关于橡胶压缩永久变形测试标准的文章

合集下载

d395橡胶压缩永久变形特性试验方法

d395橡胶压缩永久变形特性试验方法

Designation:D395–02Standard Test Methods forRubber Property—Compression Set1This standard is issued under thefixed designation D395;the number immediately following the designation indicates the year of original adoption or,in the case of revision,the year of last revision.A number in parentheses indicates the year of last reapproval.A superscript epsilon(e)indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1.Scope1.1These test methods cover the testing of rubber intended for use in applications in which the rubber will be subjected to compressive stresses in air or liquid media.They are applicable particularly to the rubber used in machinery mountings,vibra-tion dampers,and seals.Two test methods are covered as follows:Test Method Section A—Compression Set Under Constant Force in Air7–10B—Compression Set Under Constant Deflection in Air11–14 1.2The choice of test method is optional,but consideration should be given to the nature of the service for which correlation of test results may be sought.Unless otherwise stated in a detailed specification,Test Method B shall be used.1.3Test Method B is not suitable for vulcanizates harder than90IRHD.1.4The values stated in SI units are to be regarded as the standard.1.5This standard does not purport to address all of the safety concerns,if any,associated with its use.It is the responsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2.Referenced Documents2.1ASTM Standards:D1349Practice for Rubber—Standard Temperatures for Testing2D3182Practice for Rubber—Materials,Equipment,and Procedures for Mixing Standard Compounds and Prepar-ing Standard Vulcanized Sheets2D3183Practice for Rubber—Preparation of Pieces for Test Purposes from Products2D3767Practice for Rubber—Measurement of Dimensions2 D4483Practice for Determining Precision for Test Meth-ods Standards in the Rubber and Carbon Black Industries2E145Specification for Gravity-Convection and Forced-Ventilation Ovens33.Summary of Test Methods3.1A test specimen is compressed to either a deflection or by a specified force and maintained under this condition for a specified time and at a specified temperature.3.2The residual deformation of a test specimen is measured 30min after removal from a suitable compression device in which the specimen had been subjected for a definite time to compressive deformation under specified conditions.3.3After the measurement of the residual deformation,the compression set,as specified in the appropriate test method,is calculated according to Eq1and Eq2.4.Significance and Use4.1Compression set tests are intended to measure the ability of rubber compounds to retain elastic properties after pro-longed action of compressive stresses.The actual stressing service may involve the maintenance of a definite deflection, the constant application of a known force,or the rapidly repeated deformation and recovery resulting from intermittent compressive forces.Though the latter dynamic stressing,like the others,produces compression set,its effects as a whole are simulated more closely by compressionflexing or hysteresis tests.Therefore,compression set tests are considered to be mainly applicable to service conditions involving static stresses.Tests are frequently conducted at elevated tempera-tures.5.Test Specimens5.1Specimens from each sample may be tested in duplicate (Option1)or triplicate(Option2).The compression set of the sample in Option1shall be the average of the two specimens expressed as a percentage.The compression set of the sample in Option2shall be the median(middle most value)of the three specimens expressed as a percentage.5.2The standard test specimen shall be a cylindrical disk cut from a laboratory prepared slab.5.2.1The dimensions of the standard specimens shall be:1These test methods are under the jurisdiction of ASTM Committee D11onRubber and are the direct responsibility of Subcommittee D11.10on PhysicalTesting.Current edition approved Dec.10,2002.Published January2003.Originallyapproved st previous edition approved in2001as D395–01.2Annual Book of ASTM Standards,V ol09.01.3Annual Book of ASTM Standards,V ol14.04.1Copyright©ASTM International,100Barr Harbor Drive,PO Box C700,West Conshohocken,PA19428-2959,United States.Type1A2BThickness,mm(in.)12.560.5(0.4960.02)6.060.2 (0.2460.01)Diameter,mm(in.)29.060.5(1.1460.02)13.060.2 (0.5160.01)A Type1specimen is used in Test Methods A and B.B Type2specimen is used in Test Method B.5.2.2When cutting the standard specimen,the circular die having the required inside dimensions specified in5.2.1shall be rotated in a drill press or similar device and lubricated by means of a soap solution.A minimum distance of13mm(0.51 in.)shall be maintained between the cutting edge of the die and the edge of the slab.The cutting pressure shall be as light as possible to minimize cupping of the cut edges.The dies shall be maintained carefully so that the cutting edges are sharp and free of nicks.5.3An optional method of preparing the standard specimen may be the direct molding of a circular disk having the dimensions required for the test method used and specified in 5.2.1.N OTE1—It should be recognized that an equal time and temperature,if used for both the slab and molded specimen,will not produce an equivalent state of cure in the two types of specimen.A higher degree of cure will be obtained in the molded specimen.Adjustments,preferably in the time of cure,must be taken into consideration if comparisons between the specimens prepared by different methods are to be considered valid. N OTE2—It is suggested,for the purpose of uniformity and closer tolerances in the molded specimen,that the dimensions of the mold be specified and shrinkage compensated for therein.A two-plate mold with a cavity13.060.1mm(0.51060.004in.)in thickness and29.2060.05 mm(1.14860.002in.)in diameter,with overflow grooves,will provide Type1specimens for Test Method A and Test Method B.A similar mold but having a cavity of6.360.3mm(0.2560.012in.)in thickness and 13.260.1mm(0.5260.004in.)in diameter will provide Type2 specimens for Test Method B.5.4When the standard test specimen is to be replaced by a specimen taken from a vulcanized rubber part of greater thickness than the one indicated in5.2.1,the sample thickness shall be reducedfirst by cutting transversely with a sharp knife and then followed by buffing to the required thickness in accordance with Practice D3183.5.5An alternative method of preparing specimens is by plying up cylindrical disks cut from a standard sheet prepared in accordance with Practice D3182using the specimen sizes specified in5.2.1and cutting as described in5.2.2,or where a drill press is not available cutting the specimens with a single stroke from a cutting die.5.5.1The disks shall be plied,without cementing,to the thickness required.Such plies shall be smooth,flat,of uniform thickness,and shall not exceed seven in number for Type1 specimens and four in number for Type2specimens.5.5.2Care shall be taken during handling and placing of the plied test specimen in the testfixture by keeping the circular faces parallel and at right angles to the axis of the cylinder.5.5.3The results obtained on plied specimens may be different from those obtained using solid specimens and the results may be variable,particularly if air is trapped between disks.5.5.4The results obtained on the specimens prepared by one of the methods may be compared only to those prepared by the same method.5.6For routine or product specification testing,it is some-times more convenient to prepare specimens of a different size or shape,or both.When such specimens are used,the results should be compared only with those obtained from specimens of similar size and shape and not with those obtained with standard specimen.For such cases,the product specification should define the specimen as to the size and shape.If suitable specimens cannot be prepared from the product,the test method and allowable limits must be agreed upon between the producer and the purchaser.6.Conditioning6.1Store all vulcanized test specimens or product samples to be tested at least24h but not more than60days.When the date of vulcanization is not known,make tests within60days after delivery by the producer of the