POCT血糖仪与生化分析仪血糖检测结果的比对试验及分析刘加红
【期刊名称】《中国现代药物应用》
【年(卷),期】2014(000)014
【摘要】Objective To explore precision and accuracy of point-of-care testing(POCT) blood glucose meter through comparing blood glucose results between POCT glucose meter and biochemical analyzer. Methods Selected 4 sets of POCT glucose meter to detect the same sample and compared blood glucose results between Beckman CX-9 biochemical analyzer and POCT blood glucose meter. Results Measured results of Beckman CX-9 biochemical analyzer and POCT glucose meter were highly correlated(P>0.05), differences were not statistically significant. Conclusion POCT blood glucose meter is a consistent and convenient detection method for measuring blood glucose results as biochemical analyzer, so it is worthy of clinical promotion and application.%目的对比床旁检测(POCT)血糖仪与生化分析仪对血糖测定的结果,对POCT血糖仪的精密度与准确度进行探讨。
方法选用4台POCT血糖仪对同一标本进行测定,并比较贝克曼CX-9生化分析仪与POCT血糖仪的测定结果。
结果贝克曼CX-9生化分析仪与POCT血糖仪的测定结果高度相关(P>0.05),差异不具有统计学意义。
结论 POCT血糖仪与生化分析仪对血糖的测定结果具有一致性,且方便快捷,可作为急诊的检测方法,值得在临床过程中推广、应用。
【总页数】2页(P21-22)
【作者】刘加红
【作者单位】225300 江苏省泰州市中西医结合医院检验科
【正文语种】中文
【相关文献】
1.POCT血糖仪与全自动生化分析仪血糖检测结果的比对分析 [J], 羊建;周明莉
2.POCT血糖仪与生化分析仪血糖检测结果的比对试验及分析 [J], 邓芳;敖家富;王海涛
3.POCT血糖仪与全自动生化分析仪血糖检测结果的比对 [J], 赵志芳;张炳峰;陈晓婷;赵建群
4.POCT血糖仪与全自动生化分析仪血糖检测结果比对 [J], 刘云峰;刘中国;杜利容;邓清华
5.POCT血糖仪与全自动生化分析仪血糖检测结果的比对分析 [J], 景鸣;饶煌辉;夏卫珍