贝朗DIALOG+透析机水路设计概览

合集下载
  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
• 透析器内没有压力突变 超滤平稳并且避免了平衡腔切换时产生反超
• 透析液持续不断(包括液流切换时刻) 维持住弥散压力可带来更好的治疗效果(K*t/V值)
• 对膜片、阀门和泵都有保护作用 延长配件寿命
Dialog+ Wettbewerb
17/08/2020
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
17/08/2020
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Authors:
3
Guenter Nissen
Heiko Riemenschneider
Water Upline Tank 水箱
填充水箱,整个压力通路通畅
高水位限位圈 高水位触点 关闭进水阀门 中水位(低于高位触点) 打开进水阀门
经除气和加热后的 反渗水
反渗水
注意:浮球上的 三角标识指示 装配方向
Authors:
11
Guenter Nissen
Heiko Riemenschneider
Advantages of the Dialog Balance Chamber Control Dialog 改良的平衡腔控制系统的优点
• 持续监测平衡腔膜片位置 持续控制,腔体内压力稳定,阀门切换对流速影响小
Authors:
7
Guenter Nissen
Heiko Riemenschneider
Main Flow / Bypass 主水路/旁路
17/08/2020
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Authors:
4
Guenter Nissen
Heiko Riemenschneider
Water Upline Tank水箱
除气和加热后的 反渗水
部分溢流至3/2部分 经除气和加热后 的反渗水
反渗水进水
进水减压阀DMV 控制进水箱反渗水 压力不高于1.3Bar
去配比透析液阶段
去除气和加热 节段
17/08/2020
Dialog+ Flow Diagrams水路设计概览
贝朗爱敦(上海)贸易有限公司 维修部 技术经理 崔波 2010年4月8日
Flow Diagram
TSHE T
水路图可分成六个节段:
1、进水节段 2、除气和加热节段 3、配比透析液节段 4、平衡腔容量控制节段 5、超滤控制节段 6、冲洗桥节段
17/08/2020
透析器
17/08/2020
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Authors:
9
Guenter Nissen
Heiko Riemenschneider
Balance Chamber Phase 1 平衡腔容量平衡 1阶段
Authors:
13
Guenter Nissen
Heiko Riemenschneider
Flow Diagram Disinfection消毒循环回路
TSHE T
VD:消毒阀 VZ:循环阀
water inlet dialysate outlet
17/08/2020
通大气 软管
消毒液
Authors:
17/08/2020
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
低水位触点 自动停止加热, 泵停转
17/08/2020
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Authors:
10
Guenter Nissen
Heiko Riemenschneider
Balance Chamber Phase 2 平衡腔容量平衡 2阶段
正压泵
平衡腔
空气分离室
BLD
负压泵
PDA
超滤泵
17/08/2020
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Authors:
2
Guenter Nissen
Heiko Riemenschneider
Water Inlet, Degassing and Heating 进水节段,除气和加热节段
TSHE T
进水口 排水口
进水部分 DMV:减压阀 RVVB:进水电磁阀 WT:热交换器 VB:水箱 NSVB:水箱液位传感器 除气部分 RVE:除气比例阀 PE:除气压力传感器 EK:除气室 EP:除气泵 加热部分 TSHE:加热前温度传感器 H:加热棒 TSH:加热棒高温保护 TSE:加热后温度传感器
Authors:
5
Guenter Nissen
Heiko Riemenschneider
Dialysate Preparation 配比透析液节段
TSHE T
进水口
17/08/2020
排水口
BICP:B液泵 RVB:B液单向阀 TSBIC:B液温度传感器 BICLF:B液电导度传感器 KP:A液泵 RVK:A液单向阀 ENDLF:总电导度传感器 ENDLF_S: ENDLF的检测信号 TSD:透析液温度传感器 TSD_S:TSD的检测信号 FPE:正压泵 RVFPE:FPE的高压保护阀
用后透析液 + 超滤液
VABK:废液出平衡腔电磁阀
17/08/2020
ห้องสมุดไป่ตู้
Training document, no update service.
This document may not be copied, partly or whole, or made be accessible to third parties without
正压泵
平衡腔 DDE:透析液节流阀 RVDA:废液节流阀 PDA:废液压力传感器 BLD:漏血传感器
LA (空气分离室)
BLD
FPA(负压泵)
PDA
UFP(超滤泵)
17/08/2020
Training document, no update service. This document may not be copied, partly or whole, or made be accessible to third parties without our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Authors:
6
Guenter Nissen
Heiko Riemenschneider
Main Flow / Bypass 主水路/旁路
Authors:
12
Guenter Nissen
Heiko Riemenschneider
Main Flow / Bypass 主水路/旁路(安全性设计)
净的透析液
通向除气部分
废液: 透析液+超滤液
VBP:旁路阀 VDE:透析液出机器电磁阀 VDA:废液回机器电磁阀 LA:空气分离室 VLA:空气分离阀
Authors:
8
Guenter Nissen
Heiko Riemenschneider
Balance Chamber平衡腔
透析液
dialyser 透析器
FPE:正压泵
FPA:负压泵
BK1/2:平衡腔室1/2
MSBK1/2:平衡腔膜位移传感器
VEBK:透析液进平衡腔电磁阀
VDEBK:透析液出平衡腔电磁阀 VDABK:废液进平衡腔电磁阀
Bic
acidic
concentrate concentrate
B液
A液
Training document, no update service.
This document may not be copied, partly or whole, or made be accessible to third parties without
相关文档