article represented by the specimen.6.2Allow buffed specimens to rest at least30min before specimens are cut for testing.6.3Condition all specimens before testing for a minimum of 3h at2362°C(73.463.6°F).Specimens whose compression set properties are affected by atmospheric moisture shall be conditioned for a minimum of24h in an atmosphere controlled to5065%relative humidity.7.Precision and Bias47.1These precision statements have been prepared in ac-cordance with Practice D4483.Please refer to Practice D4483 for terminology and other testing and statistical concepts.7.2Prepared test specimens of two rubbers for Test MethodsA andB were supplied tofive laboratories.These were tested in duplicate each day on two separate testing days.A test result, therefore,is the average of two test specimens,for both Test Methods A and B.7.3One laboratory did not run the Test Method A testing; therefore,the precision for Test Method A is derived from four laboratories.7.4The Type1precision results are given in Table1and Table2.4Supporting data are available from ASTM Headquarters.Request RR: D11–1138.TABLE1Type1Precision Results,%Compression Set—TestMethod AMaterialMeanLevelWithin Laboratory A Between Laboratory AS r r(r)S R R(R)1 1.73(%)0.0500.1428.20.1900.5431.1 226.10.898 2.549.7 2.37 6.7125.7A Sr=within laboratory standard deviation.r=repeatability(in measurement units).(r)=repeatability(in percent).S R=between laboratory standard deviation. R=reproducibility(in measurement units). (R)=reproducibility(in percent).7.5Bias—In test method statistical terminology,bias is the difference between an average test value and the reference or true test property value.Reference values do not exist for these test methods since the value or level of the test property is exclusively defined by the test method.Bias,therefore,cannot be determined.TEST METHOD A—COMPRESSION SET UNDERCONSTANT FORCE IN AIR8.Apparatus8.1Dial Micrometer—A dial micrometer,for measuring specimen thickness,in accordance with Practice3767,Method A1.8.2Compression Device,consisting of a force application spring and two parallel compression plates assembled bymeans of a frame or threaded bolt in such a manner that the device shall be portable and self-contained after the force has been applied and that the parallelism of the plates shall be maintained.The force may be applied in accordance with either 8.2.1or8.2.2.8.2.1Calibrated Spring Force Application—The required force shall be applied by a screw mechanism for compressing a calibrated spring the proper amount.The spring shall be of properly heat-treated spring steel with ends ground and per-pendicular to the longitudinal axis of the spring.A suitable compression device is shown in Fig. 1.The spring shall conform to the following requirements:8.2.1.1The spring shall be calibrated at room temperature 2365°C(73.469°F)by applying successive increments of force not exceeding250N(50lbf)and measuring the corresponding deflection to the nearest0.2mm(0.01in.).The curve obtained by plotting the forces against the corresponding deflections shall have a slope of7063.5kN/m(400620 lbf/in.)at1.8kN(400lbf).The slope is obtained by dividing the two forces above and below1.8kN by the difference between the corresponding deflections.8.2.1.2The original dimensions of the spring shall not change due to fatigue by more than0.3mm(0.01in.)after it has been mounted in the compression device,compressed under a force of1.8kN(400lbf),and heated in the oven for one week at70°C62°C(15863.6°F).In ordinary use,a weekly check of the dimensions shall show no greater change than this over a period of1year.8.2.1.3The minimum force required to close the spring (solid)shall be2.4kN(530lbf).8.2.2External Force Application—The required force shall be applied to the compression plates and spring by external means after the test specimen is mounted in the apparatus. Either a calibrated compression machine or known masses may be used for force application.Provision shall be made by the use of bolts and nuts or other devices to prevent the specimen and spring from losing their initial deflections when the external force is removed.The spring shall have essentially the same characteristics as described in8.2.1,but calibration is not required.A suitable compression device is shown in Fig.2.8.3Plates—The plates between which the test specimen is compressed shall be made of steel of sufficient thickness to withstand the compressive stresses without bending.8.3.1The surfaces against which the specimen is held shall have a chromium platedfinish and shall be cleaned thoroughly and wiped dry before each test.8.3.2The steel surfaces contacting the rubber specimens shall be ground to a maximum roughness of250µm(10µin.) and then chromium plated and polished.8.3.3The chromium plating and subsequent polishing shall not affect thefinalfinish beyond the tolerance stated in8.3.2.8.4Oven,conforming to the specification for a Type IIB laboratory oven given in Specification E145.8.4.1Type IIB ovens specified in Test Method E145are satisfactory for use through70°C.For higher Temperatures Type II A ovens are necessary.8.4.2The interior size shall be as follows or of an equivalent volume:TABLE2Type1Precision Results,%Compression Set—TestMethod BMaterial MeanLevelWithin Laboratory A Between Laboratory AS r r(r)S R R(R)113.7(%)0.591 1.6712.2 1.54 4.3631.8 252.80.567 1.60 3.0 5.9216.831.7 A Sr=within laboratory standard deviation.r=repeatability(in measurement units).(r)=repeatability(in percent).S R=between laboratory standard deviation.R=reproducibility(in measurement units).(R)=reproducibility(in percent).Interior size of air oven:min.300bt300mm by300mm(12by12by12in.)max.900by900by1200mm(36by36by48in.) 8.4.3Provision shall be made for placing test specimens in the oven without touching each other or the sides of the aging chamber.8.4.4The heating medium for the aging chamber shall be air circulated within it at atmospheric pressure.8.4.5The source of heat is optional but shall be located in the air supply outside of the aging chamber.8.4.6A suitable temperature measurement device located in the upper central portion of the chamber near the test speci-mens shall be provided to record the actual aging temperature.8.4.7Automatic temperature control by means of thermo-static regulation shall be used.8.4.8The following special precautions shall be taken in order that accurate,uniform heating is obtained in all parts of the aging chamber.8.4.8.1The heated air shall be thoroughly circulated in the oven by means of mechanical agitation.When a motor driven fan is used,the air must not come in contact with the fan motor brush discharge because of danger of ozone formation.8.4.8.2Baffles shall be used as required to prevent local overheating and dead spots.8.4.8.3The thermostatic control device shall be so located as to give accurate temperature control of the heating medium. The preferred location is adjacent to the temperature measuring device listed in section8.4.6.8.4.8.4An actual check shall be made by means of maxi-mum reading thermometers placed in various parts of the oven to verify the uniformity of the heating.9.Procedure9.1Original Thickness Measurement—Measure the original thickness of the specimen to the nearest0.02mm(0.001in.). Place the specimen on the anvil of the dial micrometer so that the presser foot will indicate the thickness at the central portion of the top and bottom faces.9.2Application of Compressive Force—Assemble the specimens in the compression device,using extreme care to place them exactly in the center between the plates to avoid tilting.If the calibrated spring device(see Fig.1)is used,apply the compressive force by tightening the screw until the deflection as read from the scale is equivalent to that shown on the calibration curve for the spring corresponding to a force of 1.8kN(400lbf).With the external loading device(see Fig.2), apply this force to the assembly in the compression machine or by adding required masses,but in the latter case,take care to add the mass gradually without shock.Tighten the nuts and bolts just sufficiently to hold the initial deflections of the specimen and spring.It is imperative that no additional force be applied in tightening the bolts.9.3Test Time and Test Temperature—Choose a