AB液质联用使用说明tune mass calibration and CEM
ABSciex液质联用使用说明MSandMS-MSexperiments
ABSciex液质联⽤使⽤说明MSandMS-MSexperiments2.1 IntroductionThis chapter is intended to familiarize the user with the different scan typesavailable on the instrument.In this section, users will:Set up a Q1 MS scan experimentSet up a Product Ion Scan experiment – see differences with and without CE rampSet up a Precursor Ion Scan experimentSet up a Neutral Loss Scan experimentExplain the differences and advantages of the various MS/MSexperiments2.2 MaterialsIn this chapter, a sulfa-drug mix is used as the example, however, you can use any compound or mixture of compounds to complete this chapter.Sulfa-drug mix standards at appropriate concentration for instrument model (see Appendix for preparation instructions and suggestedconcentrations)Syringe PumpSyringePEEK tubing and union adapter2.3 ProcedureInsert the syringe into the PEEK tubing and union adapterTighten fittings to ensure there will be no leaksPlace syringe on pump and lock in placeSlide push plate up against plunger of syringeSet flow rate and diameter of syringePlumb the loose end of the PEEK tubing to the inlet of the sourcePress start to begin infusion2.4 MS/MS ExperimentsDO NOT continue to work in the "API Instrument" Project folder. It is used only during Mass CalibrationMS and MS/MS Experiments2.4.1 Create a Project FolderClick on Tools, Project, Create Project and give the project a relevant name, suchas Training. If desired, create a subproject folder by clicking the “Add All” buttonto add the required folders to the subproject folder. The name defaults as thedate, but you can change the name of the subproject folder after the requiredfolders have been added.2.4.2 Syringe Pump Set-up1.Load the syringe with the sulfa-drug mix 4 Compound Mix at anappropriate concentration for your instrument model (see Appendix,Table A-4).2.Start the infusion at a flow rate of 10 µL/min. Ensure that the appropriatesyringe diameter (Chapter 1, Table1-2) is being used.API 2000? and API 3200? Systems: Use the drop down menu to theright of Edit Ramp to access the Syringe Pump Method dialog box.Click the Start Syringe Pump to start the infusion.3.Allow enough time for the infusion rate to equilibrate.4.If needed, verify that the sprayer is positioned properly.API 2000?, API 3000? Systems: the edge of the spray is at least 2mm from the curtain plate hole. Momentarily increase flow rate to 100µL/min to visibly see the spray for proper spray adjustment relative tothe orifice.API 3200?, API 4000?, & API 5000? Systems: Settings of 10vertical, 5 horizontal are recommended.2.4.3 Q1 MS scan1.In the Navigation Bar, click on Tune and Calibrate Mode. If open, closethe Tune Method Editor window.2.Double-click on Manual Tuning to re-open the Tune Method Editorwindow. You should hear the gases turn on as the instrument becomesactive for use (depending on previous status of the instrument, gases maybe on already).3.In Scan Type, select Q1 MS scan.4.Enter a Start (amu) of 200, Stop (amu) of 400 and Time (sec) of 3. Thesevalues are sufficient for the 4 Compound Mix provided.5.Set number of Cycles to 10 and check MCA./doc/59e78a245a8102d276a22f9d.html e the default values for the Source/Gas tab for now.7.At flows of < 50 µL/min, GS2 /TurboIonSpray? Gas and the temperature(TEM) is kept at 0.8.On the Resolution Tab, select Unit. Do not change the IE1./doc/59e78a245a8102d276a22f9d.html pound tab:API 2000?, API 3000? Systems: set the DP to 50; use default forother settings.API 3200?, API 4000?, & API 5000? Systems: set the DP to 80; usedefault for other settings.MS and MS/MS Experiments10.Click Start to acquire 10 scans in MCA (Multiple Channel Acquisition) mode.11.After the data has been acquired, select the bottom right pane, then right click and select Open File to view a window similar to the one below. Notice the presence of sodium adducts (M+H+22).Figure 2-1Q1 MS scan of 4 Compound Mix12.Right Click and select Add Caption. Type in Q1 MS scan. Click OK.13.Print Pane (Ctrl-P).14.Optional: Repeat the experiment at a much lower DP (20V) and at a high DP (140V). Note changes in spectrum.15.Close the windowTip!To break up any adducts/clusters, you can increase the DP . The DP is applied to the orifice plate and therefore will remove the sodium before the ion enters Q0.Table 2-14 Compound Mix (see Appendix for structures)Name MW (amu)Observed M+H Sulfadiazine250.3251Sulfamerazine264.3265Sulfamethazine278.3279Sulfadimethoxine 310.3311MS and MS/MS Experiments2.4.4 Product Ion ScanManually Adjusting CEA Product Ion Scan is an MS/MS scan where the first quadrupole (Q1) is fixed and the third quadrupole (Q3) sweeps a range. It is an experiment that will search for all of the products of a particular precursor ion.1.Under Scan Type, select Product Ion Scan.2.In the Products Of field, enter one of the precursor ions observed in the Q1 MS scan above (see printed results from 3-3).3.Enter a Start (amu) of 40, Stop (amu) of 10 mass units above the selected precursor/parent ion and Time (sec) of 3.4.Verify that the MCA box is unchecked and enter 30 cycles.5.Source/Gas tab: Set CAD gas to 4/doc/59e78a245a8102d276a22f9d.html pound tab: For DP , same as in Q1 MS scan (3-3). Enter 20 for CE.7.Resolution tab: Q1 resolution = LOW. Q3 resolution = UNIT.These settings will increase sensitivity (low resolution) of the parent or precursor ion that is transmitted in Q1 but at the same time allow for normal isotope product ion pattern (unit resolution) in Q3.8.Click Start to begin the experiment.9.After scanning has started, increase the CE 5V at a time until 80V and observe how the fragmentation pattern shifts from high mass fragments to low mass fragments. The spectrum displayed will correlate to the CE entered.Using Software to Ramp CETo obtain a “representative” spectrum of fragment ions, the collision energy (CE) is ramped incrementally during the experiment. Again, as observed in the experiment above, at low CE values, high MW product ions are formed and most of the precursor ion is kept intact whereas increasing the CE to higher values, low MW product ions are formed.The spectra produced in the experiment outlined below will show us a cumulative “picture” of the various fragment ions that are formed from a low CE to a high CE (i.e. CE being ramped automatically with MCA on). Procedures for finding the optimal CE for each fragment will be discussed in Chapters 4 and 5.1.Under Scan Type, select Product Ion Scan.2.In the Products Of field, enter one of the precursor ions observed in the Q1 MS scan above (see printed results from 3-3).3.Enter a Start (amu) of 40, Stop (amu) of 10 mass units above the selected precursor/parent ion and Time (sec) of 3.4.Source/Gas tab: Set CAD gas to 4Note:Remember that CE is a voltage difference between Q0 and the collision cell RO2 (CE= Q0-RO2). The higher the CE value, the faster the ions are drawn into the collision cell containing nitrogen gas ions (CAD gas), hence, the smaller fragments are formed. The lower CE value keeps the precursor ion intact.MS and MS/MS Experiments/doc/59e78a245a8102d276a22f9d.html pound tab: Same as in the Q1 MS scan above (3-3).6.Resolution tab: Q1 resolution = LOW. Q3 resolution = UNIT.Again, these settings will increase sensitivity (low resolution) of the parent or precursor ion that is transmitted in Q1 but at the same time allow for normal isotope product ion pattern (unit resolution) in Q3.7.Click on the Edit Ramp button to open the Ramp Parameter Settings window.Figure 2-2Ramp Parameter Settings Window8.Select Collision Energy. Use the default Start and Stop vaules, but increase the Step to 5 volts.9.Click OK and note that the Ramp Parameter box is checked.10.Be sure the MCA box is checked (MCA Multi-Channel Averaging will add the spectra generated at each step of the CE ramp). When Ramp Parameter is active, the Period Cycles and Duration in the MS tab of the Tune Method Editor window are automatically adjusted to reflect the ramp steps.11.Click Start to begin the experiment.12.After the data has been acquired, select the bottom right pane, then right click and select Open File to view a window similar to the one in Figure 2-3.13.When the scan has finished, right click and select Add Caption.14.Type in “Product Ion Scan of “X" with CE Param Ramp”. Click OK.15.Print Pane (ctrl-P). Close the window.Note:MCA Multi-Channel Averaging will add the spectra generated at each step of the CE ramp – the optimal CE for a particular fragment cannot be obtained in the experiment outlined above. This will be performed/discussed in Chapters 4 and 5. The experiment above is meant to provide information on thevarious fragments that can be formed over a wide range of CE.MS and MS/MS ExperimentsFigure 2-3Product Ion Scan of Sulfadimethoxine with CE Ramp16.Repeat steps 2 to 16 using another precursor ion.The Product Ion Scan Spectra generated with CE being ramped andMCA on results in a “summed” type of spectra i.e. Product IonSpectra at each CE step is added to provide a “picture” of the variousfragments/product ions over a wide CE range.Remember, different fragments are formed depending on the collisionenergy.In the spectrum shown above, the parent ion was formed at low CEand the low mass fragments were formed at high CE. Since thespectrum at each CE ramp step were added, the resulting spectrumthen displays the range of fragments formed from low to high CE.The optimal CE of any particular fragment of interest can bedetermined in MRM mode. This will be discussed further in Chapters 4and 5.2.4.5 Precursor Ion ScanPrecursor Ion Scan is an MS/MS scan where the third quadrupole (Q3) is fixed andthe first quadrupole (Q1) sweeps a range. It is a mode to scan for the ion of aspecific mass-to-charge ratio (precursor) that is generating a defined product ionformed by fragmentation.From the print outs of the Product Ion Scan Spectra with CE ramped of all 4compounds from 3-4, notice that 92 amu is a strong common fragment. Thiscommon fragment can be used in the Precursor Ion Scan experiment describedbelow.1.Under Scan Type, select Precursor Ion Scan.2.In the Precursor Of field, enter the common fragment ion 92.MS and MS/MS Experiments3.Enter a Start (amu) of 200, Stop (amu) of 400 and Time (sec) of 3.4.In the MS tab, set number of Cycles to 20 and check MCA.5.Source/Gas tab: Set CAD gas to 4/doc/59e78a245a8102d276a22f9d.html pound tab: For DP , same as in Q1 MS scan (3-3). Enter 35 for CE.7.Resolution tab: Q1 resolution = UNIT. Q3 resolution = LOW. These settings allow for a normal isotope precursor ion pattern (unit resolution) but at the same time will increase sensitivity (low resolution) of the fragment ion.8.The Ramp Parameter box should be