATSIN KCHECK POWER SUPPLY 1CHECK PELTIER CONTROL BOARDCHECK STATUS BOARD82. PELTIER TEMPERATURE HI CHECK 室温,CLEAN AIR FILTER 工作温度为18度至28度CHECK PELTER REPLACE HEATSINKCHECK REAGENT CAROUSEL 内温度 REPLACE THEMISTORCHECK POWER SUPPLY 1CHECK PELTIER CONTROL BOARDSTATUS BOARD83. POWER SUPPLY FAILURE (CX3* /CX3 )CHECK INDICATOR BOARD (ONLY CX3* )REPLACEBOX 1 (POWER INPUT )---SUPPLY BOX 2~5, +17V HEATERS POWER SUPPLYBOX 2 (+5V, -5v, +12V, -12V)---LOGIC, DISK DRIVERBOX 3 (LAMP POWER SUPPLY )---CA / CREA LAMPBOX 4 (+15V, -15V, +5V, +24V)---ANALOG, SOLENOID VALVE, MOTORBOX 5 (+13V )---STEP MOTOR84. POWER SUPPLY FAILURE ( ISE ) PSI & 24V FREQUENCY BOARD 同时坏---CHEC K INDICATOR BOARDBREAKERPSI24V FREQUENCY BOARD REPLACE UPDATE KITCHECK STATUS/ SENSE BOARD & C32PSI POWER SUPPLY: +5V, -5V, +15V, -15V, +13V24V FREQUENCY BOARD: 24VAC85. DBM FATAL ERROR # 0 SHUTDOWN INSTRUNMENT 二次( POWER DOWN )REBOOTDBM FATAL ERROR # 8 DELETE DATABASEREBOOT INSTRUNMENTDBM FATAL ERROR # ? ?86.ANALOG CONTROL SENSOR ERROR CHECK PROX SENSOR RUN REACTI ON EXTENDED TESTCHECK REACTION CAROUSEL ALIGNMENT CAROUSELCHECK EVENT LOGMAC DETECTED: TYPE#5, 23 VERIFY HOME / SERVICE SENSOR( HOME STACK ) CHECK CABLE BETWEEN HOME SENSOR AND MAC BO ARDMAC BOARDMAC DETECTED: TYPE#6 VERIFY HOME SENSOR( HOME TIME-OUT) CHECK MARGINAL REACTION CAROUSEL MOTOR VOLT AGEMAC 3 BOARD, MMC BOARDMAC DETECTED: TYPE#7 CHECK SYNC SENSOR( SYNC STACK ) CABLE BETWEEN SYNC SENSOR AND MAC BOARD MAC BOARDMAC DETECTED: TYPE#20 CHECK SYNC SENSOR(NO FLASH SYNC) REACTION WHEEL MOTORMAC BOARDMAC DETECTED: TYPE#21 CHECK HOME / SERVICE SENSOR (MISSED CUVETTE ) MSC / MMC /MAC BOARDMAC DETECTED: TYPE#22 CHECK SYNC SENSORSENSOR ALIGNMENT87.ANALOG COMMUNICATION ERROR CHECK EVENT LOG---FLASH POWER FILE NOT FOUND REBOOT AFTER RESTORE BACKUP DISKETTERUN ANALOG SYSTEM ALIGNMENTCHECK DISK OPERATION SYSTEMFLASH DELAY FILE NOT FOUND 如上ERROR SENDING FLASH DATA CHECK MAC BOARD, DISK OPERATION SYSTEMREBOOTRESULTS BLOCK CHECKSUM ERROR CHECK MULTI-BUS BAC K PLANEREBOOTRESULTS MSG ADDRESS ERROR REBOOTRESULTS OUT OF SEQUENCE REBOOTRECEIVED RESULTS BEFORE READY REBOOTERROR SENDING CHEM. DATA REBOOT, CHECK MAC BO ARDERROR STOPPING MAC 如上ERROR STARTING MAC 如上ERROR SENDING WORKLIST REBOOTWORKLIST QUEUE IS EMPTY REBOOTCAN NOT START PROCESS REBOOT88.ANALOG COMMUNICATION ERROR CHECK EVENT LOG---CAN NOT CREATE SEMAPHORE