suitable temperature and time for the compression set,depending upon the conditions of the expected service.In comparative tests,use identical temperature and heating periods.It is suggested that the test temperature be chosen from those listed in Practice D1349.Suggested test periods are22h and70h.The specimen shall be at room temperature when inserted in the compression device.Place the assembled compression device in the oven within2h after completion of the assembly and allow it to remain there for the required test period in dry air at the test temperature selected.At the end of the test period,take the device from the oven and remove the specimens immedi-ately and allow it to cool.9.4Cooling Period—While cooling,allow the specimens to rest on a poor thermally conducting surface,such as wood,for 30min before making the measurement of thefinal thickness. Conduct the cooling period at a standard laboratory tempera-ture of2362°C(73.463.6°F).Specimens whose compres-sion set property is affected by atmospheric moisture shall be cooled in an atmosphere controlled to5065%relative humidity.9.5Final Thickness Measurement—After the rest period, measure thefinal thickness at the center of the specimen in accordance with9.1.10.Calculation10.1Calculate the compression set as a percentage of the original thickness as follows:C A5@~t o2t i!/t o#3100(1) where:C A=Compression set(Test Method A)as a percentage ofthe original thickness,t o=original thickness(see9.1),andt i=final thickness (see 9.5).11.Report11.1Report the following information:11.1.1Original dimensions of the test specimen,including the original thickness,t o ,11.1.2Actual compressive force on the specimen as deter-mined from the calibration curve of the spring and spring deflection reading (see 8.2.1)or as applied by an external force (see 8.2.2),11.1.3Thickness of the test specimen 30min after removal from the clamp,t i ,11.1.4Type of test specimen used,together with the time and temperature of test,11.1.5Compression set,expressed as a percentage of the original thickness,11.1.6Test method used (Test Method A),and 11.1.7Number of specimens tested.TEST METHOD B—COMPRESSION SET UNDERCONSTANT DEFLECTION IN AIR 12.Apparatus12.1Dial Micrometer —A dial micrometer,for measuring the specimen thickness,in accordance with Practice D 3767,Method A 1.N OTE 3—For vulcanizates having a hardness below 35IRHD,the force on the presser foot should be reduced to 0.260.05N (0.0460.01lbf).12.2Spacer Bars ,to maintain the constant deflection re-quired under Test Method B.12.2.1Spacer bars for Type 1samples shall have a thickness of 9.560.02mm (0.37560.001in.).12.2.2Spacer bars for Type 2samples shall have a thickness of 4.5060.01mm (0.177060.0005in.).12.3Compression Device ,consisting of two or more flat steel plates between the parallel faces of which the specimens may be compressed as shown in Fig.3.Steel spacers for the required percentage of compression given in 13.2shall be placed on each side of the rubber specimens to control their thickness while compressed.12.4Oven ,see 8.4.12.5Plates —The plates between which the test specimen is compressed shall be made of steel of sufficient thickness to withstand the compressive stresses without bending.12.5.1The surfaces against which the specimen is held shall have a chromium-plated finish and shall be cleaned thoroughly and wiped dry before each test.12.5.2The steel surfaces contacting the rubber specimens shall be ground to a maximum roughness of 250µm (10µin.)and then chromium plated and polished.12.5.3The chromium plating and subsequent polishing shall not affect the final finish beyond the tolerance stated in 12.5.2.13.Procedure13.1Original Thickness Measurement —Measure the origi-nal thickness of the specimen to the nearest 0.02mm (0.001in.).Place the specimen on the anvil of the dial micrometer so that the presser foot will indicate the thickness at the central portion of the top and bottom faces.13.2Application of Compressive Force —Place the test specimen between the plates of the compression device with the spacers on each side,allowing sufficient clearance for the bulging of the rubber when compressed (see Fig.3).Where a lubricant is applied,it shall consist of a thin coating of a lubricant having substantially no action on the rubber.For most purposes,a silicon or fluorosilicon fluid is suitable.Tighten the bolts so that the plates are drawn together uniformly until they are in contact with the spacers.The amount of compression employed shall be approximately 25%.A suitable mechanical or hydraulic device may be used to facilitate assembling and disassembling the test fixture.13.3Test Time and Temperature —Choose a suitable tem-perature and time for the compression set,depending upon the conditions of the expected service.In comparative tests,use identical temperature and test periods.It is suggested that the test temperature be chosen from those listed in Practice D 1349.Suggested test periods are 22h and 70h.The test specimen shall be at room temperature when inserted in the compression device.Place the assembled compression device in the oven within 2h after completion of the assembly and allow it to remain there for the required test period in dry air at the test temperature selected.At the end of the test period,take the device from the oven and remove the test specimen immediately and allow them to cool.13.4Cooling Period —While cooling,allow the test speci-men to rest on a poor thermally conducting surface,such as wood,for 30min before making the measurement of the final thickness.Maintain the conditions during the cooling period in accordance with 9.4.13.5Final Thickness Measurement —After the rest period,measure the final thickness at the center of the test specimen in accordance with 13.1.14.Calculation14.1Calculate the compression set expressed as a percent-age of the original deflection as follows:C B 5@~t o 2t i !/~t o 2t n !#3100(2)where:C B =compression set (Test Method B)expressed aspercentage of the original deflection,t 0=original thickness of specimen (13.1),t i =final thickness of specimen (13.5),and t n =thickness of the spacer bar used.N OTE 4—Lubrication of the operating surfaces of thecompressionFIG.3Device for Compression Set Test Under ConstantDeflection,Test MethodBdevice is optional while giving more reproducible results;lubrication may somewhat alter the compression set values.15.Report15.1Report the following information:15.1.1Original dimensions of the test specimen including the original thickness,t o,15.1.2Percentage compression of the specimen actually employed,15.1.3Thickness of the test specimen30min after removal from the clamp,t i,15.1.4Type of test specimen used,together with the time and temperature of test,15.1.5Whether or not the surfaces of the compression device are lubricated.If they are,what type lubrication was used,15.1.6Compression set,expressed as a percentage of the original deflection,15.1.7Test method used(Test Method B),and15.1.8Number of specimens tested.16.Keywords16.1compression set;compression set under constant de-flection;compression set under constant force;deflection; deformation;elastic property;hysteresis;recoveryASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned in this ers of this standard are expressly advised that determination of the validity of any such patent rights,and the risk of infringement of such rights,are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed everyfive years and if not revised,either reapproved or withdrawn.Your comments are invited either for revision of this standard or for additional standards and should be addressed to ASTM International Headquarters.Your comments will receive careful consideration at a meeting of the responsible technical committee,which you may attend.If you feel that your comments have not received a fair hearing you should make your views known to the ASTM Committee on Standards,at the address shown below.This standard is copyrighted by ASTM International,100Barr Harbor Drive,PO Box C700,West Conshohocken,PA19428-2959, United States.Individual reprints(single or multiple copies)of this standard may be obtained by contacting ASTM at the above address or at610-832-9585(phone),610-832-9555(fax),or service@(e-mail);or through the ASTM website().。