unchecked.9.Click Start.10.Select the bottom right pane, then right click and select Open File to view the Precursor Ion Scan, similar to the one in Figure 2-4.11.Right Click and Add Caption after 20 MCA scans (20 cycles) has been acquired. Type in “Precursor of 92.”12.Print Pane (ctrl-P) and close the window.Figure 2-4Precursor Ion Scan of Fragment 92 amu13.Set up another Precursor Ion Scan experiment. This time, choose a precursor ion that is not common among the 4 compounds. Check the Product Ion Scan with CE ramp printouts from 3-4 for unique fragments for each compound.Note:The Precursor Ion Scan experiment will display a spectrum of all the parent ions that produced the selected fragment. This experiment is very powerful for finding metabolites with a common structural moiety, especially in a complex mixture or matrix. Note also the spectral difference from the Q1 MSscan printout.MS and MS/MS Experiments2.4.6 Neutral Loss ScanNeutral Loss Scan is a mode of operating an MS/MS instrument so that the massof the two analyzers (Q1 and Q3) differs by a defined constant value (neutral loss).A response will be observed if the ion selected by the first analyzer fragments bylosing or gaining the mass difference.A Neutral Loss Scan is used to look for charged fragments that are produced byloss of a neutral fragment. In the case of the 4 compound mixture used in theabove experiments, 155 is a common neutral fragment produced. This type ofMS/MS experiment typically complements Precursor Ion Scan type experimentsand is especially useful in qualitative metabolism analysis.1.Under Scan Type, select Neutral Loss Scan.2.In the Loss Of field, enter 155.3.Enter an appropriate Start (amu), Stop (amu) and Time (sec). A mass rangeof 200 - 400 amu is sufficient for the compounds in the mixture.4.In the MS tab, set number of Cycles to 20 and check MCA.5.Source/Gas tab: Set CAD gas to 4/doc/59e78a245a8102d276a22f9d.html pound tab: For DP, same as in Q1 MS scan (3-3). Enter 35 for CE.7.Resolution tab: Q1 resolution = UNIT. Q3 resolution = UNIT.8.The Ramp Parameter box should be unchecked.9.Click Start.10.Select the bottom right pane, then right click and select Open File to viewthe Neutral Loss Scan, similar to the one in Figure2-5.11.Right Click and Add Caption.12.Print Pane (ctrl-P) and close the window.Figure 2-5Neutral Loss Scan of Fragment 155MS and MS/MS Experiments13.Set up another Neutral Loss Scan experiment. This time, choose a neutral ion that is not common to all the 4 compounds.Check the Product Ion Scan printouts from 3-4 for unique fragments. To determine a neutral fragment, subtract the positive ion fragment from the precursor ion. The Product Ion Scan with CE ramp printouts show these ions. This also allows hands-on opportunity for other members of the group.2.4.7 MS and MS/MS Scan Types 1.Click Tune and Calibrate Mode in the navigation bar and close all windows.2.Double click on Manual Tuning.3.In the MS tab, click on the Scan Type drop down menu. Identify which scans are MS versus MS/MS scans4.Click Source/Gas tab. In MS tab, select Q1 MS scan from the drop down menu. Then select Q3 MS scan. Write down the parameters (not values) that appear/disappear. Select an MS/MS scan. Again note appearance of other parameters.5.Click Compound tab. Repeat steps 3 and 4.6.In terms of available parameters to adjust:Is a Q1 MS scan different from Q3 MS scan? Why? Are the MS/MS scans different? Why?2.5 Discuss and Answer the following Questions:1.Why is the temperature (TEM) and GS2 kept at 0 at flow rates <50µL/min?2.How do I change the default values on the Source/Gas and Compound tabs that are displayed with each scan mode each time I open Manual Tuning?3.I have calibrated the instrument at Unit resolution. Now I’m ready to analyze my compound. Can I change the IE to improve the resolution and peak shape of my compound? Why?Note:The spectrum in Figure 2-5 shows compounds that produced the neutral fragment of 155 amu. Note that the mass spec is not capable of detecting neutral fragments. The ions that generated the 155 neutral fragment (and consequently detected by the mass spec) are listed in Table 2-2.Table 2-2Common Neutral Loss Fragment CompoundPrecursor Ion - Neutral Charged Product Ion Sulfadiazine251(+) - 155 =96(+)Sulfamerazine265(+) - 155 =110(+)Sulfamethazine 279(+) - 155 =124(+)Sulfadimethoxine 311(+) - 155 =156(+)MS and MS/MS Experiments4.Why do the spectra generated by Precursor Ion and Neutral Loss Scansseem ”cleaner” or have less background than Q1 MS scan?5.Explain why the following resolution settings were used in the MS/MSexperiments above:Product Ion Scan: Q1 Low, Q3 UnitPrecursor Ion Scan: Q1 Unit, Q3 LowNeutral Loss Scan: Q1 Unit, Q3 UnitAt the end of this chapter, you should have the following printouts:Q1 MS scan from 3-3Product Ion Scans for each compound with CE ramp from 3-4Precursor Ion Scan from 3-5Neutral Loss Scan from 3-6。
AB液质操作规程完整
第一部分开关机A A、开机1打开UPS电源,打开氮气发生器(气瓶),确认Curtain gas的输入压力为0.4MPa, gas1/2为0.7MPa,exhuast为0.4MPa。
2. 打开机械泵电源开关(5500系列直接打开质谱主机电源,仪器自动启动机械泵)3. 等机械泵工作至少30min5. 打开质谱主机电源开关6. 过夜抽真空B B、关机1•停止输液:关停针泵或液相泵,或断开输液管线(一般质谱主机Sta ndby后液相系统也自动待机,输液泵自动关停,但建议操作人员再次确认,必须保证没有任何液体再泵入质谱)2•使仪器待机,并灭活配置2•关闭质谱仪主机电源开关(5500系列只需按下电源开关旁的VENT按键排放真空)3•机械泵继续工作至少30min4•关闭机械泵上的电源开关(5500系列会自动关停机械泵,待机械泵停止后即可关闭5500系列主机上的电源开关)5•关闭气体发生器或气瓶第二部分调谐1、PPG就是正离子校正液,PPG3000负离子调谐液,本台仪器只用两瓶校正液,正离子一瓶,负离子一瓶。
洗液一瓶50%的甲醇水。
调谐前先用洗液洗针两次。
吸校正液的时候慢慢吸,避免吸入气泡,吸好以后将针泵卡住。
再将高度调到5,将管路重新连接。
不再连接六通阀,由质谱直接进样,且将针泵的速度调到10卩L/m。
2、调谐液弄好后在仪器面板工具栏上:TOO H项目一创建项目。
此项目为一个文件夹,文件夹里面包括所有此次采集的信息,方法、序列等,且所有采集的项目都固定在一个文件夹下面:D:/A nalyst Data/Project。
新建项目的时候,弹出窗口中有个Add All,可以创建孙文件夹。
3、联机:软件打开后在左上方硬件配置(单机选中,双击打开)-只需选中仪器MB active profile。
图标变为绿色,表示联机成功。
调谐只需要选中质谱。
仪器的实时状态可以点击右下方的图标,质谱和液相都是单独观察,质谱要注意真空度,为0.5.4、调谐—手动调谐—项目(文件夹)选中下拉菜单中的Installation 20161102(为系统默认的调谐文件夹),先Q仔start syring pump (打开针泵)—start (质谱),开始采集以后观察基线是否平衡,若等两分钟还未平衡好,则手动将针泵往上抬。
AB Sciex液质联用使用说明MS and MS-MS experiments
2.1 IntroductionThis chapter is intended to familiarize the user with the different scan typesavailable on the instrument.In this section, users will:•Set up a Q1 MS scan experiment•Set up a Product Ion Scan experiment – see differences with and without CE ramp•Set up a Precursor Ion Scan experiment•Set up a Neutral Loss Scan experiment•Explain the differences and advantages of the various MS/MSexperiments2.2 MaterialsIn this chapter, a sulfa-drug mix is used as the example, however, you can use any compound or mixture of compounds to complete this chapter.•Sulfa-drug mix standards at appropriate concentration for instrument model (see Appendix for preparation instructions and suggestedconcentrations)•Syringe Pump•Syringe•PEEK tubing and union adapter2.3 Procedure•Insert the syringe into the PEEK tubing and union adapter•Tighten fittings to ensure there will be no leaks•Place syringe on pump and lock in place•Slide push plate up against plunger of syringe•Set flow rate and diameter of syringe•Plumb the loose end of the PEEK tubing to the inlet of the source•Press start to begin infusion2.4 MS/MS ExperimentsDO NOT continue to work in the "API Instrument" Project folder. It is used only during Mass CalibrationMS and MS/MS Experiments2.4.1 Create a Project FolderClick on Tools, Project, Create Project and give the project a relevant name, suchas Training. If desired, create a subproject folder by clicking the “Add All” buttonto add the required folders to the subproject folder. The name defaults as thedate, but you can change the name of the subproject folder after the requiredfolders have been added.2.4.2 Syringe Pump Set-up1.Load the syringe with the sulfa-drug mix 4 Compound Mix at anappropriate concentration for your instrument model (see Appendix,Table A-4).2.Start the infusion at a flow rate of 10 μL/min. Ensure that the appropriatesyringe diameter (Chapter 1, Table1-2) is being used.API 2000™ and API 3200™ Systems: Use the drop down menu to theright of Edit Ramp to access the Syringe Pump Method dialog box.Click the Start Syringe Pump to start the infusion.3.Allow enough time for the infusion rate to equilibrate.4.If needed, verify that the sprayer is positioned properly.API 2000™, API 3000™ Systems: the edge of the spray is at least 2mm from the curtain plate hole. Momentarily increase flow rate to 100μL/min to visibly see the spray for proper spray adjustment relative tothe orifice.API 3200™, API 4000™, & API 5000™ Systems: Settings of 10vertical, 5 horizontal are recommended.2.4.3 Q1 MS scan1.In the Navigation Bar, click on Tune and Calibrate Mode. If open, closethe Tune Method Editor window.2.Double-click on Manual Tuning to re-open the Tune Method Editorwindow. You should hear the gases turn on as the instrument becomesactive for use (depending on previous status of the instrument, gases maybe on already).3.In Scan Type, select Q1 MS scan.4.Enter a Start (amu) of 200, Stop (amu) of 400 and Time (sec) of 3. Thesevalues are sufficient for the 4 Compound Mix provided.5.Set number of Cycles to 10 and check MCA.e the default values for the Source/Gas tab for now.7.At flows of < 50 μL/min, GS2 /TurboIonSpray® Gas and the temperature(TEM) is kept at 0.8.On the Resolution Tab, select Unit. Do not change the IE1.pound tab:API 2000™, API 3000™ Systems: set the DP to 50; use default forother settings.API 3200™, API 4000™, & API 5000™ Systems: set the DP to 80; usedefault for other settings.MS and MS/MS Experiments10.Click Start to acquire 10 scans in MCA (Multiple Channel Acquisition) mode.11.After the data has been acquired, select the bottom right pane, then right click and select Open File to view a window similar to the one below. Notice the presence of sodium adducts (M+H+22).Figure 2-1Q1 MS scan of 4 Compound Mix12.Right Click and select Add Caption. Type in Q1 MS scan. Click OK.13.Print Pane (Ctrl-P).14.Optional: Repeat the experiment at a much lower DP (20V) and at a high DP (140V). Note changes in spectrum.15.Close the windowTip!To break up any adducts/clusters, you can increase the DP . The DP is applied to the orifice plate and therefore will remove the sodium before the ion enters Q0.Table 2-14 Compound Mix (see Appendix for structures)Name MW (amu)Observed M+H Sulfadiazine250.3251Sulfamerazine264.3265Sulfamethazine278.3279Sulfadimethoxine 310.3311MS and MS/MS Experiments2.4.4 Product Ion ScanManually Adjusting CEA Product Ion Scan is an MS/MS scan where the first quadrupole (Q1) is fixed and the third quadrupole (Q3) sweeps a range. It is an experiment that will search for all of the products of a particular precursor ion.1.Under Scan Type, select Product Ion Scan.2.In the Products Of field, enter one of the precursor ions observed in the Q1 MS scan above (see printed results from 3-3).3.Enter a Start (amu) of 40, Stop (amu) of 10 mass units above the selected precursor/parent ion and Time (sec) of 3.4.Verify that the MCA box is unchecked and enter 30 cycles.5.Source/Gas tab: Set CAD gas to 4pound tab: For DP , same as in Q1 MS scan (3-3). Enter 20 for CE.7.Resolution tab: Q1 resolution = LOW. Q3 resolution = UNIT.These settings will increase sensitivity (low resolution) of the parent or precursor ion that is transmitted in Q1 but at the same time allow for normal isotope product ion pattern (unit resolution) in Q3.8.Click Start to begin the experiment.9.After scanning has started, increase the CE 5V at a time until 80V and observe how the fragmentation pattern shifts from high mass fragments to low mass fragments. The spectrum displayed will correlate to the CE entered.Using Software to Ramp CETo obtain a “representative” spectrum of fragment ions, the collision energy (CE) is ramped incrementally during the experiment. Again, as observed in the experiment above, at low CE values, high MW product ions are formed and most of the precursor ion is kept intact whereas increasing the CE to higher values, low MW product ions are formed.The spectra produced in the experiment outlined below will show us a cumulative “picture” of the various fragment ions that are formed from a low CE to a high CE (i.e. CE being ramped automatically with MCA on). Procedures for finding the optimal CE for each fragment will be discussed in Chapters 4 and 5.1.Under Scan Type, select Product Ion Scan.2.In the Products Of field, enter one of the precursor ions observed in the Q1 MS scan above (see printed results from 3-3).3.Enter a Start (amu) of 40, Stop (amu) of 10 mass units above the selected precursor/parent ion and Time (sec) of 3.4.Source/Gas tab: Set CAD gas to 4Note:Remember that CE is a voltage difference between Q0 and the collision cell RO2 (CE= Q0-RO2). The higher the CE value, the faster the ions are drawn into the collision cell containing nitrogen gas ions (CAD gas), hence, the smaller fragments are formed. The lower CE value keeps the precursor ion intact.MS and MS/MS Experimentspound tab: Same as in the Q1 MS scan above (3-3).6.Resolution tab: Q1 resolution = LOW. Q3 resolution = UNIT.Again, these settings will increase sensitivity (low resolution) of the parent or precursor ion that is transmitted in Q1 but at the same time allow for normal isotope product ion pattern (unit resolution) in Q3.7.Click on the Edit Ramp button to open the Ramp Parameter Settings window.Figure 2-2Ramp Parameter Settings Window8.Select Collision Energy. Use the default Start and Stop vaules, but increase the Step to 5 volts.9.Click OK and note that the Ramp Parameter box is checked.10.Be sure the MCA box is checked (MCA Multi-Channel Averaging will add the spectra generated at each step of the CE ramp). When Ramp Parameter is active, the Period Cycles and Duration in the MS tab of the Tune Method Editor window are automatically adjusted to reflect the ramp steps.11.Click Start to begin the experiment.12.After the data has been acquired, select the bottom right pane, then right click and select Open File to view a window similar to the one in Figure 2-3.13.When the scan has finished, right click and select Add Caption.14.Type in “Product Ion Scan of “X" with CE Param Ramp”. Click OK.15.Print Pane (ctrl-P). Close the window.Note:MCA Multi-Channel Averaging will add the spectra generated at each step of the CE ramp – the optimal CE for a particular fragment cannot be obtained in the experiment outlined above. This will be performed/discussed in Chapters 4 and 5. The experiment above is meant to provide information on thevarious fragments that can be formed over a wide range of CE.MS and MS/MS ExperimentsFigure 2-3Product Ion Scan of Sulfadimethoxine with CE Ramp16.Repeat steps 2 to 16 using another precursor ion.The Product Ion Scan Spectra generated with CE being ramped andMCA on results in a “summed” type of spectra i.e. Product IonSpectra at each CE step is added to provide a “picture” of the variousfragments/product ions over a wide CE range.Remember, different fragments are formed depending on the collisionenergy.In the spectrum shown above, the parent ion was formed at low CEand the low mass fragments were formed at high CE. Since thespectrum at each CE ramp step were added, the resulting spectrumthen displays the range of fragments formed from low to high CE.The optimal CE of any particular fragment of interest can bedetermined in MRM mode. This will be discussed further in Chapters 4and 5.2.4.5 Precursor Ion ScanPrecursor Ion Scan is an MS/MS scan where the third quadrupole (Q3) is fixed andthe first quadrupole (Q1) sweeps a range. It is a mode to scan for the ion of aspecific mass-to-charge ratio (precursor) that is generating a defined product ionformed by fragmentation.From the print outs of the Product Ion Scan Spectra with CE ramped of all 4compounds from 3-4, notice that 92 amu is a strong common fragment. Thiscommon fragment can be used in the Precursor Ion Scan experiment describedbelow.1.Under Scan Type, select Precursor Ion Scan.2.In the Precursor Of field, enter the common fragment ion 92.MS and MS/MS Experiments3.Enter a Start (amu) of 200, Stop (amu) of 400 and Time (sec) of 3.4.In the MS tab, set number of Cycles to 20 and check MCA.5.Source/Gas tab: Set CAD gas to 4pound tab: For DP , same as in Q1 MS scan (3-3). Enter 35 for CE.7.Resolution tab: Q1 resolution = UNIT. Q3 resolution = LOW. These settings allow for a normal isotope precursor ion pattern (unit resolution) but at the same time will increase sensitivity (low resolution) of the fragment ion.8.The Ramp Parameter box should be unchecked.9.Click Start.10.Select the bottom right pane, then right click and select Open File to view the Precursor Ion Scan, similar to the one in Figure 2-4.11.Right Click and Add Caption after 20 MCA scans (20 cycles) has been acquired. Type in “Precursor of 92.”12.Print Pane (ctrl-P) and close the window.Figure 2-4Precursor Ion Scan of Fragment 92 amu13.Set up another Precursor Ion Scan experiment. This time, choose a precursor ion that is not common among the 4 compounds. Check the Product Ion Scan with CE ramp printouts from 3-4 for unique fragments for each compound.Note:The Precursor Ion Scan experiment will display a spectrum of all the parent ions that produced the selected fragment. This experiment is very powerful for finding metabolites with a common structural moiety, especially in a complex mixture or matrix. Note also the spectral difference from the Q1 MSscan printout.MS and MS/MS Experiments2.4.6 Neutral Loss ScanNeutral Loss Scan is a mode of operating an MS/MS instrument so that the massof the two analyzers (Q1 and Q3) differs by a defined constant value (neutral loss).A response will be observed if the ion selected by the first analyzer fragments bylosing or gaining the mass difference.A Neutral Loss Scan is used to look for charged fragments that are produced byloss of a neutral fragment. In the case of the 4 compound mixture used in theabove experiments, 155 is a common neutral fragment produced. This type ofMS/MS experiment typically complements Precursor Ion Scan type experimentsand is especially useful in qualitative metabolism analysis.1.Under Scan Type, select Neutral Loss Scan.2.In the Loss Of field, enter 155.3.Enter an appropriate Start (amu), Stop (amu) and Time (sec). A mass rangeof 200 - 400 amu is sufficient for the compounds in the mixture.4.In the MS tab, set number of Cycles to 20 and check MCA.5.Source/Gas tab: Set CAD gas to 4pound tab: For DP, same as in Q1 MS scan (3-3). Enter 35 for CE.7.Resolution tab: Q1 resolution = UNIT. Q3 resolution = UNIT.8.The Ramp Parameter box should be unchecked.9.Click Start.10.Select the bottom right pane, then right click and select Open File to viewthe Neutral Loss Scan, similar to the one in Figure2-5.11.Right Click and Add Caption.12.Print Pane (ctrl-P) and close the window.Figure 2-5Neutral Loss Scan of Fragment 155MS and MS/MS Experiments13.Set up another Neutral Loss Scan experiment. This time, choose a neutral ion that is not common to all the 4 compounds. Check the Product Ion Scan printouts from 3-4 for unique fragments. To determine a neutral fragment, subtract the positive ion fragment from the precursor ion. The Product Ion Scan with CE ramp printouts show these ions. This also allows hands-on opportunity for other members of the group.2.4.7 MS and MS/MS Scan Types1.Click Tune and Calibrate Mode in the navigation bar and close all windows.2.Double click on Manual Tuning.3.In the MS tab, click on the Scan Type drop down menu. Identify which scans are MS versus MS/MS scans4.Click Source/Gas tab. In MS tab, select Q1 MS scan from the drop down menu. Then select Q3 MS scan. Write down the parameters (not values) that appear/disappear. Select an MS/MS scan. Again note appearance of other parameters.5.Click Compound tab. Repeat steps 3 and 4.6.In terms of available parameters to adjust:Is a Q1 MS scan different from Q3 MS scan? Why? Are the MS/MS scans different? Why?2.5 Discuss and Answer the following Questions:1.Why is the temperature (TEM) and GS2 kept at 0 at flow rates <50μL/min?2.How do I change the default values on the Source/Gas and Compound tabs that are displayed with each scan mode each time I open Manual Tuning?3.I have calibrated the instrument at Unit resolution. Now I’m ready to analyze my compound. Can I change the IE to improve the resolution and peak shape of my compound? Why?Note:The spectrum in Figure 2-5 shows compounds that produced the neutral fragment of 155 amu. Note that the mass spec is not capable of detecting neutral fragments. The ions that generated the 155 neutral fragment (and consequently detected by the mass spec) are listed in Table 2-2.Table 2-2Common Neutral Loss Fragment CompoundPrecursor Ion - Neutral Charged Product Ion Sulfadiazine251(+) - 155 =96(+)Sulfamerazine265(+) - 155 =110(+)Sulfamethazine 279(+) - 155 =124(+)Sulfadimethoxine 311(+) - 155 =156(+)MS and MS/MS Experiments4.Why do the spectra generated by Precursor Ion and Neutral Loss Scansseem ”cleaner” or have less background than Q1 MS scan?5.Explain why the following resolution settings were used in the MS/MSexperiments above:Product Ion Scan: Q1 Low, Q3 UnitPrecursor Ion Scan: Q1 Unit, Q3 LowNeutral Loss Scan: Q1 Unit, Q3 UnitAt the end of this chapter, you should have the following printouts:•Q1 MS scan from 3-3•Product Ion Scans for each compound with CE ramp from 3-4•Precursor Ion Scan from 3-5•Neutral Loss Scan from 3-6。
液质联用操作规程完整
1.适用范围本设备配备ACQUITY UPLC液相色谱仪、TQS MS/MS 质谱仪,适用于食品、药品中各种有机物的定性、定量分析,是一种具有高灵敏度的检测仪器,仪器由主机、计算机和数据处理软件等组成。
2. 职责2.1操作人员按照本规程操作仪器,认真填写实验使用记录。
2.2保管人员负责对仪器进行定期维护和保养。
2.3 科室负责人负责监督检查规程的执行。
