成年大鼠脑透析液和血清中褪黑素含量的测定
孟涛;郑志竑;叶婧;林玲
【摘要】目的:采用高效液相色谱(HPLC)‐荧光法和酶联免疫(ELISA)法测定成年大鼠脑微透析液及血清中的褪黑素(MLT)含量,比较2种检测方法的灵敏性、准确性及可靠性。方法将微透析引导管置入正常成年大鼠纹状体内1周后,采用脑微透析技术动态收集2:00,6:00,10:00,14:00,18:00和22:00的透析液,并于次日22:00行尾静脉取血;用HPLC‐荧光法和ELISA法分别测定透析液及血清样本中的MLT水平。结果HPLC‐荧光法测定MLT检出限为0.1 pg/mL ,在1.0~500 pg/mL之间呈线性关系,回归方程 y=10446 x +8603.5(r=0.9987,P< 0.001),日间、日内精密度均<6%;ELISA 法检出限为5.92 pg/mL ,检测线性范围为12.35~1000 pg/mL。用HPLC‐荧光法测定2:00,6:00,10:00,14:00,18:00和22:00大鼠脑透析液中的MLT含量分别为(32.99 ± 2.22),(18.25 ± 2.27),(4.03 ± 1.70),(2.92 ± 1.08),(12.08 ± 1.77)和(21.83 ± 2.04)pg/m L ,具有明显的昼夜节律性,探针回收率为12%~13%;22:00血清中的MLT含量为(82.62 ± 8.09)pg/mL ,提取回收率>90%,且血清与透析液中的MLT水平呈高度正相关(r=0.987,P<0.001);用ELISA法测定血清中的MLT含量为(81.31±9.62)pg/mL ,且用2种方法测定血清中MLT差别无统计学意义(P=0.115)。结论用HPLC‐荧光法测定大鼠脑微透析液和血清中的MLT 含量具有灵敏、快速、准确、重复性好等特点,且HPLC‐荧光法和ELISA法测定血清中的MLT具有较好的一致性。%Objective To determine brain dialysate and serum melatonin (MLT ) levels in adult rats by high‐performance liquid chromatography (HPLC) with fluorescence
detection and enzyme‐linked immunosorbent a ssay (ELISA) ,and compare the sensitivity ,accuracy and reliability of the two detection methods . Methods The guide tube was implanted into the striatum of healthy adult rats . One week af‐ter implantation ,dialysate samples were collected at
2 :00 ,6:00 ,10 :00 ,14:00 ,18 :00 ,22 :00 h in stria‐tum by brain microdialysis . Blood samples were collected from the rat tail vein at 22 :00 h on the follow‐ing day . HPLC with fluorescence detection and ELISA were applied to detect dialysate and serum MLT levels . Results The lowest concentration of MLT that can be detected by HPLC with fluorescence de‐tection was 0 .1 pg/mL . There was a regressive equation y=10 446 x+8 60
3 .5 (r=0 .998 7 ,P<0 .05) in linear range 1 .0~500 pg/mL of MLT
levels ,with an intra‐assay and inter‐assay of below 6% . The detection limit of MLT by ELISA was 5 .92 pg/mL ,and the detection range was 12 .35~1 000 pg/mL . The mean concentrations of dialysate MLT that showed a distinct fluctuation were (in pg/mL )
(32.99±2.22),(18.25±2.27),(4.03±1.70),(2.92±1.08),(12.08±1.77),(21.83±2.04 )at2:00, 6 :00 ,10:00 ,14:00 ,18:00 and 22 :00 h by HPLC with fluorescence detection . The recovery of the probes was from 12% to 13% . Besides ,the mean concentration of serum MLT was (82 .62 ± 8 .09) pg/mL by t his method . The extraction recovery of serum MLT was more than 90% . There was a distinct positive correlation between serum MLT and dialysate MLT levels at 22:00 pm ,and pearson’s correlation coeffi‐cient between the two was 0 .987 at the 0 .001 level (t wo‐tailed) .In addition ,the mean concentration of serum MLT was (81 .31 ± 9 .62) pg/mL by ELISA . There