REBOOTCAN NOT DOWNLOAD CODE REBOOTMAC IS DEAD CHECK MAC BOARD, REBOOT89.FATAL MMC ERROR:MMC DETECTED NO COMM. FROM HOSTTOO MANY SUB SEQUENCESBAD TERNINATE FUNCTIMEOUT COURTING MSCTIMEOUT READING MSC BAD PORTBAD DIAG SEQUENCE FUNCBAD SEQUENCE FUNCBAD DEVICE--- REBOOT REPLACE MMC BOARD90.CHECKSUM ERROR POWER DOWN, REBOOT CHECK CPU / RAM BOARDSTACK OVERFLOW POWER DOWN, REBOOT CHECK CPU / RAM BO ARDHARDWARE MENORY PARITY ERROR CHECK CPU / RAM BOARDPOWER DOWN, REBOOTSOFTWARE ERROR DETECTED CHECK DISKETTE ( CX3 )DISK CONTROL BOARDHARD DISK ( CX4/5 )SCSI BOSRD, HP PCOPERATING SYSTEM ERROR 如上UNDEFINED APPLICATION ERROR 如上91.MMC TIMOUT ERROR REBOOTCHECK MMC BOARD92.UNRECOGNIZED COMMAND SENT TO MMC REBOOTREPLACE OPERATOR DISKETTE ( CX3 )UNRECOGNIZED COMMAND RECEIVED TO MMC CHECK MMC BOARDCPU BOARDPARAMSTER SENT TO MMC OUT OF RANGE DISK DRIVER BOARD93. MOTOMETER SYSTEM NOT READY CHECK MAC BOARDINTERFACE BOARD94. ANALOG WAS DEAD AT INITIALIZATION CR23 CR24 亮起---REPLACE MSC BO ARDCHECK POWER SUPPLY 3MMC BOARD95.FATAL NOTE- IOM MODULE: IDLEIOM: Input/Output Module NOT RECOGNIZED MESSAGE RECEIVEDBAD IOM MESSAGEBAD PRINT TASK CREATEIOM TASK IS NOT RUNNING ( BAD SENT )PRINT TASK IS NOT RUNNING ( BAD SENT )NAME ATTACH FOR IOM FAILEDCAN’T OPEN PRINT TASK FILE (BAD FOPEN )UNABLE TO ALLOCATE BUFFER MEMORY----- REBOOT INSTRUMENT AFTER INSTRUMENT IDLE96. FATAL ERROR--- COULD NOT POST ALARM EVENTALARM REQUEST LIST FULLCOULD NOT OPEN ALARM REQUEST QUEUECOULD NOT OPEN EVENT QUEUECOULD NOT CREATE ALARM AQENT TASKCOULD NOT POST ALARM REQUESTCOULD NOT REPORT ALARM EVENTALARM NOT RUNNING---- REBOOT, CHECK HP PC97.SCSI PROBLEM CHECK SCSI BOARD (CX3 / CX4)—JUMPERSCSI BOARD (HP PC)CABLEISOLATER BOARD98.CANNOT START HEATER CALIBRTION CODE CHECK HEATER CONTROL BOARD---CA LIBRATION CODE99. DOES NOT MATCH DATA ON DISK ENTER DIAGNOSTIC---EXTENDED (6 ) TEMP ERATURE CONTROL SYSTEM改正或重输CODERESET INSTRUMENTREPLACE REART CONTROL BOARD100. INCOMPATIBLE TEMPERATURE BOARDS (如上)101. CANNOT START: FATAL ANALOG ERROR CHECK MAC BOARDRESET INSTRUMENT102. CANNOT START: CX4 ANALOG SYSTEM NOT READY REBOOT INSTRUMENT THRO UGH SYSTEM IDLE103. CANNOT START: DATABASE ERROR IDLE, REBOOTDELETE DATABASEREINSTALL, SOFTWARECHECK STATUS MONTTOR104. ERROR SYNCHRONIZING ISE PROBE HOMEAND CX4 SAMPLE CAROUSEL INSTALL SOFTWARE VERSION 1.8/2.2 CHECK CX4 MOTOR CONTROL SYSTEM。