3.操作程序日常操作步骤:准备UPLC —→设置样品表—→运行样品—→定量—→打印报告。
注:如果一个星期内不运行样品请不要关质谱仪,使其保持真空。
建立新方法和project的操作步骤:准备UPLC —→建立新的project —→用标准品调谐—→编辑质谱方法—→编辑UPLC方法—→设置样品表—→运行样品—→定量—→打印报告。
3.1开机:3.1.1 彻底开机顺序(仪器已关闭)确定MS及其它仪器电源电缆已连接,开氮气发生器、开氩气,小表<0.1mpa。
打开计算机电源 > 等待windows正常启动 >电脑界面右下角网络图标红叉。
打开UPLC自动进样器电源,等到电脑界面右下角网络图标出现感叹号!。
打开UPLC泵电源,等约30s或者是有响声。
打开质谱电源,等待5min,离子源透视镜里面亮。
打开Masslynx软件,masslynx主界面 -----左侧instrument----Mass tune---界面菜单栏vacuum---pump 同样界面左侧偏上diagnostics---vacuum---analyser MS1 turbo speed[%]要在5分钟内升到80。
至少抽真空4个小时 > 查看真空状态主界面mass console---界面左侧xevo tq ms detector加号展开---ms display > 碰撞室真空度 >达到 7.x e-5mbar 。
3.1.2 日常开机顺序(仪器未关闭)开氮气发生器、开氩气,小表<0.1mpa,打开Masslynx软件,进入工作站。
液质联用操作规程
1.适用范围本设备配备ACQUITY UPLC液相色谱仪、TQS MS/MS 质谱仪,适用于食品、药品中各种有机物的定性、定量分析,是一种具有高灵敏度的检测仪器,仪器由主机、计算机和数据处理软件等组成。
2. 职责2.1操作人员按照本规程操作仪器,认真填写实验使用记录。
2.2保管人员负责对仪器进行定期维护和保养。
2.3科室负责人负责监督检查规程的执行。
3.操作程序日常操作步骤:准备UPLC —→设置样品表—→运行样品—→定量—→打印报告。
注:如果一个星期内不运行样品请不要关质谱仪,使其保持真空。
建立新方法和project的操作步骤:准备UPLC —→建立新的project —→用标准品调谐—→编辑质谱方法—→编辑UPLC方法—→设置样品表—→运行样品—→定量—→打印报告。
3.1开机:3.1.1 彻底开机顺序(仪器已关闭)确定MS及其它仪器电源电缆已连接,开氮气发生器、开氩气,小表<0.1mpa。
打开计算机电源> 等待windows正常启动>电脑界面右下角网络图标红叉。
打开UPLC自动进样器电源,等到电脑界面右下角网络图标出现感叹号!。
打开UPLC泵电源,等约30s或者是有响声。
打开质谱电源,等待5min,离子源透视镜里面亮。
打开Masslynx软件,masslynx主界面-----左侧instrument----Mass tune---界面菜单栏vacuum---pump 同样界面左侧偏上diagnostics---vacuum---analyser MS1 turbo speed[%]要在5分钟内升到80。
至少抽真空4个小时> 查看真空状态主界面mass console---界面左侧xevo tq ms detector加号展开---ms display > 碰撞室真空度>达到7.x e-5mbar 。
3.1.2 日常开机顺序(仪器未关闭)开氮气发生器、开氩气,小表<0.1mpa,打开Masslynx软件,进入工作站。
AB液质操作规程
第一部分开关机A A 、开机1.打开UPS电源,打开氮气发生器(气瓶),确认Curtain gas的输入压力为0.4MPa,gas1/2为0.7MPa,exhuast为0.4MPa。
2.打开机械泵电源开关(5500系列直接打开质谱主机电源,仪器自动启动机械泵)3.等机械泵工作至少30min5.打开质谱主机电源开关6.过夜抽真空B B 、关机1.停止输液:关停针泵或液相泵,或断开输液管线(一般质谱主机Standby后液相系统也自动待机,输液泵自动关停,但建议操作人员再次确认,必须保证没有任何液体再泵入质谱)2.使仪器待机,并灭活配置2.关闭质谱仪主机电源开关(5500系列只需按下电源开关旁的VENT按键排放真空)3.机械泵继续工作至少30min4.关闭机械泵上的电源开关(5500系列会自动关停机械泵,待机械泵停止后即可关闭5500系列主机上的电源开关)5.关闭气体发生器或气瓶第二部分调谐1、PPG就是正离子校正液,PPG3000负离子调谐液,本台仪器只用两瓶校正液,正离子一瓶,负离子一瓶。
洗液一瓶50%的甲醇水。
调谐前先用洗液洗针两次。
吸校正液的时候慢慢吸,避免吸入气泡,吸好以后将针泵卡住。
再将高度调到5,将管路重新连接。
不再连接六通阀,由质谱直接进样,且将针泵的速度调到10μL/ml。
2、调谐液弄好后在仪器面板工具栏上:TOOl→项目→创建项目。
此项目为一个文件夹,文件夹里面包括所有此次采集的信息,方法、序列等,且所有采集的项目都固定在一个文件夹下面:D:/Analyst Data/Project。
新建项目的时候,弹出窗口中有个Add All,可以创建文件夹。
3、联机:软件打开后在左上方硬件配置(单机选中,双击打开)→只需选中仪器MS→active profile。
图标变为绿色,表示联机成功。
调谐只需要选中质谱。
仪器的实时状态可以点击右下方的图标,质谱和液相都是单独观察,质谱要注意真空度,为0.5.4、调谐→手动调谐→项目(文件夹)选中下拉菜单中的Installation 20161102(为系统默认的调谐文件夹),先Q1→start syring pump(打开针泵)→start (质谱),开始采集以后观察基线是否平衡,若等两分钟还未平衡好,则手动将针泵往上抬。
AB液质操作规程完整
第一部分开关机A A 、开机1.打开UPS电源,打开氮气发生器(气瓶),确认Curtain gas的输入压力为,gas1/2为,exhuast为。
2.打开机械泵电源开关(5500系列直接打开质谱主机电源,仪器自动启动机械泵)3.等机械泵工作至少30min5.打开质谱主机电源开关6.过夜抽真空B B 、关机1.停止输液:关停针泵或液相泵,或断开输液管线(一般质谱主机Standby后液相系统也自动待机,输液泵自动关停,但建议操作人员再次确认,必须保证没有任何液体再泵入质谱)2.使仪器待机,并灭活配置2.关闭质谱仪主机电源开关(5500系列只需按下电源开关旁的VENT按键排放真空)3.机械泵继续工作至少30min4.关闭机械泵上的电源开关(5500系列会自动关停机械泵,待机械泵停止后即可关闭5500系列主机上的电源开关)5.关闭气体发生器或气瓶第二部分调谐1、PPG就是正离子校正液,PPG3000负离子调谐液,本台仪器只用两瓶校正液,正离子一瓶,负离子一瓶。
洗液一瓶50%的甲醇水。
调谐前先用洗液洗针两次。
吸校正液的时候慢慢吸,避免吸入气泡,吸好以后将针泵卡住。
再将高度调到5,将管路重新连接。
不再连接六通阀,由质谱直接进样,且将针泵的速度调到10μL/ml。
2、调谐液弄好后在仪器面板工具栏上:TOOl→项目→创建项目。
此项目为一个文件夹,文件夹里面包括所有此次采集的信息,方法、序列等,且所有采集的项目都固定在一个文件夹下面:D:/Analyst Data/Project。
新建项目的时候,弹出窗口中有个Add All,可以创建孙文件夹。
3、联机:软件打开后在左上方硬件配置(单机选中,双击打开)→只需选中仪器MS→active profile。
图标变为绿色,表示联机成功。
调谐只需要选中质谱。
仪器的实时状态可以点击右下方的图标,质谱和液相都是单独观察,质谱要注意真空度,为.4、调谐→手动调谐→项目(文件夹)选中下拉菜单中的(为系统默认的调谐文件夹),先Q1→start syring pump(打开针泵)→start(质谱),开始采集以后观察基线是否平衡,若等两分钟还未平衡好,则手动将针泵往上抬。
ABSciex液质联用使用说明quantitate
ABSciex液质联⽤使⽤说明quantitate8.1 IntroductionThis chapter is intended to familiarize new users of the mass spectrometer withthe Quantitate mode of Analyst? software. This tutorial utilizes sample data in the Example Project folder provided. However, the user may use data collectedduring earlier lab exercises.8.2 MaterialsAnalyst softwareAcquired data files8.3 ProcedureSelect inegration algorithmBuild a Results Table using Quantitation WizardBuild a Results Table using Quick QuantView and process the Results TableEdit Table SettingsView and adjust Calibration CurveView Peak Integration & Integration MethodUse Statistics functionBuild/Execute a QueryBuild/Execute a SortBuild/Execute a Metric PlotCreate a Formula Column8.4 Integration Algorithms8.4.1 Analyst Classic and IntelliQuan AlgorithmsThere are two integration algorithms available to Analyst? software users. TheAnalyst Classic algorithm in version 1.4 has all the same features as theintegration algorithm in previous versions of Analyst? software. IntelliQuan is an algorithm new to Analyst? 1.4 software and incorporates some features from the MacQuan software. The user must specify which algorithm to use prior to using the Quantitation Wizard. Because of the improvements incorporated into theIntelliQuan algorithm, it is recommended for best results.Analyst? Software Quantitate8.4.2 Selecting the AlgorithmTo specify the algorithm to be used for integration in the Quantitate mode, click on Tools – Settings – Quantitation Integration Algorithm and choose between Analyst Classic and IntelliQuan. The Quantitation Wizard will automatically use the chosen algorithm. IntelliQuan is the recommended algorithm.Figure 8-1Selecting Integration AlgorithmThe different integration algorithms have different integration parameters. The default parameters dialog box for Analyst Classic is shown in Figure 8-2.Figure 8-2Analyst Classic ParametersThe IntelliQuan algorithm uses one of two Peak-Finding Parameters: Automatic IQA or Specify Parameters-MQII. The Automatic IQA is a “parameterless” setting, while the user can set parameters such as noise percentage in MQII. TheNote:Regardless of which algorithm chosen to best meet the needs of a particular lab, it is advisable to use the same algorithm for integrating subsequent analytical runs within a project. This allows comparable integration parameters for comparing data from different studies in a project. This factor is especially important for those labsworking in a regulate environment.Analyst? Software QuantitateIntelliQuan default integration parameters dialog box is shown in Figure8-3 for both the Automatic-IQA and the Specify Parameters-MQII options.Figure 8-3IntelliQuan Parameters8.4.3 Building a Results Table using Quantitation Wizard1.If not already in the Example project, change to this project. Double Clickon Quantitation Wizard from the navigation bar of Analyst? software. The Select Samples window similar to Figure8-4 appears.Figure 8-4Quantitation Wizard - Sample Selection2.Select the batch of samples to process from the available batch files (QuanData.wiff in the Triple Quad folder or other data). Choose the filesto analyze from the list of data files in the batch and add them to the selected sample list (if using QuanData.wiff, select Add All). Choose the Next option at the bottom of the window.Analyst? Software Quantitate3.The Select Settings & Query window is now open. In this window set the Settings to Use to Default and the Default Query option to None as shown in Figure8-5 (these will be defined later in the tutorial). Choose the Next option at the bottom of the window to move to the next phase in building the method.Figure 8-5Select Settings and Query4.The Select Method window is now open. Three different options are available:Choose Existing Method - selects an existing method from within the current projectCreate Automatic Method - sets the integration parameters from a chosen data file, then automatically integrates chromatograms to generate peak area data. This option is only available for the Analyst Classic and IntelliQuan MQII algorithms.Create New Method - to build a method from a chosen data set5.Choose Create New Method and name the method as Training (Figure8-6). Click Next.Figure 8-6Select MethodAnalyst? Software Quantitate 6.The Select Representative Sample dialog appears. The chosenrepresentative data file will be used to determine retention time of the compounds of interest. Therefore, choose a middle concentration from the data or API3-016 file in QuanData.wiff (Figure8-7). Choose the Next option at the bottom of the window to move on to the next phase.Figure 8-7Select Representative Sample7.The Define Peaks window appears. In the Internal Standards and Analytessections, define the name in the Name box. Choose the corresponding MRM in the Q1/Q3 box. In this example the Internal standard is named, internal std. and has a MRM transition of 420.0/220.0, and the analyte is named analyte with a transition of 400.0/200.0. (Figure8-8). If the Quick Quant option of the Acquisition Batch was used, the names would appear in the drop down menu. If this information was not entered into the batch,a name for each compound can now be defined in this table. Theconcentration of the internal standard will then have to be typed into the final Results Table after the internal standard peaks have been integrated.Figure 8-8Define Peaks For MassSpec8.In the column labeled “Internal Standard”, assign an internal standard tothe analyte(s), if present. In this example, “internal std.” was selectedAnalyst? Software Quantitatefrom the drop down menu.Define the name of the analyte(s) and their MRM or SIM transition(s) in period 1 and assign an internal standard to the analyte(s) if present. In this example the analyte is called analyte and has a MRM transition of 400.0/200.0.9.Define the Default Number of Smooths to be zero. This can be adjusted later if needed. It is always best to see the data in raw form.10.If additional MS experiments and/or periods are present repeat steps 6, 7 and 8 for each and choose the Next option at the bottom of the window to move on when finished.11.The Define Integration window of the first peak, the internal standard in this example, is now open (Figure 8-9). Here the baseline noise is already defined and the peak has been integrated automatically. The default RT Window (retention time window) is 30.0 sec.Figure 8-9Define Integration12.If the desired peak was not found automatically select the background range (Figure 8-10)Note:The names will not appear in drop down menus if Quick Quant was not used before data acquisition. If Quick Quant was not used, enter a name for each compound and internal standard (if present) and select the correspondingQ1/Q3 from the drop down menu in this table.Analyst? Software QuantitateFigure 8-10Select Background13.OR select the peak of interest (Figure8-11) by making a selection in thechromatogram and selecting the corresponding icon in the top left hand corner of this window.Figure 8-11Select Peak14.Select the Advanced button to open up a menu of more advancedintegration parameters. The displayed box will depend on the Algorithm chosen. Set the parameters for IntelliQuan MQII and Analyst Classicparameters as shown in Figures 9-2 and 