• 能用面板, Prolink II, HC275, AMS更改 • 默认为不自动翻滚



“SCROLL” 键: – 在菜单间滚动, “NO”/取消, 增加数字 “SELECT” 键: – 选择菜单, “Yes”/确认, 移位 “SCROLL” + “SELECT”: – 从“OPERATIONS”菜单进入“ALARMS / MAINTENANCE”菜 单
Micro Motion Confidential Page 21
常见诊断代码
代码
A19 A20 A21 A22 A23 A24 A25 A26 A100 A101 A102 A103 A104 A105 A107 A108 A109 A110 A111 A112
275显示信息
RAM Error - 1000/2000 Calibration Factor Unentered (Flocal) Unrecognized/Unentered Sensor Type (K1) EEPROM Config Corrupt - Core Processor EEPROM Totals Corrupt - Core Processor EEPROM Program Corrupt - Core Processor Core Processor Boot Sector Fault Sensor/Xmtr Communication Error Analog output 1 saturated Analog output 1 fixed Drive over range Data loss possible Calibration in progress Slug flow Power reset occurred Event 1 triggered Event 2 triggered Frequency over range Freq output fixed Series 1000/2000 software upgrade recommended

所有线圈对地要求开路 LPO 和 RPO阻值差别应小于 < 10%
Micro Motion Confidential Page 36
传感器故障诊断

RTD 阻值
– CMF, D-系列, F-系列, R-系列, CNG传感器, T-系列的紫色黄 色或紫色到橙色线阻值:
- 流量计OK - 信息 - 报警
支持所有基本开车功能:
流量计调零
模拟输出 输出量程 安全密码
用于C1D1 / Zone 1 带反馈的按键, 易于操作.
Micro Motion Confidential Page 9
面板操作 (LDO)

参数自动翻滚 – 翻滚间隙可在1-10秒间选择 (默认为10 秒)
智能诊断技术

传感器诊断 变送器诊断 核心处理器诊断 组态&过程诊断
Micro Motion Confidential Page 23
故障处理基本程序

检查报警及观察故障现象,记录产品序列号 采取措施要由简单到繁琐

先查看报警及过程变量 检查组态及外部连接 传感器等部件测试
B Q
Right P/O (-) Temp (-)
3
RTD RTD COMP
Right P/O (+)
6
8
9
Left P/O (-)
Temp Shield
4 5
Left P/O (+)
Temp (+)
7 1
2
Drive (-)
Drive (+)
Numbers correspond to terminal strip.
Micro Motion Confidential Page 35
传感器故障诊断

驱动线圈和检测线圈阻值
(T)/(others) – 驱动 – LPO – RPO – 棕色到红色: – 绿色到白色: – 蓝色到灰色: 8 to 500/2650 ohms 16 to 1000/300 ohms 16 to 1000/300 ohms
MVD多变量数字变送器 -- 智能化传感器的组成

核心处理器
– 一体化/分体安装 – 存储传感器系数 – 18-30 volts DC – 3 watts – Modbus 输出
Micro Motion Confidential Page 4
MVD多变量数字变送器 -- 智能化传感器的组成

变送器
MMI质量流量计故障诊断
Micro Motion 产品系列
Micro Motion Confidential Page 2
MVD多变量数字变送器 -- 智能化传感器的组成

Micro Motion 9Βιβλιοθήκη Baidu电缆
– – – – 所有型号 所有材质 D, Elite, R, F, T 所有口径
Micro Motion Confidential Page 3
传感器特性化报警

未输入正确的传感器参数 传感器/核心处理器损坏
Micro Motion Confidential Page 31
传感器故障报警

LPO&RPO小于目标幅值的%5~10% 仅表示一个结果 重新上电(如果有A2) 充满/清洗传感器
Micro Motion Confidential Page 32
Micro Motion Confidential Page 28
mA输出固定

HART地址不为零 处于回路测试模式 关闭回路电流模式 PV设定错误
Micro Motion Confidential Page 29
零点故障

有流量的时候执行了调零操作 上电重新启动
Micro Motion Confidential Page 30
275显示信息
EEPROM Checksum - Core Processor RAM Error - Core Processor Sensor failure Temperature out of range Input over range Field device not characterizeded Real time interrupt failure Density outside limits Field device warming up Calibration failed Excess calibration correction, zero too low Excess calibration correction, zero too high Process too noisy to perform auto zero Electronics failure Data loss possible Line RTD Overrange Case RTD Overrange EEPROM Checksum - 1000/2000
4-线接口 (2-电源, 2-通讯) RS-485通讯
– Modbus RTU 协议 – Auto-侦测波特率、奇偶和停止位