橡胶制品的压缩永久变形如何调整

橡胶制品的压缩永久变形如何调整

橡胶制品的压缩永久变形如何调整摘要:密封橡胶制品是橡胶工业制品的重要组成部分,为获得可靠的长期密封性,一般对胶料的压缩永久变形都有严格的要求。

本文选用了几种常用作密封制品的橡胶作为综述对象,浅述了目前对此类橡胶压缩永久变形的研究概况。

前言:压缩永久变形是橡胶制品的重要性能指标之一,与橡胶密封制品的密封性能密切相关,因此技术人员在设计配方时总是希望能够尽可能地降低压缩永久变形,以达到最佳的密封效果。

硫化橡胶压缩永久变形的大小,涉及到硫化橡胶的弹性与恢复。

有些人往往简单地认为橡胶的弹性好,其恢复就快,永久变形就小。

这种理解是不够的,弹性与恢复是相互关联的两种性质。

但有时候,橡胶的本质没有发生根本的变化,永久变形的大小主要是受橡胶恢复能力的变化所支配。

影响恢复能力的因素有分子之问的作用力、网络结构的变化、分子间的位移等【1】。

当橡胶的变形是由于分子链的伸张引起的,它的恢复(或永久变形的大小)主要由橡胶的弹性所决定,如果橡胶的变形还伴有网络的破坏和分子链的相对划移,这部分可以说是不可恢复的,它是与弹性无关的。

所以,凡是影响橡胶弹性与恢复的因素,都是影响硫化橡胶压缩永久变形的因素。

当然橡胶压缩变形的测试方法一定意义上决定了所测数值的大小。

如杨红卫等人【2】根据对不同形状的试样进行研究,发现由于B型试样截面直径较小,而相反它的曲率半径较大,顶部受压缩的程度也就越严重,且在相同体积下,B型试样与空气接触面积是A型试样的2.2倍,这就是说在实验过程中,B型试样的老化机会要大于A型试样,因此B型试样的压缩永久变形大于A型,同时橡胶的热空气老化是由表及里的,试样越大,内部的老化就会越慢,这也是A型试样的压缩永久变形小于B型试样的一个因素。

而对于10×10mm试样,因为是在室温下恢复,此时的橡胶分子活性较低,难以充分恢复,因此压缩变形相对于A型、B型的高温下恢复而较大。

因此,按GB/T 7759—1996进行试验,B型试样的压缩永久变形大于A型试样;按GB/T7759—1996对B型试样进行试,按GB/T 1683—1981对10×l0mm 试样进行试验,10×l0mm试样的压缩永久变形大于B型试样。

三元乙丙橡胶力学及压缩永久变形性能研究

三元乙丙橡胶力学及压缩永久变形性能研究

三元乙丙橡胶力学及压缩永久变形性能研究硫化胶力学性能越好,但耐老化性能差;蒙脱土和纳米凹凸棒粒径越小,硫化胶力学性能和耐老化性能越好,但压缩永久变形性能差;BaSO_4粒径越小,硫化胶力学性能和耐老化性能越好,压缩永久变形性能先增加后减小。

最后,本文通过扫描电镜(SEM)、透射电镜(TEM)、X射线衍射(XRD)、傅里叶变换红外(FTIR)等手段,对填料的微观形貌、结构和化学性质进行了表征和分析,为填料的选择及应用提供了理论和实验依据。

4.3 结果与讨论4.3.1 硫化特性在本研究中,我们使用了两种不同的硫化系统,即常规硫化系统和高效硫化系统,并对其进行了比较。

结果表明,使用高效硫化系统可以显著提高硫化速率和交联密度,从而改善EPDM的硬度和强度。

此外,我们还研究了硫化剂种类和用量对硫化特性的影响。

结果表明,使用过量的硫化剂会导致硫化密度过高,从而降低EPDM的拉伸性能。

4.3.2 力学性能我们测试了EPDM的拉伸强度、断裂伸长率、硬度和抗撕裂性能,并比较了不同硫化条件下的结果。

结果表明,使用高效硫化系统可以显著提高EPDM的拉伸强度和硬度,但对断裂伸长率和抗撕裂性能的影响不大。

此外,我们还研究了填料对EPDM力学性能的影响。

结果表明,添加适量的填料可以显著提高EPDM的强度和硬度,但过量的填料会导致EPDM的断裂伸长率和抗撕裂性能下降。

4.3.3 耐老化性能我们使用热氧老化试验和紫外线老化试验来研究EPDM的耐老化性能,并比较了不同硫化条件下的结果。

结果表明,使用高效硫化系统可以显著提高EPDM的耐热老化性能和耐紫外线老化性能。

此外,我们还研究了填料对EPDM耐老化性能的影响。

结果表明,添加适量的填料可以显著提高EPDM的耐老化性能,但过量的填料会导致EPDM的老化速度加快。

4.3.4 压缩永久变形性能我们测试了EPDM的压缩永久变形率,并比较了不同硫化条件下的结果。

结果表明,使用高效硫化系统可以显著降低EPDM的压缩永久变形率。

橡胶压缩永久变形标准

橡胶压缩永久变形标准

橡胶压缩永久变形标准
橡胶是一种常见的弹性材料,广泛应用于汽车制造、建筑工程、电子产品等领域。

然而,橡胶制品在使用过程中可能会出现压缩永久变形的情况,这对产品的质量和可靠性都会造成影响。

因此,制定橡胶压缩永久变形标准具有重要意义。

首先,橡胶制品的压缩永久变形是指在一定温度、一定时间的条件下,橡胶制品在受到压缩后,无法完全恢复原状的现象。

这种变形会导致橡胶制品的尺寸稳定性下降,影响其正常使用。

因此,制定橡胶压缩永久变形标准可以帮助生产厂家和用户更好地评估产品质量,提高产品的可靠性和耐久性。

其次,橡胶压缩永久变形标准应包括以下内容,首先是测试方法,即确定测试温度、测试时间和测试压力等参数,以保证测试结果的准确性和可比性。

其次是标准数值,即规定橡胶制品在特定条件下的压缩永久变形限值,通常以百分比表示。

最后是评定标准,即根据测试结果对橡胶制品的压缩永久变形进行等级评定,以便用户选择和使用。

在制定橡胶压缩永久变形标准时,需要考虑以下因素,首先是橡胶材料的种类和用途,不同种类的橡胶在受压后的变形特性可能有所不同,因此需要针对不同的应用领域进行标准制定。

其次是环境因素,如温度、湿度等环境条件对橡胶制品的压缩永久变形也会产生影响,因此需要在标准中对环境因素进行考虑。

最后是用户需求,不同用户对橡胶制品的要求也会有所不同,因此标准制定过程中需要充分考虑用户的实际需求。

总之,制定橡胶压缩永久变形标准对于保障产品质量、提高产品可靠性具有重要意义。

通过科学、严谨的标准制定,可以有效地评估和监控橡胶制品的压缩永久变形性能,为用户提供更加可靠和耐用的产品,推动橡胶制品行业的健康发展。

ASTMD395-2003压缩永久变形中文版

ASTMD395-2003压缩永久变形中文版

ASTMD395-2003压缩永久变形中文版编号:D 395-03橡胶性能的标准试验方法----------压缩永久变形1此项标准在固定编号B 117下发布,紧随编号的数字表示标准采纳的年度,如果是修正,数字表示最后一次修正的年度。

在括号内的数字表示最后一次重申批准的年度。

上标表示自最后一次修正或重申批准以来的编辑改动。

此项标准已被批准供美国国防部下属机构使用。

1范围1.1本测试方法测试应用中会在气体或液体媒介中承受压力的橡胶。

本测试方法特别适用于在机械固定器件,减震器,封条中使用的橡胶。

本测试方法包含以下两种方法:1.2测试方法可以选择,但是应考虑用于与测试结果关联的实际情况下使用的橡胶的性质。

除非在具体的规范中有其他规定,应使用测试方法B。

测试方法B不适用于硬度大于90IRHD的硫化橡胶。

以国际单位(SI)为单位的数值应被认为是标准。

在括号内的数值起参照作用。

此项标准不包括与其应用有关的所有的安全隐患。

此项标准的使用者有责任在使用前建立合适的安全健康规范以及决定法规限制是否适用2 参考文件ASTM标准2:D1349 橡胶规范---测试的标准温度D 3182D 3183D 3767D 4483E 145---------------------------------------1此测试方法属于ASTM D 11橡胶委员会的工作范围,是其下属物理测试子委员会的直接责任。