9-3. For IntelliQuan IQA, there are no parameters to set. Make sure the Concentration units match that of the standard, in this example ng/ml. Select OK to review the new peak integration.15.The peak integration can be reviewed by zooming in on the baseline bydragging the curser outside the x- and y-axes. If the integrationparameters are suitable, click Next to move to the next peak. If they need to be changed, click Back, change the appropriate parameter, and re-examine the integration. The various parameters will be defined at the end of this section.16.Repeat steps 10 through 13 for each peak to be used in the quantitationmethod. In this tutorial the only other peak to be used is that of theanalyte. Make sure that the concentration units match that of theAnalyst? Software Quantitatestandard, ng/ml. After defining the integration parameters for the analyte,choose Next to move on to the final phase of building the quantitationmethod.17.In this final window, Specify Calibration, define the Fit to be Linear, theWeighting to be 1/x and the Regression parameters to be based on Area.Figure 8-12Specify Calibration18.Click Finish to complete the building of the quantitation method. Thesample files originally added in Step 2 are then automatically integratedand a table of results is generated. Save the new Results Table as Training.The next section will navigate through the Results Table.8.4.4 Definition of Integration ParametersIntelliQuan MQIINoise Percentage: The level at which Analyst software considers thesignal background. Used to determine the start and end points forintegration. Higher values will result in a peak being integrated with ahigher baseline, generally resulting in smaller peak areas. However, toohigh a value may result in the failure to integrate very small peaks. Lowervalues will extend the start and stop points of the integration, generallyresulting in higher peak areas. Lower values also increase the chance thata very small peaks will be identified and integrated.Baseline Subtraction Window: A time window (in minutes) around eachdata point used to determine the height of the baseline correction to beapplied to that point. This feature removes large “humps” from thechromatogram. The baseline is defined as the point of minimum intensityon the left side of a given point to the point of minimum intensity on theright side, within the specified window.Peak Splitting Factor: This parameter controls whether a given peakcluster consists of a single peak or multiple adjacent peaks. If the intensitydip is less than the value specified, a single peak is reported. If theintensity dip is not less than the value specified, the cluster will beintegrated as two separate peaks. Setting a large factor will preventclusters from being integrated as more than one peak.Analyst? Software QuantitateSmoothing Width: The number of points used in data smoothing Analyst Classic Bunching Factor: Number of points to be averaged together and considered as a single point for the purpose of peak-finding. Increasing this number extends the start and stop points for the integration, resulting in larger peak areas. Bunching is used to find the peak, but all points are considered for integration.Noise Threshold: The minimum height difference required between two adjacent points for the potential start of a peak to be detected. Used to determine which points are noise and which are part of a peak.Area Threshold: Minimum distance between a series of points to be considered a peak. Used to help identify peaks.Number of Smooths: Smoothing reduces spectral noise. This parameter determines the number of times a chromatogram is smoothed.8.4.5 Building a Results Table using Quick Quant (Optional – Can go directly to Viewing and Processing the Results Table )Quick Quant method created before data acquisition can be used. Use of these methods allows Analyst ? software to define much of the integration method automatically. These parameters can be fine-tuned once the data is in the Results Table. Use of a pre-existing quantitation method follows these general steps as well.1.Click on the Quantitation Wizard and move the appropriate Available Samples to the Selected Samples box. Then Click the Next button.2.Leave the default values in the Query and Settings window; this option will be described later in the tutorial. Click Next, the Select Method window will then appear. Click on Choose Existing Method and select the Quick Quant method created earlier during setup of the samples. After clicking Finish, a Results Table with integrated MRMs, corresponding sample type and concentration information will be prepared. Save the Results Table.8.4.6 Viewing and Processing the Results Table1.Open up the Results Table Training.rdb in the Example project by selecting the Review Results Table option in Quantitate and selecting the previously saved Training.rdb file (Figure 8-13). Maximize the Results Table window by closing the Navigation Bar. Point to the View menu, and remove the check next to Navigation Bar.Note:For much more detailed information on algorithm parameters, refer to the Help button in the Quantitation Wizard window, or from the main Help menu (F1).Analyst? Software QuantitateFigure 8-13The Unedited Results Table2.Click the right mouse button in the table and select Table Settings andthen EditFigure 8-14Editing Table Settings3.The Table Settings dialog appears. Click on Columns, then the Editbutton to the right, and the Results Table Columns dialog appears(Figure8-14). In this dialog, you can define the columns that will appear inthe Results Table. The columns are separated into Field Groups in thedrop down menu at the top of this window: Sample, Analyte, InternalStandard, Record, Formula, and Custom. In this menu, choose Sampleand in the resulting dialog, uncheck Sample ID and File Name. This actionremoves the Sample ID and File Name columns (or Fields) from thedisplayed Results Table.Analyst? Software Quantitate4.Choose Analyte in the menu. In the resulting dialog, uncheck Analyte Peak Height and Standard Query Status. Adjust the Precision of the Analyte Peak Area and Analyte Concentration from 3 to 5 significant figures (Figure 8-15)Figure 8-15Editing Results Table Column Settings5.Choose Internal Standard in the drop down menu and uncheck Peak Height and set the Precision for Peak Area to 5. Click OK to use the new settings then Done to return to the Results Table with the adjusted columns (Figure 8-16).Note:A Custom column must be added in the Acquisition Batch prior to running the sample to be accessible in the Table Settings. Analyst? Software QuantitateFigure 8-16Edited Results Table and Exporting New Settings6.To save these new column settings, right click and choose Table Settingsthen Export to New Table Settings. Enter Training as the name for thistemplate. This saved template can now be accessed whenever a newResults Table is created (in this project only) by using the same right clickmenu and selecting the desired name under the Table Settings options orby selecting it in the Settings to Use drop down menu of the SelectSettings and Query dialog when using the Quantitation Wizard (Figure8-5).8.4.7 Viewing a Calibration LineThere are two options under Quantitate that can be used to view a calibrationline: Calibration – Pane, which opens a new pane for the calibration at thebottom of the window which holds the Results Table; and Calibration – Window,which opens a new window for the calibration over the Results Table. Select theCalibration – Window icon in the Quantitate tool bar to view the calibrationcurve.Analyst? Software QuantitateFigure 8-17Quantitation Mode Tool Bar and Calibration - Window1.From within the calibration window, individual analyte calibration curvescan be viewed (under the Analyte drop down menu); also the regressionplaced on the data can be altered. In this example there is only oneanalyte in the Analyte menu. Change the regression used for the data by selecting the Regression button in the window and viewing theRegression Options dialog (Figure8-18). Under the drop down menus in the Regression Options window, change the Fit to Quadratic and theWeighting to 1/x. Select OK to view the changes.Figure 8-18Regression Options - Fit and Weighting2.The calibration line now uses the new regression settings, which cause theequation for the calibration line and the “r” value to change. To use these new settings, click the Accept button in the window. For this tutorial,select Revert to use the previous settings, then close the calibration curve.Analyst? Software QuantitateFigure 8-19New calibration window using a quadratic regression and 1/x weighting fit8.4.8 Viewing Peak Integration and Updating IntegrationMethods1.The integrated peaks can be viewed in a format much similar to the calibration line, that is, as a window or pane. Both window and pane canbe accessed by Tools – Peak Review or by clicking on the icons. Also you can double-click on a row and the Peak Review Pane will open.2.To view the integrated peaks, select the Peak Review – Pane option under the Tools menu of Analyst? software or click the Peak Review - Pane button on the toolbar to view the peak review pane (Figure8-20). Thepeak review pane will now be opened with the data file API3-012 active (the active data file is the one in bold font).Analyst? Software QuantitateFigure 8-20Peak Review Buttons and Peak Review - Pane3.Right click in the peak review pane and choose Options in the new menu. The Peak Review Options window now appears (Figure8-21).4.In this dialog box, set:Appearance section: Num. Rows 1, Num. Columns 2Automatic Zooming section: Choose Zoom Y axis to 100% of largest peak and Zoom time axis to view peak to a 2.00 min window (Figure8-21). Select OK to use the new settings.Figure 8-21Peak Review Options WindowAnalyst? Software Quantitate5.The peak review pane has now changed to the new settings (Figure8-22).These new settings apply to this Peak Review window only. If a newwindow is opened, the default settings will be used. To change thedefault settings, click Tools – Settings – Quantitation Peak Review Settingsand make changes in this box.Figure 8-22Peak Review Pane With New Settings6.Select the Show or Hide Parameters icon at the top of the peak reviewpane to view the peak integration parameters (Figure8-23).Note:The parameters displayed depend upon the algorithm used. The parameters forIntelliQuan are shown in Figure8-23.Analyst? Software QuantitateFigure 8-23Peak Review Pane Showing Integration Parameters7.The size of peak review pane can be adjusted by dragging the bar to thedesired size. Move the mouse to the top of the review pane. A double arrow should appear. Drag this arrow