12 - 30 VDC 输入 LED 状态指示 包含仪表标定系数 所有的累积、事件寄存器、增强变量计算、和仪表诊断信息等都包 含在核心处理器中 快速的DSP信号处理,增强性能,等等
Micro Motion Confidential Page 24
常见报警



A26&A14&A9 A102&A3 A102&A100&A110 A101 A10&A11&A12 A6&A20&A21 A102&A3&A2 A5&A3A102 A107
Micro Motion Confidential Page 25
用Prolink II故障诊断
Micro Motion Confidential Page 15
用Prolink II故障诊断
Micro Motion Confidential Page 16
Prolink II – 诊断信息
Micro Motion Confidential Page 17
核心处理器诊断信息
Micro Motion Confidential Page 10
面板操作 (LDO)

“Alarms/Maintenance”菜单
– See Alarms(观看报警) • Alarm Code (s): - 报警代码 – 关键报警 A1 ~ A99 – 一般报警 A100 ~ A200 • ACK all? (确认所有报警) • Exit(退出) – Off-Line Maintenance(离线组态) • xxxx Code? (如设置为Enable, 默认为1234) • Sim(模拟输出)), Zero(调零), Unit & Output Configuration(单位和输出组态), Exit(退出)
Micro Motion Confidential Page 7
利用面板开车/组态
Micro Motion Confidential Page 8
易于开车
显示所有过程参数 能够启动/停止/复位累计量 彩色LED提供流量计状态显示
外壳和显示可以 90 旋转360 .
• 绿色 • 黄色 • 红色
A26报警

通讯错误报警
Micro Motion Confidential Page 26
驱动增益报警

不满管 气液两相 传感器堵塞/挂壁 其他
Micro Motion Confidential Page 27
输出饱和报警

测量值超出组态输出的范围 零点不准确 不稳定/驱动增益饱和
Micro Motion Confidential Page 11
17002700的手操器连接
Micro Motion Confidential Page 12
传感器诊断信息
Micro Motion Confidential Page 13
17002700的Prolink连接
Micro Motion Confidential Page 14
Micro Motion Confidential Page 18
用DeltaV AMS故障诊断
Micro Motion Confidential Page 19
AMS – 诊断信息

关键报警
Micro Motion Confidential Page 20
常见诊断代码
代码
A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18
可能的故障
流量计需要返修 重新上电/流量计需要返修 检查测试点 检查测试点 检查测试点 检查特性化. 特别要检查FCF和K1值 流量计需要返修 检查测试点 流量计预热 重新上电, 然后重新标定 重新上电, 然后重新调零 重新上电, 然后重新调零 消除或降低干扰, 重新调零 变送器需要返修 重新上电,检查组态数据 检查测试点 检查测试点 变送器需要返修
传感器输入错误

测量驱动、检测线圈及RTD的阻值 测量驱动、检测线圈及RTD对外壳 检查9线连接电缆(如果有)
Micro Motion Confidential Page 33
断电报警

检查供电电源 报警本身无需处理
Micro Motion Confidential Page 34
核心处理器 - 九芯插针
– 本质安全栅 – 显示 – 输出
Micro Motion Confidential Page 5
MVD多变量数字变送器 -- 核心处理器 分解图
Cover Ring Puck Assy Base Feed thru
Micro Motion Confidential Page 6
核心处理器

Micro Motion Confidential Page 22
可能的故障
重新上电或返修 检查特性化。 特别要检查FCF值 检查特性化。 特别要检查K1值 流量计需要返修 流量计需要返修 流量计需要返修 流量计需要返修 检查变送器和核心处理器间的接线是 否正确 调整模拟输出量程 检查HART地址。或正处于回路测试状 态 驱动增益超限 重新上电,检查组态数据 流量计处于标定阶段 消除团流,调整团流限制和团流时间 无需任何措施 报警触发 报警触发 调整频率输出量程 处于频率测试状态 变送器需要升级 4