目前的版本在批准,出版。

原始的版本在1934年批准。

上一个版本在2003年批准,编号为D395-03.2如需参照ASTM 标准,访问ASTM网站,. 如需要《ASTM标准年鉴》的内容信息,浏览ASTM网站的标准索引页。

3 测试方法概要用挠力或规定的力压缩试样,并在规定的温度下保持规定的时间。

在试样在合适的装置内,在规定的条件下经过特定时间的压缩变形后,取出试样,等待30分钟,测量试样的残留变形。

在测量残留变形后,根据Eq1和Eq2计算压缩永久变形。

橡胶制品压缩永久变形测试

橡胶制品压缩永久变形测试

创作编号:GB8878185555334563BT9125XW创作者:凤呜大王*橡胶制品压缩永久变形测试1.定义和方法橡胶压缩永久变形,是指压缩橡胶试样在完全去掉引起其压缩形变的力之后所剩余的变形。

其用于判定橡胶材料的交织密度,受力状况下的物性。

试验方法通常有三种:1)方法A:在恒定压力作用下,空气中作压缩试验2)方法B:在空气中恒定形变压缩试验3)方法C:在空气(气体)或液体中,恒定形变压缩试验在方法的选择中一般选用B,但是方法B、C不适合于IRHD>90℃的硬度胶料中。

以上三种方法可以做常温、高温、低温或溶液中的形变测试。

2.简单的测试步骤如下:1)按照要求制作压缩永久变形的试块或直接用产品或部分产品(如O-ring,Washer,Disc等);2)用夹具将试块固定并压缩到一定的压缩量(压缩率),在一定试验条件(通常是一定温度和时间,有时会浸泡在溶液中测试)后取出;3)在2的操作过程中记录相应数据,同时记录取出的产品在室温下放置30分钟后的数值(有些客户要求不松开夹具放置30分钟,后松开30分钟后测量);4)按照压缩永久变形的公式计算在要求温度时间和变形量的前提下的压缩永久变形。

3.压缩永久变形CS的计算方法:CS=(h0-h2)/(h0-h1)h0:压缩前试样的高度,mmh1:限制器的高度,mmh2:试样恢复后的高度,mm4.结果判定:在压缩永久变形中,对于所测的每一个样品,都要在标准内,否则视为不合格。

在每一个数据都在标准内时,一般测三个样品的试验,最后数值以平均值记录,如果五个样品,一般去掉最大和最小的数值,其余求平均值一般测试需要4-5样品。

5. 压缩永久变形的影响因素:1)橡胶配方,此决定压缩永久变形好坏的最大关键;如过氧化物硫化的EPDM压缩永久变形比硫磺硫化的小非常多,而且可以通过更高温度的测试;2)加硫程度,取决于橡胶成型三大因素-温度,时间,压力。

正常的橡胶随加硫程度的增加而压缩永久变形变小,到最低值后就开始变大,这时意味着橡胶产品开始过硫化了;特别需要说明的是硫磺硫化的NBR,EPDM等,一次加硫和二次加硫均对此影响非常大(尤其是温度);而过氧化物硫化的NBR,EPDM,一次成型的温度尤其重要,建议在180摄氏度以上,如果一次加硫不足,二次加硫的补足有限;3)测试样品的形状和测试夹具,量具和试验设备的精密性均会影响最后的结果,但不是橡胶压缩永久变形的真因。

ASTM_D395_2003压缩永久变形中文版

ASTM_D395_2003压缩永久变形中文版

编号:D 395-03橡胶性能的标准试验方法----------压缩永久变形1此项标准在固定编号B 117下发布,紧随编号的数字表示标准采纳的年度,如果是修正,数字表示最后一次修正的年度。

在括号的数字表示最后一次重申批准的年度。

上标 表示自最后一次修正或重申批准以来的编辑改动。

此项标准已被批准供美国国防部下属机构使用。

1围1.1本测试方法测试应用中会在气体或液体媒介中承受压力的橡胶。

本测试方法特别适用于在机械固定器件,减震器,封条中使用的橡胶。

本测试方法包含以下两种方法:1.2测试方法可以选择,但是应考虑用于与测试结果关联的实际情况下使用的橡胶的性质。

除非在具体的规中有其他规定,应使用测试方法B。

1.3测试方法B不适用于硬度大于90IRHD的硫化橡胶。

1.4以国际单位(SI)为单位的数值应被认为是标准。

在括号的数值起参照作用。

1.5此项标准不包括与其应用有关的所有的安全隐患。

此项标准的使用者有责任在使用前建立合适的安全健康规以及决定法规限制是否适用2 参考文件2.1 ASTM标准2:D1349 橡胶规---测试的标准温度D 3182 混合标准化合物及制备标准硫化橡胶薄片用橡胶材料、设备及工序的标准实施规程D 3183 橡胶实施规---从橡胶制品中制备试验目的用试片D 3767 橡胶的标准规程----尺寸测量D 4483 评定橡胶和炭黑制造工业试验方法标准的精度的实施规程E 145 重力对流式和强制通风式烘炉的规---------------------------------------1此测试方法属于ASTM D 11橡胶委员会的工作围,是其下属D11.10物理测试子委员会的直接责任。

目前的版本在2008.3.1批准,2008.07出版。

原始的版本在1934年批准。

上一个版本在2003年批准,编号为D395-03.2如需参照ASTM 标准,访问,或联系ASTM客户服务. 如需要《ASTM标准年鉴》的容信息,浏览ASTM的标准索引页。

橡胶压缩永久变形率测定常用标准分析与解读

橡胶压缩永久变形率测定常用标准分析与解读

橡胶压缩永久变形率测定常用标准分析与解读橡胶是一种具有良好性能和灵活性的多功能材料,由于其变形性能良好,可以用于制造适用于汽车、机器人、海洋平台和机器臂等各种复杂结构的组件。

但是,由于橡胶的变形特性受到温度、湿度和外力的影响,因此在设计过程中,精确测定橡胶的变形特性是非常必要的。

因此,本文将聚焦于橡胶压缩变形率测定,并全面分析、解读常用的测定方法和标准。

首先,介绍一般正常温度下橡胶压缩变形率测定的基本原理,包括其变形机理、测试仪器设备、测试手段、计算数据以及主要参数的确定。

其次,系统地介绍国际、国家和行业标准对橡胶压缩变形率的要求,以及根据不同标准的不同测试方法及其特点进行详细分析。

接下来,从技术角度讨论橡胶压缩变形率测定的常见问题,如变形温度、载荷变化速率、变形容量有限性等,并针对各种材料和应用场景提出改进方案。

最后,介绍橡胶压缩变形率测定可能会遇到的不确定性和误差,并从统计学的角度介绍测试结果的评估方法及其应用。

总之,本文从宏观上介绍了橡胶压缩变形率测定的基本原理、常用标准、常见问题及其解决方案,以及变形测试结果的评估方法,旨在为橡胶压缩变形率测定提供较为完整的依据和解读,为橡胶材料的研制及其应用提供有力的技术支持。