towards the top of the screen to magnify the peak review pane.8.Change the Bunching Factor to 1 and the Num. Smooths to 1 in theAnalyst Classic integration parameters or Noise Percentage (must select Specify Parameters - MQII) to 90% and Smoothing Width to 5 inIntelliQuan. Select the Apply button to use these new settings. The peak is re-integrated with the new integration parameters. Choose Accept to save the new integration parameters for this sample. Notice that theRecord Modified column in the Results Table is checked for that sample.The other data files still use the original integration method.9.To apply the new integration parameters to all the data files in the ResultsTable, right-click in the Peak Review pane and select Update Method. All the data files will then be re-integrated using the new method and the old integration method will be over written. This method change applies to the selected MRM only.10.A method can also be updated while viewing the Results Table byselecting the Modify Method icon of the Quantitate Tool Bar. Uponselecting this icon, the integration parameters for each compound, the fit for the calibration line and the components information for the data files,i.e. the peak names and corresponding MRM transition for eachcompound can be changed (Figure8-24).Analyst? Software QuantitateFigure 8-24Modify Method Button and Window11.To use the Manual Integration Mode, select the Manual Integration icon inthe peak review pane and use the mouse to drag a line from the base ofone side of the peak of interest to the other. The peak is now manuallyintegrated, the Manual Integration box is checked, and the integrationparameters are greyed out and inactive (Figure8-25). Unchecking theManual Integration box restores the integration parameters. However, therecord is still noted as modified in the Results Table. You can also revert tothe method integration parameters by right-clicking in the Peak Reviewpane and selecting the Revert to Method optionFigure 8-25Manual IntegrationAnalyst? Software Quantitate 12.A comment can be added to the sample chromatogram by right-clickingin the Peak Review window of the sample and selecting SampleAnnotation. Type “Training Alteration” and select OK. The altered data file will now be annotated and this annotation will be printed on thechromatogram for that data file (Figure8-26).Figure 8-26Sample Annotation on Printed Chromatogram13.Integrated peaks for each sample can also be reviewed using the PeakReview Slide Show. Right-click in the peak review pane or window and select the Slide Show Peak Review option. The Peak Review Slide Show controls appear (Figure8-27). Change the delay time to 2 seconds and press the green circle button to start the slide show. The peak reviewwindow then automatically scrolls through the data files with a 2-second delay.Figure 8-27Peak Review Slide Show Controls14.To stop reviewing the peaks, select the red X in the slide show options.Peak review can also be paused, sent back to the beginning of the Results Table, or a previous page can be viewed. Alternatively, the peak review for the first or last data file can also be viewed by right clicking in the peak review pane or window and selecting Show First (or Last) Page.15.The chromatograms for all data files can be printed by selecting File -Print - Pane from the menu when the Peak Review Pane (or Window) is active. This will bring up a menu to choose a printer, and finally a window will appear to choose what is to be printed (Figure8-28). You can choose。
液质联用仪操作说明
型号:LCQ DECA 美国热电集团
用途:主要应用在中药成分研究,药代动力学研究,毒品和兴奋剂的检查,农药残留量的分析,生物大分子的研究等领域。
操作规程:
1、开启氦气高压钢瓶,打开真空泵开关,抽真空5~6小时。
2、打开液相泵、自动进样器和PDA检测器,打开质谱仪开关(按住按钮几秒种)进行通讯,若通讯不畅可按reset键。
3、打开Xcalibur软件,根据要分析的样品设定程序。
4、将预处理好的样品放入自动进样器的样品盘中,开启氮气高压钢瓶,运行3中所设程序
5、利用Xcalibur或Bioworks等软件进行定性、定量分析。
6、检测完样品后,须用合适的溶剂清洗仪器,对仪器进行正常维护,同时避免上次样品对下次测定造成影响。
7、将仪器打到Standby状态,关掉质谱开关,停30~60S后再关真空泵开关。
8、所有进质谱的试剂必需用进口色谱纯的,尤其是在做生物大分子时,其前处理试剂也必需用进口色谱纯的,而且一定要进行脱盐处理。
AB液质操作规程
第一部分开关机A A 、开机1.打开UPS电源,打开氮气发生器(气瓶),确认Curtain gas的输入压力为0.4MPa,gas1/2为0.7MPa,exhuast为0.4MPa。
2.打开机械泵电源开关(5500系列直接打开质谱主机电源,仪器自动启动机械泵)3.等机械泵工作至少30min5.打开质谱主机电源开关6.过夜抽真空B B 、关机1.停止输液:关停针泵或液相泵,或断开输液管线(一般质谱主机Standby后液相系统也自动待机,输液泵自动关停,但建议操作人员再次确认,必须保证没有任何液体再泵入质谱)2.使仪器待机,并灭活配置2.关闭质谱仪主机电源开关(5500系列只需按下电源开关旁的VENT按键排放真空)3.机械泵继续工作至少30min4.关闭机械泵上的电源开关(5500系列会自动关停机械泵,待机械泵停止后即可关闭5500系列主机上的电源开关)5.关闭气体发生器或气瓶第二部分调谐1、PPG就是正离子校正液,PPG3000负离子调谐液,本台仪器只用两瓶校正液,正离子一瓶,负离子一瓶。
洗液一瓶50%的甲醇水。
调谐前先用洗液洗针两次。
吸校正液的时候慢慢吸,避免吸入气泡,吸好以后将针泵卡住。
再将高度调到5,将管路重新连接。
不再连接六通阀,由质谱直接进样,且将针泵的速度调到10μL/ml。
2、调谐液弄好后在仪器面板工具栏上:TOOl→项目→创建项目。
此项目为一个文件夹,文件夹里面包括所有此次采集的信息,方法、序列等,且所有采集的项目都固定在一个文件夹下面:D:/Analyst Data/Project。
新建项目的时候,弹出窗口中有个Add All,可以创建孙文件夹。
3、联机:软件打开后在左上方硬件配置(单机选中,双击打开)→只需选中仪器MS→active profile。
图标变为绿色,表示联机成功。
调谐只需要选中质谱。
仪器的实时状态可以点击右下方的图标,质谱和液相都是单独观察,质谱要注意真空度,为0.5.4、调谐→手动调谐→项目(文件夹)选中下拉菜单中的Installation 20161102(为系统默认的调谐文件夹),先Q1→start syring pump(打开针泵)→start(质谱),开始采集以后观察基线是否平衡,若等两分钟还未平衡好,则手动将针泵往上抬。
ab液质离子净培训手册
ab液质离子净培训手册
AB液质离子净培训手册
一、产品介绍
AB液质离子净是一种高效、环保的净水处理系统,通过独特的离子交换技术,能够有效去除水中的有害物质,提供安全的饮用水。
本手册将为您详细介绍AB液质离子净的使用、操作和维护方法。
二、使用步骤
1. 开机前的准备
确保电源已连接,检查所有管路连接是否紧固,无渗漏现象。
检查储水桶是否清洁,必要时进行清洗。
2. 开机操作
按下主机面板上的电源开关,系统将自动进行自检。
自检完成后,将自动进行离子交换过程。
此时,指示灯亮起,表示系统正在运行。
3. 关机操作
完成水处理后,按下主机面板上的关机按钮,系统将停止运行。
关闭电源开关,拔下电源插头。
4. 日常维护
定期检查管路连接是否紧固,确保无渗漏现象。
定期清洗储水桶,保持清洁卫生。
三、注意事项
1. 请勿在水中添加任何添加剂,以免影响离子交换效果。
2. 若长时间不使用本产品,请将电源插头拔下,避免造成能源浪费或造成安全隐患。
3. 当出现异常声音或异常情况时,应立即停止使用,并联系专业人员进行检修。
4. 在使用过程中,请勿随意拆卸、修理或改造本产品,以免造成安全隐患或损坏产品。
四、常见问题及解决方法
1. 问题:出水流量小
解决方法:检查管路是否堵塞,清洗管路或更换滤芯。
2. 问题:出水口感不佳
解决方法:检查离子交换树脂是否需要更换,或检查储水桶是否清洁。
3. 问题:指示灯不亮或无法开机
解决方法:检查电源插头是否插好,检查电源线是否损坏,如有问题,请联系专业人员进行检修。
AB液质操作规程完整
AB液质操作规程完整第一部分开关机A A 、开机1.打开UPS电源,打开氮气发生器(气瓶),确认Curtain gas 的输入压力为,gas1/2为,exhuast为。
2.打开机械泵电源开关(5500系列直接打开质谱主机电源,仪器自动启动机械泵)3.等机械泵工作至少30min5.打开质谱主机电源开关6.过夜抽真空B B 、关机1.停止输液:关停针泵或液相泵,或断开输液管线(一般质谱主机Standby后液相系统也自动待机,输液泵自动关停,但建议操作人员再次确认,必须保证没有任何液体再泵入质谱)2.使仪器待机,并灭活配置2.关闭质谱仪主机电源开关(5500系列只需按下电源开关旁的VENT按键排放真空)3.机械泵继续工作至少30min4.关闭机械泵上的电源开关(5500系列会自动关停机械泵,待机械泵停止后即可关闭5500系列主机上的电源开关)5.关闭气体发生器或气瓶第二部分调谐1、PPG就是正离子校正液,PPG3000负离子调谐液,本台仪器只用两瓶校正液,正离子一瓶,负离子一瓶。
洗液一瓶50%的甲醇水。
调谐前先用洗液洗针两次。
吸校正液的时候慢慢吸,避免吸入气泡,吸好以后将针泵卡住。
再将高度调到5,将管路重新连接。
不再连接六通阀,由质谱直接进样,且将针泵的速度调到10μL/ml。
2、调谐液弄好后在仪器面板工具栏上:TOOl→项目→创建项目。
此项目为一个文件夹,文件夹里面包括所有此次采集的信息,方法、序列等,且所有采集的项目都固定在一个文件夹下面:D:/Analyst Data/Project。
新建项目的时候,弹出窗口中有个Add All,可以创建孙文件夹。
3、联机:软件打开后在左上方硬件配置(单机选中,双击打开)→只需选中仪器MS→active profile。
图标变为绿色,表示联机成功。
调谐只需要选中质谱。
仪器的实时状态可以点击右下方的图标,质谱和液相都是单独观察,质谱要注意真空度,为.4、调谐→手动调谐→项目(文件夹)选中下拉菜单中的(为系统默认的调谐文件夹),先Q1→start syring pump(打开针泵)→start(质谱),开始采集以后观察基线是否平衡,若等两分钟还未平衡好,则手动将针泵往上抬。
液质联用操作说明
液质联用操作说明液质联用操作说明一、1.使用前先检查流动相是否充足,若不够,请及时添加。
(B瓶是甲醇,C瓶是乙腈)2.排气:把液相泵打开,将里面的开关逆时针旋转180°,打开显示器,打开软件“LCQ”,单击窗口左上角的按钮联机,单击purge 排气,待purge重新变回黑色,表明排气完成。
将液相泵开关顺时针旋转180°,直至旋不动为止。
3.选择流动相溶剂,“solvent value B”:100%(配样溶剂应与流动相保持一致),“flow rate”:0.2mlL/min,“Download”,“pump on”,“close”。
二、1.观察电喷雾源里面是否有液体排出,等有液体流出后将里面的针取出放于固定的位置。
2.点击软件窗口中的“on/standby”按钮,使其处于“on”状态。
此时窗口中会出现很多杂乱无章的峰,仪器上“scan”灯亮蓝光,表示已经接通高压电流,其余的灯亮黄绿色。
3.先用溶剂洗涤注射器,然后吸取一定量的溶剂,在“inject”状态,将注射器插入进样孔,切换成“load”状态,将注射器中的液体注入进样口,进完样之后切换到“inject”状态。
观察窗口中显示出的背景峰,重复注射溶剂,直到背景峰杂乱无章,没有出现特别明显的峰即可。
4.点击“Define scan”选择分子量范围,尽可能使需要的峰位于横坐标的中间,点击“apply”,然后点击“OK”5.点击“Acquire Date”,选择文件保存路径。
在E盘的“teachers”中选择自己导师的文件夹(没有的就自己新建一个),然后建议创建一个以日期命名的文件夹,例如:“E/teachers/wangzhaoyang/20161231”,6.进样:将化合物名字改为对应的名称,单击“start”,再点击“view”,然后再点击下窗口右上角的按钮,使其变为绿色,进样。
7.在上窗口出现峰的位置拖动鼠标,下窗口就可以显示出该处峰的具体分子量情况,同时可以不停的按“F5”对窗口显示的界面进行刷新。
液质联用操作规程完整
1.适用范围本设备配备ACQUITY UPLC液相色谱仪、TQS MS/MS 质谱仪,适用于食品、药品中各种有机物的定性、定量分析,是一种具有高灵敏度的检测仪器,仪器由主机、计算机和数据处理软件等组成。
2. 职责2.1操作人员按照本规程操作仪器,认真填写实验使用记录。
2.2保管人员负责对仪器进行定期维护和保养。
2.3 科室负责人负责监督检查规程的执行。
3.操作程序日常操作步骤:准备UPLC —→设置样品表—→运行样品—→定量—→打印报告。
注:如果一个星期内不运行样品请不要关质谱仪,使其保持真空。
建立新方法和project的操作步骤:准备UPLC —→建立新的project —→用标准品调谐—→编辑质谱方法—→编辑UPLC方法—→设置样品表—→运行样品—→定量—→打印报告。
3.1开机:3.1.1 彻底开机顺序(仪器已关闭)确定MS及其它仪器电源电缆已连接,开氮气发生器、开氩气,小表<0.1mpa。
打开计算机电源 > 等待windows正常启动 >电脑界面右下角网络图标红叉。
打开UPLC自动进样器电源,等到电脑界面右下角网络图标出现感叹号!。
打开UPLC泵电源,等约30s或者是有响声。
打开质谱电源,等待5min,离子源透视镜里面亮。
打开Masslynx软件,masslynx主界面 -----左侧instrument----Mass tune---界面菜单栏vacuum---pump 同样界面左侧偏上diagnostics---vacuum---analyser MS1 turbo speed[%]要在5分钟内升到80。
至少抽真空4个小时 > 查看真空状态主界面mass console---界面左侧xevo tq ms detector加号展开---ms display > 碰撞室真空度 >达到 7.x e-5mbar 。
3.1.2 日常开机顺序(仪器未关闭)开氮气发生器、开氩气,小表<0.1mpa,打开Masslynx软件,进入工作站。
液质联用操作规程完整
1.适用范围本设备配备ACQUITY UPLC液相色谱仪、TQS MS/MS 质谱仪,适用于食品、药品中各种有机物的定性、定量分析,是一种具有高灵敏度的检测仪器,仪器由主机、计算机和数据处理软件等组成。
2. 职责2.1操作人员按照本规程操作仪器,认真填写实验使用记录。
2.2保管人员负责对仪器进行定期维护和保养。
2.3 科室负责人负责监督检查规程的执行。
3.操作程序日常操作步骤:准备UPLC —→设置样品表—→运行样品—→定量—→打印报告。
注:如果一个星期内不运行样品请不要关质谱仪,使其保持真空。
建立新方法和project的操作步骤:准备UPLC —→建立新的project —→用标准品调谐—→编辑质谱方法—→编辑UPLC方法—→设置样品表—→运行样品—→定量—→打印报告。
3.1开机:3.1.1 彻底开机顺序(仪器已关闭)确定MS及其它仪器电源电缆已连接,开氮气发生器、开氩气,小表<0.1mpa。
打开计算机电源 > 等待windows正常启动 >电脑界面右下角网络图标红叉。
打开UPLC自动进样器电源,等到电脑界面右下角网络图标出现感叹号!。
打开UPLC泵电源,等约30s或者是有响声。
打开质谱电源,等待5min,离子源透视镜里面亮。
打开Masslynx软件,masslynx主界面 -----左侧instrument----Mass tune---界面菜单栏vacuum---pump 同样界面左侧偏上diagnostics---vacuum---analyser MS1 turbo speed[%]要在5分钟内升到80。
至少抽真空4个小时 > 查看真空状态主界面mass console---界面左侧xevo tq ms detector加号展开---ms display > 碰撞室真空度 >达到 7.x e-5mbar 。
3.1.2 日常开机顺序(仪器未关闭)开氮气发生器、开氩气,小表<0.1mpa,打开Masslynx软件,进入工作站。
液质联用操作规程
1.适用范围本设备配备ACQUITY UPLC液相色谱仪、TQS MS/MS 质谱仪,适用于食品、药品中各种有机物的定性、定量分析,是一种具有高灵敏度的检测仪器,仪器由主机、计算机和数据处理软件等组成。
2. 职责2.1操作人员按照本规程操作仪器,认真填写实验使用记录。
2.2保管人员负责对仪器进行定期维护和保养。
2.3科室负责人负责监督检查规程的执行。
3.操作程序日常操作步骤:准备UPLC —→设置样品表—→运行样品—→定量—→打印报告。
注:如果一个星期内不运行样品请不要关质谱仪,使其保持真空。
建立新方法和project的操作步骤:准备UPLC —→建立新的project —→用标准品调谐—→编辑质谱方法—→编辑UPLC方法—→设置样品表—→运行样品—→定量—→打印报告。
3.1开机:3.1.1 彻底开机顺序(仪器已关闭)确定MS及其它仪器电源电缆已连接,开氮气发生器、开氩气,小表<0.1mpa。
打开计算机电源> 等待windows正常启动>电脑界面右下角网络图标红叉。
打开UPLC自动进样器电源,等到电脑界面右下角网络图标出现感叹号!。
打开UPLC泵电源,等约30s或者是有响声。
打开质谱电源,等待5min,离子源透视镜里面亮。
打开Masslynx软件,masslynx主界面-----左侧instrument----Mass tune---界面菜单栏vacuum---pump 同样界面左侧偏上diagnostics---vacuum---analyser MS1 turbo speed[%]要在5分钟内升到80。
至少抽真空4个小时> 查看真空状态主界面mass console---界面左侧xevo tq ms detector加号展开---ms display > 碰撞室真空度>达到7.x e-5mbar 。
3.1.2 日常开机顺序(仪器未关闭)开氮气发生器、开氩气,小表<0.1mpa,打开Masslynx软件,进入工作站。
液质联用操作规程
1.适用范围本设备配备ACQUITY UPLC液相色谱仪、TQS MS/MS 质谱仪,适用于食品、药品中各种有机物的定性、定量分析,是一种具有高灵敏度的检测仪器,仪器由主机、计算机和数据处理软件等组成。
2. 职责2。
1操作人员按照本规程操作仪器,认真填写实验使用记录。
2.2保管人员负责对仪器进行定期维护和保养。
2.3科室负责人负责监督检查规程的执行。
3.操作程序日常操作步骤:准备UPLC —→设置样品表—→运行样品-→定量—→打印报告。
注:如果一个星期内不运行样品请不要关质谱仪,使其保持真空。
建立新方法和project的操作步骤:准备UPLC —→建立新的project -→用标准品调谐-→编辑质谱方法-→编辑UPLC方法—→设置样品表-→运行样品-→定量—→打印报告。
3。
1开机:3.1.1 彻底开机顺序(仪器已关闭)确定MS及其它仪器电源电缆已连接,开氮气发生器、开氩气,小表〈0。
1mpa。
打开计算机电源> 等待windows正常启动>电脑界面右下角网络图标红叉。
打开UPLC自动进样器电源,等到电脑界面右下角网络图标出现感叹号!。
打开UPLC泵电源,等约30s或者是有响声。
打开质谱电源,等待5min,离子源透视镜里面亮.打开Masslynx软件,masslynx主界面--———左侧instrument-——-Mass tune——-界面菜单栏vacuum-——pump 同样界面左侧偏上diagnostics-——vacuum—--analyser MS1 turbo speed[%]要在5分钟内升到80.至少抽真空4个小时〉查看真空状态主界面mass console—-—界面左侧xevo tq ms detector加号展开-——ms display 〉碰撞室真空度>达到7.x e-5mbar 。