全自动生化分析仪贝克曼AU仪器报警列表中文报警翻译完整版

全自动生化分析仪贝克曼AU仪器报警列表中文报警翻译完整版
5105
REAGENT ID NOT DEFINED
未定义试剂条码
5107
R2 REAGENT FOUND IN THE R1 REFRIGERATOR
在R1试剂仓发现R2试剂
5108
R1 REAGENT FOUND IN THE R2 REFRIGERATOR (A, B, C)
在R2试剂仓发现R1试剂
全自动生化分析仪贝克曼AU仪器报警列表中文报警翻译
报警表
3038
Power Failure Detect
电源出现断电
3039
Reagent Refrigerator Temp. out of Range
试剂冰箱温度超出范围
3212
R1 Reagent Run Out
R1试剂耗尽
3213
R1 Reagent Empty
R1试剂空
3214
Probe Detergent Short
探针清洗液少了
3215
Probe Detergent Empty
探针清洗液空了
3216
ReagentProbeW2DetergentShort
试剂针W2清洗液少了
3217
R2 Reagent Short
R2试剂少了
3218
R2 Reagent Run Out
5109
ILLEGAL BOTTLE TYPE
不规则的试剂瓶
5110
ILLEGAL REAGENT POSITION
不规则的试剂位置
5111
REAGENT EXPIRED
试剂失效
5112
DIFFERENTLOTNO. REAGENT FOUND

离心机常见故障及处理办法

离心机常见故障及处理办法

离心机常见故障及处理办法1. 急停开关消失特别问题分析此类报警一般在离心机消失紧急故障按下急停开关后或设备重启时消失此类报警。

处理方法只需对设备进行复位键,如复位失败需等30s后驱动变频器开关能自动切断后再进行复位即可。

2. 分散单元位置不对问题分析此类报警一般在离心机开启后预备进行操作时或在分散过程中刮刀走到位后消失,由于分散时设备特别停车导致分散循环没有走完,刮刀没有回到起始位置造成;其次,在检修或清理设备完成后及其他特别状况下造成的接近开关与刮刀接近金属片间距较大,从而导致即使刮刀到位也无法检测到,造成假报警;再次,接近开关特别损坏,无法正常工作。

处理方法(1)开启前消失报警:检查刮刀位置,不在起始位置,马上手动分散,将刮刀复位; 使用金属物遮拦住接近开关,若报警消退,则证明是接近开关间隙较大,略微调整间隙,消退报警即可。

(2)假如金属物接触接近开关,报警信号无法消退,说明接近开关已坏,更换备件即可解决。

(3)将2个开关调换,如能消退报警,则可能是在安装时将2个开关位置混淆所致。

3. 浆料进口阀/EG进口阀消失特别问题分析此类报警一般消失在离心开头时或离心结束转分散的过程中,由于浆料进口阀不能打开或关闭,或由于阀门卡阻造成打开、关闭不到位,其次就是位置开关的误报警。

处理方法(1)检查阀门是否开关到位,如开关到位便是位置开关消失故障,调整位置开关位置或更换位置开关; 如阀门没有开关到位,开关到一半的位置极大可能是阀门卡阻,检查气源压力是否满意动作需求,其次检查电磁阀是否动作正常,供电是否正常,手动是否能正常动作,再次检查气缸是否能敏捷动作,最终拆开阀芯检查,是否有异物卡住阀芯。

(2)阀门没有动作检查同阀门没有开关到位步骤全都。

4. 轴温消失特别问题分析此类报警基本都是在离心机运行过程中消失,由于电机温度过高或PT100的不正常工作造成报警或跳停。

处理方法(1)检查电机温度是否达到报警设定值,如未达到则检查PT100是否有损坏,接线是否有松动、脱落或接线断开和破损状况,确认有无特别,检查平安栅有*,输入输出是否全都。