- 1 -。

压缩永久变形试验标准分析解读

压缩永久变形试验标准分析解读

作者简介:尤黎明(1993-),男,从事胶筒模具设计与胶筒制造工艺方面工作。

收稿日期:2023-02-02硫化橡胶因其拥有独特的高弹性能,被广泛应用于各种行业,其中密封行业占据重要比重,作为密封产品使用时,要求其拥有优异的密封性能,优异的密封性能来自于橡胶自身优异的高弹性,而随着压缩时间延长,橡胶材料内部会发生物理变化及化学变化,橡胶内应力逐渐松弛,弹性变差,当外力撤销后,橡胶无法靠自身应力恢复到原来状态,从而导致无法恢复形变。

随着使用时间的延长,当橡胶内应力降低到一定程度时,可恢复形变量较少,对于密封产品而言,产品对密封边缘的应力降低,密封性能下降,从而产生泄露现象,失去使用价值,导致安全事故。

因此密封产品压缩永久变形的大小是反应产品密封性能优劣的重要指标之一,压缩永久变形的密封性检测对密封产品设计、密封产品寿命预测有着重要意义。

橡胶密封产品压缩永久变形测试结果与众多因素有关,产品配方设计、加工工艺、测试温度、测试时间、试样尺寸等因素对压缩永久变形影响较为明显。

压缩永久变形是橡胶密封制品最为关注的性能之一,不同试验标准测得压缩永久变形试验结果有所不同,下面将对4个常用压缩永久变形标准进行对比与总结。

1 压缩永久变形常用标准简介(1)IS0 815-1:2019(常温或高温条件下,硫化橡胶或热塑性橡胶压缩永久变形测定):于2008年进行第三次修订,此次修订将压缩永久变形常温和高温试验方法分开,形成了ISO 815-1和ISO815-2两个标准。

本次解读最新修订版本IS0 815-1:2019,与上一版相比,主要变化如下:在第2条中更新了规范参压缩永久变形试验标准分析解读尤黎明,程文佳,刘明泰,靳浩楠,李权(中海油田服务股份有限公司油田生产事业部,天津 300450)摘要:为对橡胶密封产品压缩永久变性能进行研究,找到合适实验标准,对O 型密封圈进行压缩永久变形试验,对比分析了4种常用的压缩永久变形标准(IS0 815-1:2019,ASTM D395-2018,GB/T 7759.1-2015,GB/T 1683-2018),从实验方法、试样尺寸、试验步骤等角度进行分析,为选取合适的试验标准提供参考。

三元乙丙橡胶力学及压缩永久变形性能研究

三元乙丙橡胶力学及压缩永久变形性能研究

三元乙丙橡胶力学及压缩永久变形性能研究三元乙丙橡胶力学及压缩永久变形性能研究三元乙丙橡胶力学及压缩永久变形性能研究首先,本文对EPDM常用硫化体系进行对比考察,并对过氧化物硫化体系中助交联剂的单用及并用、常用防老剂的单用及并用对EPDM硫化特性、力学性能、耐老化性能及耐压缩永久变形性能的影响进行系统研究,以探索助交联剂的交联反应对DCP主交联作用的影响及防老剂的合理选用。

研究表明:1)、有效硫化体系及复合硫化体系力学性能较好,但耐老化及压缩永久性能差,DCP硫化体系综合性能较好。

2)、PDM、TAIC和S,可增大硫化胶的硫化速度和交联密度,并改善耐老化性能,且TAIC可改善加工性能;PDM、TAIC及适量S分别与DCP并用,可较好改善硫化胶压缩永久变形。

3)、采用单一防老剂,MB的综合性能及抗老化效果最好,硫化胶压缩永久变形最小;采用两种防老剂同份量并用,MB与NBC并用抗老化效果最好,而RD与MB并用压缩永久变形性能最好;MB与NBC不同份量并用时,其份量比为1/1时硫化胶综合性能较好,而份量比为0.5/1.5时压缩永久变形最小。

其次,本文分别从硫化温度、硫化时间、硫化工艺考察对EPDM硫化特性、力学性能、耐老化性能及耐压缩永久变形性能的影响。

研究表明:随硫化温度增大,填料间相互作用、交联密度减小,T10和T90明显缩短,压缩永久变形增大。

合理选择硫化温度和硫化时间,可使得硫化胶具有较好的综合性能,采用二段硫化能提高硫化胶的力学性能,二段硫化时间为2h时综合性能较佳。

随硫化时间和二段硫化时间增大,其压缩永久变形降低。

再次,本文研究了炭黑(N234、N330、N550及N774)、蒙脱土(DK2-OMMT、DK3-OMMT)、纳米重晶石(BaSO_4)及和纳米凹凸棒(AT)对EPDM的硫化特性、力学性能、耐老化性能及压缩永久变形性能的影响,并从粒径、结构性、微观形貌等因素考察,从而为配方筛选、新型填料的探索、纳米材料的应用提供理论和实验依据。

压缩永久变形测试标准

压缩永久变形测试标准

压缩永久变形测试标准
压缩永久变形测试标准
压缩永久变形是指物体在经受一定压力后所产生的永久变形,这里
“永久变形”指物体受到强压力后不能复原。

压缩永久变形测试标准
可以帮助我们了解材料性能,以及开发新型材料。

那么,具体的压缩
永久变形测试标准如下:
首先,要准备好测试样品。

根据不同的实验要求,测试样品可以是具
有不同直径,长度,形状和尺寸的各种材料,包括金属,塑料,玻璃,和其他材料。

第二,要确保测试条件的恒定性。

在开始测试前,首先要检查样品温度,湿度,和温度梯度,确保它们处于稳定状态,以确保测试结果的
准确性。

第三,使用合适的测试设备和工具。

实验中,将研究对象固定在测试
仪表上,调整好参数,然后向样品施加压力,并且使用传感器检测样
品的变形。

第四,统计和计算变形量。

测量完成后,要记录初始和测试后的样品
尺寸,以便计算出样品的压缩永久变形量。

最后,报告测试结果。

经过计算后,可以得出样品的压缩永久变形极限,测试报告中包含了样品的变形量和试验条件的记录,便于对样品
的性能做出准确的判断。

以上就是压缩永久变形测试标准,此标准适用于各种材料的检测,为
我们了解这些材料的性能,以及开发新型材料提供了参考依据。

在实际测试时,还要注意安全措施,以确保测试的顺利进行。

橡胶压缩永久变形测试标准

橡胶压缩永久变形测试标准

橡胶压缩永久变形测试标准橡胶压缩永久变形测试是评估橡胶材料压缩性能的一种常见试验方法。

本文将介绍橡胶压缩永久变形测试的相关参考内容和标准,帮助读者了解该测试方法的基本原理和操作规范。

一、ASTM D395-03标准ASTM D395-03是美国材料和试验协会(ASTM)发布的一项标准,名为"Standard Test Methods for Rubber Property—Compression Set",中文译为《橡胶性能试验方法—压缩永久变形》。

该标准规定了橡胶材料在一定压缩载荷下的压缩永久变形测试方法。

ASTM D395-03标准主要包括以下内容:1. 试样的准备:包括试样的尺寸、形状和制备方法;2. 试验设备和仪器:包括压缩试验机、压缩板和计算压缩变形的测量仪器;3. 试验过程:包括试样的压缩速率、保持时间和温度条件等;4. 数据处理和结果计算:包括压缩永久变形的计算方法和结果的统计分析。

二、ISO 815:2008标准ISO 815:2008是国际标准化组织(ISO)发布的一项标准,名为"Rubber, vulcanized or thermoplastic -- Determination of compression set at ambient, elevated or low temperatures",中文译为《硫化橡胶或热塑性橡胶——在常温、高温或低温下测定压缩永久变形》。