3。
1。
2 日常开机顺序(仪器未关闭)开氮气发生器、开氩气,小表〈0.1mpa,打开Masslynx软件,进入工作站。
液质联用操作规程
1.适用范围本设备配备ACQUITY UPLC液相色谱仪、TQS MS/MS 质谱仪,适用于食品、药品中各种有机物的定性、定量分析,是一种具有高灵敏度的检测仪器,仪器由主机、计算机和数据处理软件等组成.2. 职责2.1操作人员按照本规程操作仪器,认真填写实验使用记录。
2.2保管人员负责对仪器进行定期维护和保养。
2。
3科室负责人负责监督检查规程的执行。
3.操作程序日常操作步骤:准备UPLC —→设置样品表-→运行样品—→定量-→打印报告。
注:如果一个星期内不运行样品请不要关质谱仪,使其保持真空。
建立新方法和project的操作步骤:准备UPLC -→建立新的project -→用标准品调谐-→编辑质谱方法—→编辑UPLC方法—→设置样品表—→运行样品—→定量—→打印报告.3.1开机:3。
1.1 彻底开机顺序(仪器已关闭)确定MS及其它仪器电源电缆已连接,开氮气发生器、开氩气,小表〈0.1mpa。
打开计算机电源〉等待windows正常启动〉电脑界面右下角网络图标红叉。
打开UPLC自动进样器电源,等到电脑界面右下角网络图标出现感叹号!.打开UPLC泵电源,等约30s或者是有响声.打开质谱电源,等待5min,离子源透视镜里面亮。
打开Masslynx软件,masslynx主界面-—-—-左侧instrument----Mass tune—--界面菜单栏vacuum--—pump 同样界面左侧偏上diagnostics—-—vacuum-—-analyser MS1 turbo speed[%]要在5分钟内升到80。
至少抽真空4个小时> 查看真空状态主界面mass console——-界面左侧xevo tq ms detector加号展开—--ms display > 碰撞室真空度〉达到7.x e—5mbar .3.1。
2 日常开机顺序(仪器未关闭)开氮气发生器、开氩气,小表<0。
1mpa,打开Masslynx软件,进入工作站。
- 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
- 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
- 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
1.1 IntroductionThis chapter is intended for the user to become familiar with manually tuning and calibrating the instrument.In this section, users will:•Open a PPG acquisition method•Acquire 10 MCA scans of PPG spectrum•Adjust Peak Widths to achieve Unit Resolution•Manually tune and calibrate the instrument using the Calibrate from Spectrum icon•Replace Mass Calibration•Review Calibration Results with instructor•Optimize the CEM1.2 Materials•Syringe Pump•Syringe•PEEK tubing and union adapter•Appropriate PPG calibrant solution1.3 Procedure•Insert the syringe into the PEEK tubing and union adapter•Tighten fittings to ensure there will be no leaks•Place syringe on pump and lock in place•Slide push plate up against plunger of syringe•Set flow rate and diameter of syringe•Plumb the loose end of the PEEK tubing to the inlet of the source•Press start to begin infusion1.4 Manual Tuning and Mass Calibration1.4.1 Choosing Hardware Configuration1.Choose the API Instrument Project folder from the drop down menu nextto the yellow folder icon on the task bar.2.Double-click the Hardware Configuration and activate the Mass SpecOnly Profile. After a green check appears for the selected profile, closethe window.Manual Tune, Mass Calibration and CEM1.4.2 Manual Tuning Set-up1.Click on the Tune and Calibrate Mode on the Navigation Bar. You shouldhear the gases turn on as the instrument becomes active for use. A greenmass spec status icon also appears in the bottom right corner of thewindow.2.Under the View menu, activate the Sample Queue or click on the ViewQueue icon to open the Queue Manager window. Minimize the QueueManager window.3.Double-click on Manual Tuning. Under File, select Open and choose anacquisition method (*.dam) for calibration. Start with Q1 positivecalibration. For positive Q1, select Q1PosPPG.dam, for negative Q3,select Q3NegPPG.dam and so on. The method then comes up in theTune Method Editor window, shown in Figure1-1.Figure 1-1Manual Tune Window with “Q1 PosPPG.dam” Method Open4.The settings on the Source/Gas, Compound, Resolution and Detectorsettings found on the tabs on the left side of the window as well as the MSand Advanced MS settings on the right side are values established by theinstallation engineer. Consult the help files if you have any questionsregarding the settings in this window.5.The API 2000™ and API 3200™ Systems have integrated syringe pumps,be sure to create the hardware profile with it activated on the Mass SpecConfiguration tab. Click on the Mass Spec and click Set Up.Manual Tune, Mass Calibration and CEMFigure 1-2Integrated Syringe Pump Set Up1.4.3 PPG Dilution and Syringe Pump Set UpDifferent calibration solutions must be used, depending on the specific instrument and the polarity being calibrated.1.Load a syringe with the appropriate PPG solution.Table 1-1PPG Solutions and Dilutions By Instrument InstrumentTypeNegative ModePositive ModeHigh MassAPI 2000™ System PPG 3000 Standard PPG Standard (10-4M)NA API 3000™ System PPG 3000 Standard PPG Standard, Diluted 1:10 (10-5M)PPG 3000 Standard API 3200™ System PPG 3000 Standard PPG Standard, Diluted 1:10 (10-5M)PPG 3000 StandardAPI 4000™ System PPG 3000 Standard PPG Standard, Diluted 1:50(2x10-6M)PPG 3000Standard API 5000™ SystemPPG 3000 Standard, Diluted 1:10PPG Standard, Diluted 1:500(2x10-7M)PPG 3000 Standard Note:The PPG Standard should be diluted with the appropriate diluent. DONOT use the vial marked “Diluent” in the Mass Spec Standards Kit. The appropriate dilution solution for the PPG Standard is: dissolve 15.4 mg of ammonium acetate in 49.9mL of water, then add 49.9 mL of methanol, 0.1 mL acetonitrile, and 0.1 mL formic acid. The appropriate dilution solution for the PPG 3000 standard is: dissolve 385 mg ammonium acetate in 49.9 mL of water, then add, 49.9 mL of methanol, 0.1 mL acetonitrile, and 0.1 mL formic acid.Manual Tune, Mass Calibration and CEM2.API 3000™, API 4000™, API 5000™ Systems: Start the infusion pump at aflow rate of 10 μL/min. Ensure that the appropriate syringe diameter is being used.3.API 2000™ and API 3200™ Systems: Use the drop down menu to theright of Edit Ramp to access the Syringe Pump Method dialog box. Input a rate of 10 μL/min and an appropriate diameter for the syringe being used. Click Set Flow Rate or Start Syringe Pump to start the infusion. Then go back to the MS Method Editor by selecting MS Method from the drop down menu to the right of Edit Ramp.Figure 1-3API 2000™ and API 3200™ Systems Syringe Pump Method Set-up4.Allow enough time for the infusion rate to equilibrate.5.Verify that the sprayer is positioned properly.API 2000™, API 3000™ Systems: The edge of the spray should be at least 2 mm from the curtain plate hole. Momentarily increase flow rate to 100 μL/min to visibly see the spray for proper spray adjustment relative to the orifice.API 3200™, API 4000™, & API 5000™ Systems: Settings of 10 vertical and 4 to 6 horizontal are recommended.1.4.4 Verify PPG Ion Intensities and Peak Shape1.On the MS tab, enter 10 cycles and verify that MCA box is checked. ClickStart to acquire 10 scans in MCA mode.Table 1-2Syringe DiametersSyringe Volume (μL)Diameter (mm)1000 4.61500 3.262502.30Manual Tune, Mass Calibration and CEM2.Right click in the bottom-right pane and select Open File.3.Right click in one of the spectral panes and select List Data.A new pane will open below the spectra with three tabs: Data List, Calibration Peak List, and Peak List.Figure 1-4Calibration Peak List Tabpare the PPG intensities displayed to a previous calibration of theinstrument. Guidelines for intensities at Unit resolution can be found at the end of this chapter.5.If the instrument is operating correctly, all intensity values should bewithin 15% of the previous calibration values. 6.Observe the isotopic pattern of each panel. For all PPG ions >1700 amu,the second isotope is the ion of interest for calibration purposes. Also, forNote:MCA is Multi-Channel Acquisition and when it is activated the successive scans will be summed in a running total of intensities starting from the first scan up to the current scan. This is most useful when acquiring data from low intensity peaks, because adding several scans can produce better signal-to-noise ratios. This mode of operation is used for discrete sample analysis by infusion. For PPG tuning and calibration, 10 MCA scans is the standard approach.Tip!If the Calibration Peak List tab is not visible, navigate to the Appearance Options to activate it by following Figure1-5.Manual Tune, Mass Calibration and CEMall PPG ions >1700 amu, a change in the isotope pattern might beobserved if solution is old.Figure 1-5Activating Mass Calibration Peak List Tab1.4.5 Adjusting Peak Widths to Achieve Unit ResolutionOn the Calibration Peak List tab, examine the peak width column. Peak widths orresolution should be consistent across the entire mass calibration range. Toachieve “Unit” resolution, the DC offsets of the PPG ions whose peak widthvalues are outside 0.6 - 0.8 amu should be adjusted.To adjust Peak Widths:1.Return to the Tune Method Editor window. Click on the Resolution tab.The Ion Energy (IE1) value was established at installation and is seldomchanged. This parameter affects peak shape, sensitivity and resolution.Click the Advanced button. The Resolution Table (see Figure1-6) thatappears contains a list of all the PPG ions being calibrated on with theircorresponding DC Offset values.2.Capture the offset values prior to making any changes by usingCtrl+PrntScrn and paste the capture into NotePad (that way you have astarting point to return to). Make small changes (+/- 0.005) in the offsetvalues of the PPG ions whose peak widths are out of range. Click Apply,then re-acquire another 10 MCA scans (click YES when prompted to“Save changes made in manual tune to instrument data file”). Go back tothe spectral panes, right click and List Data again and observe if therewere any changes to the peak width.Try making big changes (+/- 0.05),click Apply, re-acquire 10 MCA scans, and note how peak width changes.Increasing the offset will increase the resolution, which in turndecreases the Peak WidthsManual Tune, Mass Calibration and CEMDecreasing the offset will decrease the resolution, which in turn increases the peak widths3.After changes are made, re-acquire 10 scans in MCA mode, right click inbottom right pane and select Open File, then use the List Data tab again, and examine the peak widths for Unit Resolution.Figure 1-6Resolution Table for Q1 Positive4.Determine if any changes need to be made and adjust the offsets asnecessary, re-acquiring 10 MCA scans to observe the changes in peak widths.5.Continue to adjust Offsets to achieve Unit resolution for the peak widthsacross the entire mass range.Table 1-3Peak Resolution Specifications ResolutionPeak Width (amu)FWHHLow ~1.1 amu Unit 0.6 - 0.8 amu High0.4 - 0.6 amuNote:You MUST Tune and Calibrate High Resolution if you are going to use it. Each has its own resolution and calibration tables. Low Resolution is a specified drop of all the Offsets from Unit Resolution Offsets and cannot be calibrated.Manual Tune, Mass Calibration and CEM1.4.6 Replacing the Mass Calibration1.Once Unit Resolution is achieved across the mass range, use the Toolsmenu to select Calibrate from Spectrum, or click the Calibrate from Spectrum icon on the toolbar to begin manual mass calibration.2.The Mass Calibration Options dialogue box appears (Figure 1-7). Selectthe appropriate standard from the drop down list. Click on Edit to update the Reference Table if needed. Be sure to enter an appropriate Search Range and Threshold for the calibration.Figure 1-7Mass Calibration