Ultrasound
A Deep Thermal & Non-thermal
Mechanical Modality
CAI BIN
▪对于一位前臂开刀半年后产生黏连性疤痕组织的病患,下列超声波治疗参数的组合,何着最为合适?
▪A. 間歇性:1MHZ ;0.5watt/cm2
▪B. 間歇性:1MHZ ;1-2 watt/cm2
▪C. 連續性:3MHZ ;0.5 watt/cm2
▪D. 連續性:3MHZ ;1-2 watt/cm2
纲要
▪基础部分▪临床部分
What is Ultrasound?
▪Located in the Acoustical Spectrum
▪May be used for diagnostic imaging, therapeutic tissue healing, or tissue destruction
▪Thermal & Non-thermal effects
▪We use it for therapeutic effects
▪Can deliver medicine to subcutaneous tissues (phonophoresis)
超声波
▪超声波是指频率在20KHz以上,不能引起正常人听觉反应(16~20KHz)的机械振动波
▪Therapeutic ultrasound waves range from 750,000 to 3,000,000 Hz (0.75 to 3 MHz)▪近年多采用1MHz、3MHz超声
物理特性▪超声波与声波的本质
相同,都是物体的机
械振动在弹性介质中
传播所形成的机械振
动波。
▪在传播时产生一种疏
密交替的弹性纵波,
具有一定的方向性;
Frequency
▪Frequency:number of times an event occurs in 1 second; expressed in Hertz or pulses per second
•Hertz:cycles per second
•Megahertz:1,000,000 cycles per second
▪In the U.S., mainly use ultrasound frequencies of 1,
2 and
3 MHz
▪1 = low frequency; 3 = high frequency
▪↓frequency = ↑depth of penetration
▪↑frequency = sound waves are absorbed in more superficial tissues (3 MHz)
Velocity
▪The speed of sound wave is directly related to the density (↑velocity = ↑density)
▪Denser & more rigid materials have a higher velocity of transmission
▪At 1 MHz, sound travels through soft tissue @ 1540 m/sec and 4000 m/sec through compact bone
物理特性
超声波的传播速度与介质的特性有关,而与声波的频率无关;
介質的密度越大,則音波的傳遞速度就愈
快。例如:空氣(344m/s)<水
(1410m/s)<肌肉、軟組織(1540m/s)
<骨骼(4000m/s)
Influences on the Transmission of
Energy
▪Reflection–occurs when the wave can’t pass through the next density
▪Refraction–is the bending of waves as a result of a change in the speed of a wave as it enters a medium with a different density
▪Absorption–occurs by the tissue collecting the wave’s energy
Attenuation
▪Decrease in a wave’s intensity resulting from absorption, reflection, & refraction
▪US penetrates through tissue high in water content & is absorbed in dense tissues high in protein
▪the half value depth
•The average 1/2 value depth of 3MHz ultrasound is
taken at 2.5cm and that of 1MHz ultrasound as 4.0 cm
组织对超声波的吸收能力
▪組織中的含水量如果愈高,則能量的吸收就越少
▪人體組織中,蛋白質的含量愈高則能量的吸收就愈多
▪對超音波的吸收力:骨頭>軟骨>肌腱>皮膚>肌肉>神經>脂肪>血液
Attenuation: Acoustic Impedance ▪Determines amount of US energy reflected at tissue interfaces
•If acoustic impedance of the 2 materials forming the
interface is the same, all sound will be transmitted
•The larger the difference, the more energy is reflected & the less energy that can enter the 2nd medium
▪US passing through air = almost all reflected (99%)▪US through fat = 1% reflected
▪Both reflected/refracted @ m. interface
▪Soft-tissue: bone interfaced = much reflected