该标准主要用于测定硫化橡胶或热塑性橡胶在不同温度条件下的压缩永久变形性能。

ISO 815:2008标准包括以下内容:1. 试样的准备:包括试样的尺寸和形状;2. 试验设备和仪器:包括压缩试验机、压缩板和计算压缩变形的测量仪器;3. 试验过程:包括试样的压缩速率、保持时间和温度条件等;4. 数据处理和结果计算:包括压缩永久变形的计算方法和结果的报告要求。

三、GB/T 7759-2009标准GB/T 7759-2009是中国国家标准,名为"硫化橡胶及热塑性橡胶常温、高温和低温压缩永久变形试验方法"。

橡胶压缩永久变形标准

橡胶压缩永久变形标准

橡胶压缩永久变形标准摘要:1.橡胶压缩永久变形率的定义2.常用橡胶压缩永久变形率测定标准简介3.试样尺寸、试验步骤和试验数据处理对测试结果的影响4.GB/T 7759-1996 标准在某些条款规定方面的不当之处及原因分析5.结论与展望正文:橡胶压缩永久变形率是指橡胶材料在受到一定的压缩力后,其形状发生的永久性变化。

这一指标是衡量橡胶材料性能的重要参数,对于研究橡胶材料的压缩性能和设计橡胶制品具有重要意义。

在实际应用中,橡胶压缩永久变形率的测定通常需要参考相关标准进行。

本文将对橡胶压缩永久变形率测定常用的标准试验方法进行对比分析,并探讨试样尺寸、试验步骤和试验数据处理等对测试结果影响较大的测试因素及其对测试结果的影响规律。

一、常用橡胶压缩永久变形率测定标准简介目前,橡胶压缩永久变形率测定常用的标准试验方法有4 种,分别是:1.ISO 815-1:20082.ASTM D395-033.JIS K6262:20134.GB/T 7759-1996这些标准分别代表了国际、美国、日本和中国的橡胶压缩永久变形率测试方法。

虽然这些标准在试样尺寸、试验步骤和试验数据处理等方面存在一定的差异,但它们的核心思想和测试原理基本一致。

二、试样尺寸、试验步骤和试验数据处理对测试结果的影响试样尺寸、试验步骤和试验数据处理是影响橡胶压缩永久变形率测试结果的主要因素。

试样尺寸的大小直接影响到测试结果的可靠性和准确性。

不同的标准对试样尺寸的规定也不尽相同,因此在进行测试时需要严格按照相应标准的要求进行。

试验步骤是影响测试结果的另一个重要因素。

在试验过程中,需要保证橡胶试样在恒定的温度和压力下进行测试,以确保测试结果的准确性。

此外,试验过程中需要遵循标准的操作步骤,以避免因操作不当导致的测试结果偏差。

试验数据处理是对测试结果进行分析和解释的关键环节。

在数据处理过程中,需要对测试数据进行合理的修约和处理,以确保测试结果的准确性和可靠性。

ASTM D395-2003压缩永久变形中文版

ASTM D395-2003压缩永久变形中文版

编号:D 395-03橡胶性能的标准试验方法----------压缩永久变形1此项标准在固定编号B 117下发布,紧随编号的数字表示标准采纳的年度,如果是修正,数字表示最后一次修正的年度。

在括号内的数字表示最后一次重申批准的年度。

上标 表示自最后一次修正或重申批准以来的编辑改动。

此项标准已被批准供美国国防部下属机构使用。

1范围1.1本测试方法测试应用中会在气体或液体媒介中承受压力的橡胶。

本测试方法特别适用于在机械固定器件,减震器,封条中使用的橡胶。

本测试方法包含以下两种方法:1.2测试方法可以选择,但是应考虑用于与测试结果关联的实际情况下使用的橡胶的性质。

除非在具体的规范中有其他规定,应使用测试方法B。

1.3测试方法B不适用于硬度大于90IRHD的硫化橡胶。

1.4以国际单位(SI)为单位的数值应被认为是标准。

在括号内的数值起参照作用。

1.5此项标准不包括与其应用有关的所有的安全隐患。

此项标准的使用者有责任在使用前建立合适的安全健康规范以及决定法规限制是否适用2 参考文件2.1 ASTM标准2:D1349 橡胶规范---测试的标准温度D 3182 混合标准化合物及制备标准硫化橡胶薄片用橡胶材料、设备及工序的标准实施规程D 3183 橡胶实施规范---从橡胶制品中制备试验目的用试片D 3767 橡胶的标准规程----尺寸测量D 4483 评定橡胶和炭黑制造工业试验方法标准的精度的实施规程E 145 重力对流式和强制通风式烘炉的规范---------------------------------------1此测试方法属于ASTM D 11橡胶委员会的工作范围,是其下属D11.10物理测试子委员会的直接责任。

目前的版本在2008.3.1批准,2008.07出版。

原始的版本在1934年批准。

上一个版本在2003年批准,编号为D395-03.2如需参照ASTM 标准,访问ASTM网站,或联系ASTM客户服务Service@. 如需要《ASTM标准年鉴》的内容信息,浏览ASTM网站的标准索引页。