Options3.Click Start.The Mass Calibration Report window opens (Figure 1-8) whichcontains three graphs.Tip!There is a tendency for resolution to decrease as ion mass increases;therefore, it is necessary to adjust the ratio of the RF to DC voltages applied to the quadrupole. Increasing the DC voltage as a function of mass allows you to achieve a constant peak width for each ion over a large mass range. The rate at which the DC voltage is varied with mass is set using the Mass and Offset settings in the Resolution Table window. In general, as the resolution is increased (i.e narrow peak widths), the ion intensity decreases (i.e less sensitive). The normal trade-off between high ion transmission (increasedsensitivity) and narrow peak widths (increased resolution) should be optimized for each application. Most applications will require “unit” mass resolution (i.e., isotope peaks one amu apart are clearly defined) that corresponds to peak widths of approximately 0.7 +/- 0.1 amu at 50% intensity (Full Width at Half Height, FWHH). It is usually desirable to have a consistent peak width over theentire mass range.Manual Tune, Mass Calibration and CEMFigure 1-8Mass Calibration Report GraphsThe top graph is a plot of the mass shift from the previous calibration, where the solid line represents a 0 amu shift and the dotted lines arethe limits of + and - 0.1 amu shift.The middle graph is a plot of the full peak widths at half height for all of the PPG ions used in the calibration. The solid line represents 0.7amu peak width, and the dotted lines are the limits of + and - 0.1 amu.The bottom graph is a plot of the % change in intensity of the PPG ions from the previous calibration.4.If there are any outliers in the top two graphs, you must not attemptrecalibration, instead you must determine the cause of the outliers. If the peak widths are outside the specifications of Unit Resolution, then you must adjust the peak widths appropriately. If the peak widths are meet Unit Resolution and there are still outliers in the top graph, thenrecalibration is necessary.5.To replace the existing calibration, click on the bottom red arrow iconlocated at the top left corner of the results window (See Figure1-8). Click Yes when prompted to accept the new calibration.The top red arrow is to update the calibration, for example, if there were masses added to the method and Reference Table to extend the calibration.6.Print (Ctrl+P) the Mass Calibration Report.7.There is also a Mass Calibration Results text window (Figure1-9) thatopens behind the graphs. Minimize the graph window to access the text report.This text report contains the same information in tabular form and also has the updated DAC values which are the RF/DC values for massManual Tune, Mass Calibration and CEMcalibration on the quad, as well as another parameter called SlopeVariation.If you plot the masses vs. DAC values, you should get a straight lineand that is where the Slope Variation is derived. This value should bebetween 0.998 and 1.002. If it is not, then the instrument may havebeen calibrated on the wrong isotopic peak for that ion.Figure 1-9Partial Mass Calibration Results Text WindowTune and Calibration Review•You now have successfully calibrated the instrument in the specifiedpolarity and quadrupole. Remember that for each resolution setting (unitand high), there are 4 instrument calibrations: Q1 positive, Q3 positive,Q1 negative, Q3 negative.•The DC offsets and DAC values are stored in the Instrument Data Filefolder which is located in the API Instrument Project folder. You can accessthese values in any project in Analyst® software. Click Configure in thenavigation bar. Click on Tools in the taskbar and select Settings thenInstrument Options. Click Quad 1, Positive, Unit. Then click CalibrationTable. The DAC values are displayed. Click Resolution Table. The DCoffsets and Ion Energy are displayed.•Each combination of quad, polarity and resolution will have its owncalibration and resolution settings. These settings will automatically getupdated whenever the instrument is calibrated in that polarity andresolution.•Click Configure in the navigation bar. Click on Tools in the taskbar andselect Settings then Tuning Options. The Resolution tab allows the user toset the Resolution specifications for Unit, High and Low. The Lowresolution offsets are automatically generated by taking the UnitResolution DC offsets and lowering each offset by the specified OffsetDrop.1.4.7 Automatic Tuning and CalibrationThe instrument can be tuned and calibrated automatically using the InstrumentOptimization function under the Tune and Calibrate mode. Please refer to theGetting Started Guide for detailed instructions for using the automatic feature.1.4.8 CEM OptimizationThe Channel Electron Multiplier (CEM) degrades with time and continued use and must therefore be re-optimized periodically. Manual CEM Optimization can be done using the standard positive PPG calibration solution. PPG mass 1254 amu is monitored for the optimization (mass 906 amu for API 5000™ System).1.Infuse the standard positive PPG calibration solution at 10μL/min.2.Open the PPG Q1 Pos method in Manual Tune. Modify the method asfollows:Check the “Center/Width” (if it isn’t already) to alter how the masses are displayed (toggle the check mark to familiarize yourself with thedifferent displays). Uncheck Parameter Range if it is checked.Highlight and delete all rows except the one corresponding to mass 1254. You can save the method with a different method name, such as“CEM Opt.dam”.Set the duration to 10 minutes and uncheck MCA.3.In the Detector tab, drop the CEM voltage by 200 V. Click Start and beginacquiring data.When a stable baseline has been established, increase CEM voltage in increments of 100 V, allowing enough time between changes todetermine intensity at a given setting. When the increase in signalintensity is 20 – 40% (optimal is 30%) with a 200 V increase, theoptimum CEM voltage has been determined. Remember – increasingthe CEM voltage not only increases the signal, but also the noise.Figure 1-10CEM Gain Check TICNote:API 2000™ and API 3200™ Systems have a different kind of detector, so the procedure for optimization is slightly different. Instead of a 200 V decrease, use a 100 V decrease and 50 V increments to gather the data points. Theoptimal setting will be determined with a 100 V difference producing the 20% - 40% gain in signal.4.In this example, 2150 V would be the optimal CEM setting, with 36%gain.The percentage signal gain will vary depending on the mass of the ion. In general, the gain is higher at higher mass. Therefore, it is very important to use the proper criteria for determining the CEM setting to be used.1.4.9 Intensity Guidelines For Positive ModeThe intensity values in these tables serve as guidelines only. The calibration intensities from installation are the best reference. All values are based on MCA of 10 scans.WARNING!Setting the bias voltage above the optimized voltage shortens the CEM life.Table 1-4Intensity Guidelines for API 5000™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)59NA NA 0.6 - 0.8175NA NA 0.6 - 0.8500 1.1e7 1.1e70.6 - 0.8906 1.7e7 1.4e70.6 - 0.811964.3e63.4e60.6 - 0.8Table 1-5Intensity Guidelines for API 4000™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)59 3.0e7 2.0e70.6 - 0.8175 1.0e7 1.0e70.6 - 0.8616 1.5e7 1.0e70.6 - 0.8906 2.0e7 2.0e70.6 - 0.81254 3.0e6 3.0e60.6 - 0.81545 1.0e6 5.0e50.6 - 0.82010 1.5e6 1.0E60.6 - 0.8522421.0E68.0E50.6 - 0.85Table 1-6Intensity Guidelines for API 3200™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)599.0E69.0E60.6 - 0.81759.0E67.7E60.6 - 0.8616 4.5E6 3.8E60.6 - 0.8906 1.1E78.1E60.6 - 0.81254 1.3E68.6E50.6 - 0.815452.7E51.4E50.6 - 0.81.4.10 Intensity Guidelines for Negative ModeThe intensity values in these tables serve as guidelines only. The calibrationintensities from installation is the best reference. All values are based on MCA of 10 scans.Table 1-7Intensity Guidelines for API 3000™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)59 1.2E7 1.2E70.6 - 0.8175 1.2E7 1.2E70.6 - 0.8616 1.0E79.1E60.6 - 0.8906 1.9E7 1.5E70.6 - 0.81254 1.3E6 1.2E60.6 - 0.81545 4.5E5 3.7E50.6 - 0.82010 1.5E6 1.0E60.6 - 0.8522421.0e66.1e50.6 - 0.85Table 1-8Intensity Guidelines for API 2000™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)59 2.0e61.0E6 1.2E70.6 - 0.8175 2.0E6 1.0E60.6 - 0.8906 2.0E78.0E60.6 - 0.81254 2.0E6 1.0E50.6 - 0.81545 1.0E6 1.0E50.6 - 0.817781.0E61.0E50.6 - 0.85Table 1-9Intensity Guidelines for API 5000™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)9341.1e79.0e60.6 - 0.8Table 1-10Intensity Guidelines for API 4000™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)934 2.0e7 1.0e70.6 - 0.820363.0e6-0.6 - 0.85Table 1-11Intensity Guidelines for API 3200™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)934 1.5e67.2e50.6 - 0.815721.4e5-0.6 - 0.81.5 Discuss and Answer the following Questions:1.In most labs, the service engineer performs the calibration every 6 monthsduring PM service. Why do I need to know how to calibrate the instrument?2.How often should the instrument be calibrated?3.If you would like to add another PPG ion to the acquisition method, wherecan you find a list of PPG ions to choose from? How can you calculate the DAC value for that ion and estimate the DC offset?4.What should you do so that in the event of a major computer crash, youcan easily recover your mass calibration?5.Under the Resolution tab, what is the function of the parameter IonEnergy (IE1)? See lecture presentation.6.I opened a Q1Pos PPG method that was set at Unit Resolution and wassuccessful at manual calibration. I used the same method but changed Resolution to High --- the peaks disappeared. What should I do? I used the same method but changed Resolution to Low ---- what should I expect?7.What is the difference in the mass displayed when the Center/Widthcheck box is activated or not?8.How often should I optimize CEM?9.What are some factors that can lead to a short CEM life?Table 1-12Intensity Guidelines for API 3000™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)45 2.0e5 1.2e50.6 - 0.8585 2.0e5 1.2e50.6 - 0.89347.1e6 5.5e60.6 - 0.81224 2.0e5 1.2e50.6 - 0.81572 2.0e5 2.0e50.6 - 0.82036 1.5e67.5e50.6 - 0.822115.0e54.0e40.6 - 0.85Table 1-13Intensity Guidelines for API 2000™ System Nominal m/zQ1 (cps)Q3 (cps)FWHH (amu)45 2.0e4 2.0e30.6 - 0.8585 2.0e4 2.0e30.6 - 0.8934 1.0e5 2.0e40.6 - 0.81224 4.0e4 1.0e30.6 - 0.81572 4.0e4 1.0e30.6 - 0.817471.0e41.0e30.6 - 0.810.What is the maximum CEM voltage before replacement is needed? At the end of this chapter, you should have the following printouts:•Mass Calibration Results text file.•Mass Calibration Report Graphs.。