离心机常见故障排除方法及日常维护

离心机常见故障排除方法及日常维护

离心机常见故障排除方法及日常维护本文列举了离心机常见故障排除及日常维护方法:维修案例一:客户反馈某进口品牌离心机在低速时无杂音,高速时杂音很大。

检查发现该离心机使用年限较长,内部电机老化,工程师给电机涂抹润滑油后,离心机运行无噪音。

离心机电机运转会有磨损,并且越到后期磨损越大,经常给电机做维护保养,可以延长离心机使用寿命。

维修案例二:客户反馈一进口离心机在进行离心时,离心管松动,或者飞出离心机。

经检查发现,离心管的直径太小,与离心机转子配合不良导致松动。

解决方法是更换适合的离心管,或者更换配套的离心机转子。

在使用过程中,应经常检查离心管尺寸是否合适,是否有损坏,以免出现松动或飞出的现象。

维修案例三:客户反馈某离心机在旋转时出现不平衡的情况,导致离心机晃动或者产生噪音,影响了实验的准确性。

经过检查,发现离心机转子的安装不正确,或者离心管重量分布不均匀。

解决方法是重新正确安装转子,或者重新分配样品以保证均衡分布,从而达到稳定运行的效果。

在安装过程中,一定要注意将离心管对称放置,避免不均衡的情况发生。

离心机的日常维护:1.保持离心机清洁卫生,离心结束后,应及时擦拭离心机内外表面。

2.检查离心机各部位是否存在损坏或松动,如有损坏或松动应及时修理。

3.离心机使用后,要及时关机并拔掉电源线,以避免意外发生。

4.定期检查和维护离心机,如清洗离心机转子和离心管。

定期检查内容主要包括以下几个方面:1、定期检查离心机管套,若有异物、污垢,或橡胶脱落等情况需及时处理;2、定期检查转头,若出现腐蚀斑点、凹槽、细小裂纹,应停用;需定期对转头进行保养,清洗转头时需注意,防止表面氧化层破坏;3、定期检查门盖气动支撑杆和门锁开关是否正常,检查仪器各处的橡胶密封圈,若出现变形或老化现象需及时更换;4、定期检查散热风扇是否能正常工作;5、定期清灰,清除离心机通风窗和冷凝器上的灰尘,保证通风和冷凝器的效果;6、定期检查整流子和电刷的磨损情况,有磨损过度的需立即更换;7、电动机的轴承应定期加注润滑油。

高速离心机故障排除常见问题解答说明书

高速离心机故障排除常见问题解答说明书

高速离心机故障排除常见问题解答说明书一、故障现象与处理方法1. 故障现象:离心机运行时发出异常噪音。

处理方法:首先停机检查,排除外部异物的可能性。

若外部无异常,进一步检查离心机内部零部件是否松动,进行适当的紧固和润滑处理。

2. 故障现象:离心机无法正常启动。

处理方法:检查电源是否正常供电,确保电路连接无异常。

若电路无问题,检查电机是否损坏,需要更换。

3. 故障现象:离心机转速异常波动。

处理方法:首先检查是否有负载过重或不均匀的情况,调整负载分配。

如果问题仍然存在,检查离心机内部齿轮传动装置是否损坏或部分磨损,需要修复或更换。

4. 故障现象:离心机加速度不稳定。

处理方法:检查控制系统是否正常工作,对加速度传感器和监控设备进行检修。

若故障依然存在,需要检查电源是否稳定,若电源质量较差,需更换稳定电源供应。

二、常见问题的原因分析与解决方法1. 问题:离心机运转不平稳。

原因:可能是离心机基础不稳固,导致振动。

解决方法:加强离心机基础支撑,确保其稳定性。

2. 问题:离心机运行过热。

原因:可能是轴承缺乏润滑或润滑不足,导致摩擦加热。

解决方法:定期检查轴承润滑状况,根据需要进行润滑补充或更换。

3. 问题:离心机无法达到预设转速。

原因:可能是电机损坏或控制系统故障。

解决方法:更换损坏的电机或修复控制系统故障,确保电机能够可靠运行并达到所需转速。

4. 问题:离心机加速时间过长。

原因:可能是供电电源电压过低或控制系统设置有误。

解决方法:调整电源电压至合适范围内,检查控制系统设置是否正确。

5. 问题:离心机离心力不均匀。

原因:可能是离心机转子不平衡或加速度传感器故障。

解决方法:进行动平衡处理或更换失效的加速度传感器,确保离心力均匀分布。

三、故障预防与维护建议1. 定期检查离心机各部件的紧固程度,确保其稳定性。

2. 定期清洁离心机内部和外部,防止灰尘和污垢对机器造成损害。

3. 定期检查润滑油的使用情况,及时进行补充或更换。