橡胶材料的耐压缩变形能力测试方法

橡胶材料的耐压缩变形能力测试方法

橡胶材料的耐压缩变形能力测试方法橡胶材料是一种常用的弹性材料,其耐压缩变形能力是评估其性能的重要指标之一。

通过对橡胶材料的耐压缩变形能力进行测试,可以有效地评估其在承受压力时的可靠性和耐久性。

本文将介绍几种常见的橡胶材料耐压缩变形能力测试方法。

一、压缩性能测试装置为了对橡胶材料的压缩性能进行测试,需要使用专门的测试装置。

一般而言,该装置由压力源、压力传感器、压力控制系统和压缩样本夹具等组成。

其中,压力源通常采用液压或气压形式,通过调节压力传感器和压力控制系统,可以精确控制施加在橡胶样本上的压缩力。

二、静态压缩测试方法静态压缩测试是最常见的橡胶材料耐压缩变形能力测试方法之一。

测试步骤如下:1. 样本制备:按照标准规定的尺寸和几何形状,将橡胶样本切割或制备成所需的形状。

2. 样本安装:将制备好的橡胶样本放入压缩样本夹具中,并确保样本与夹具接触牢固。

3. 施加压力:通过测试装置施加预定的压缩力,在保持恒定的温度和湿度条件下,让样本在压缩状态下保持一段时间。

4. 测试结果记录:记录橡胶样本在不同压缩力下的变形情况,包括样本的压缩率、变形应力等数据。

5. 分析结果:根据测试结果,可对橡胶材料的耐压缩能力进行评估和比较分析。

三、循环压缩测试方法循环压缩测试是测试橡胶材料耐久性和可持续性的方法之一。

该测试方法模拟了实际使用中橡胶材料所承受的周期性压力。

具体步骤如下:1. 样本制备:与静态压缩测试相同。

2. 样本安装:与静态压缩测试相同。

3. 施加循环压力:通过测试装置施加周期性循环压力,其中包括峰值压力、保持时间和循环次数等参数。

4. 测试结果记录:记录每个循环周期中橡胶样本的变形情况,包括样本的压缩率、应力和恢复性等数据。

5. 分析结果:根据测试结果,可评估橡胶材料在循环压力下的耐久性和恢复性能。

四、动态压缩测试方法动态压缩测试是模拟橡胶材料在实际使用中受到的动态压力作用的方法,可以更真实地评估橡胶材料的性能。

橡胶材料的拉伸和压缩性能研究

橡胶材料的拉伸和压缩性能研究

橡胶材料的拉伸和压缩性能研究橡胶作为一种重要的弹性材料,在各个领域中都有广泛的应用。

其中,橡胶材料的拉伸和压缩性能是评价其机械性能的关键指标。

本文将对橡胶材料的拉伸和压缩性能进行探究。

一、拉伸性能研究拉伸性能是指橡胶材料在外力作用下的延展性和抗拉强度。

首先,我们来讨论橡胶材料的延展性。

橡胶在拉伸过程中具有明显的拉伸特性,当外力作用于橡胶材料上时,橡胶会随之延展。

这种延展性的大小与橡胶的材料特性和结构有关。

一般来说,硬度较低的橡胶具有更好的延展性。

此外,橡胶材料的延展性还与温度有关,温度越高,橡胶的延展性越好。

接下来,我们来研究橡胶材料的抗拉强度。

抗拉强度即橡胶材料在拉伸过程中能承受的最大张力。

抗拉强度是评价橡胶材料抗拉能力的指标之一,通常用来衡量橡胶材料在外力作用下的稳定性和耐久性。

二、压缩性能研究压缩性能是指橡胶材料在外力作用下的压缩变形和弹性恢复能力。

与拉伸性能不同,压缩性能主要涉及橡胶材料在受力情况下的变形和回弹过程。

橡胶材料在受压时具有很好的变形能力,可以承受较大的压力而不会出现永久性变形。

压缩弹性模量是评价橡胶材料压缩性能的一项重要指标。

压缩弹性模量反映了橡胶材料在受压过程中的变形能力和回弹能力。

一般来说,压缩弹性模量越小,橡胶材料的弹性恢复能力越好。

三、拉伸与压缩性能的相关性研究橡胶材料的拉伸性能和压缩性能之间存在一定的相关性。

通常情况下,延展性较好的橡胶材料其压缩性能也比较好。

这是因为延展性好的橡胶材料在受力过程中具有较大的变形能力,能够更好地承受外力的作用。

同时,延展性好的橡胶材料回弹能力也较好,能够较好地恢复到初始状态。

然而,并非所有延展性好的橡胶材料都具有良好的压缩性能,因为压缩性能还与材料的组成和结构有关。

有些橡胶材料在拉伸过程中延展性好,但在受压过程中会出现易变形和永久性变形的问题。

为了更深入地研究拉伸和压缩性能之间的关系,需要考虑橡胶材料的材料特性、结构设计和工艺参数等因素,并进行系统的实验研究。

橡胶压缩国标

橡胶压缩国标

橡胶压缩国标1. 引言橡胶是一种重要的工业原料,广泛应用于汽车制造、建筑、电子产品等领域。

橡胶的质量直接影响到相关产品的性能和安全性。

为了确保橡胶产品的质量和可靠性,各国都制定了相应的国家标准。

本文将介绍橡胶压缩国标,包括其定义、制定过程、标准内容等方面。

2. 定义橡胶压缩国标是指针对橡胶材料进行压缩测试的国家标准。

该标准规定了测试方法、设备要求、试样制备等内容,旨在评估橡胶材料在不同温度和压力下的变形特性。

3. 制定过程3.1 标准委员会成立制定橡胶压缩国标的第一步是成立一个专门负责该领域标准制定的委员会。

委员会由相关行业专家、企业代表、科研机构代表等组成,他们通过集体讨论和投票决定最终的标准内容。

3.2 调研和分析标准委员会在制定标准之前,需要进行大量的调研和分析工作。

他们会收集国内外相关行业的标准、技术规范、研究报告等资料,并进行综合分析。

通过对市场需求、技术发展趋势等的了解,委员会可以确定标准的制定方向和重点。

3.3 制定草案在调研和分析的基础上,标准委员会开始制定橡胶压缩国标的草案。

草案包括标准编号、引言、范围、术语和定义、试验方法、结果判定等内容。

委员会成员根据各自的专业知识和经验,在草案中提出自己的意见和建议。

3.4 内部审查完成草案后,标准委员会进行内部审查。

审查过程中,委员会成员对草案中的各项内容进行仔细检查,确保其科学性、合理性和可操作性。

同时,他们也对可能存在的问题或争议进行讨论,并做出相应的修改。

3.5 公开征求意见完成内部审查后,标准委员会将草案公开征求意见。

他们可以通过发布公告、组织座谈会、邀请专家评审等方式,向社会各界征求对草案的意见和建议。

这些意见和建议将被认真研究和分析,对草案进行进一步的修订。

3.6 最终确定在经过多次修订和完善后,标准委员会最终确定橡胶压缩国标的正式版本。

该版本将由相关部门进行审核,并最终发布实施。

4. 标准内容橡胶压缩国标主要包括以下内容: - 标准编号和引言:用于唯一标识该国标,并对其背景和目的进行简要介绍。

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。

关于橡胶压缩永久变形测试标准的文章橡胶压缩永久变形测试标准
橡胶是一种常见的材料,具有优异的弹性和耐磨性。

然而,长期使用或受到压力的橡胶制品可能会出现压缩永久
变形的问题。

为了确保橡胶制品的质量和可靠性,压缩永
久变形测试标准被广泛应用于橡胶制品的生产和质量控制
过程中。

压缩永久变形是指橡胶在受到持续压力后无法完全恢复
到原始形状和尺寸的现象。

这种变形可能会导致橡胶制品
在使用过程中失去原有的弹性和功能,从而影响其性能和
寿命。

因此,对于需要长期承受压力或重复加载的橡胶制品,如密封圈、垫片、管道等,进行压缩永久变形测试是
至关重要的。

在进行压缩永久变形测试时,需要遵循一系列标准和规范。

其中最常用的标准是ASTM D395-03(2018)《Standard Test Methods for Rubber Property—Compression Set》。

该标准规定了测试样品的准备、测试条件、测试方法和结
果评估等方面的要求。

首先,测试样品的准备非常重要。

样品应该是充分固化
和成型的橡胶制品,具有一定的尺寸和形状。

在进行测试
之前,样品应该经过一段时间的恢复,以确保其处于稳定
状态。

其次,测试条件也需要严格控制。

根据ASTM D395-
03(2018)标准,测试温度通常为70℃或100℃,并且需要
在一定的压力下进行。

这些条件可以根据实际应用需求进
行调整。

在进行测试时,需要将样品置于特定的试验装置中,并
施加一定的压力。

通常情况下,压力会持续施加一段时间,然后释放并观察样品恢复程度。

通过测量样品在释放后的
高度或厚度变化,可以计算出压缩永久变形率。

最后,在得到测试结果后,需要对其进行评估和分析。

通常情况下,压缩永久变形率应该控制在一定范围内,并与产品设计要求相匹配。

如果测试结果超出了规定范围,则需要对橡胶材料或制造工艺进行调整。

总之,橡胶压缩永久变形测试标准是确保橡胶制品质量和可靠性的重要手段。

通过遵循相关标准和规范,可以有效评估橡胶材料的性能,并为产品设计和制造提供参考依据。

这将有助于提高橡胶制品的使用寿命和性能,满足用户的需求。

相关文档
最新文档