LI Jun, ZHANG Xi-zhi, TAN Fei (Department of Radiation Oncology , The People's Hospital of Subei , Yangzhou Jiangsu 225001, China )
Abstr act: objective : To measure and calculate absorbed dose of high energy X-ray and electron beam for medical linear ac- celerator ,assurance the accuracy of the output dose .Methods : Apply water phantom for measured medium ,measure and cal- culate absorbed dose by methods of IAEA 277 report and principle of dosimetry .Result : The errors between the measured absorbed dose and standard absorbed dose in the effective point of measurement were within 0.17%-0.54%.less than 2.0% of national standard .Conclusion : The measurement results of absorbed dose of high energy X-ray and electron beam were satis- fied with clinical dosimetry regulation . Key wor ds: high energy; X- ray; electron beam; absorbed dose; measurement; calculation
D( dmax) =100 cGy 时 , 剂 量 仪 读 数 M=100/1.105=90.50
R,这样可以直观的通过剂量仪读数 M 反映最大剂量
深度处 dmax 的吸收剂量情况, 平时检测时不再推算, 便可知道校准时误差情况.同理, 对于 15 MV X 线, 水
深为 10 cm 处的 PDD=0.7707,Cx =1.214, 剂量仪校准 时读数 M=82.37 R.
VARIAN Clinac 23EX 直线加速器高能 X 线有 6 MV 和 15 MV 两种。测量吸收剂量的条件为:标准源
中国医学物理学杂志 第 25 卷 第 2 期 2008 年 3 月
- 565 -
皮距 SSD=100 cm,水模体表面照射野 Field=10 cm×10
cm, 电离室的有效测量点在射线束轴上距水模体表
度 dc 取各能量电子束的最大剂量深度 dmax,同理, 高能
电子束在水中有效测量点处的吸收剂 DW 仍为:
Dw=M·Nc·(W/e)·(Kat·t km)·SW, ai·r
Pu·Pcel×2.58×10- 4×100
( 2)
公式( 2) 中除 SW, air 和 Pu 数值不同于公式(1)外, 其它均 相同, 下面就推导 SW, air 和 Pu 做简单描述, 具体推导过 程如下:
( 1)
对于公式( 1) 而言, Dw 的单位为 cGy;M 为标准剂 量仪读数, 单位为伦琴( R) ; Nc 为剂量仪校准因子, 每 年 剂 量 仪 送 检 一 次 , 由 国 家 标 准 实 验 室 给 出 , Nc=
1.000; W/e 为在空气中形成每对离子 (其电荷为 1 个
电子的电荷)所消耗的平均能量, 为一常数 W/e=33.97
J/C;Katt 为在校准电离室时, 电离室及平衡帽对校准辐 射 ( 一般为 60Co 的 ! 射线) 的吸收和散射的修正; Km
为电离室室壁及平衡帽的材料对校准辐射空气等效
不充分而引起的修正, 这里 Kat·t Km 数值[3]由剂量仪厂 家 给 出 Kat·t Km=0.973;Sw,air 为 校 准 深 度 水 对 空 气 的 平 均阻止本领比, 其数值由辐射质剂量比 D20/D10 或组 织模体比 TPR1020 导出, 对 6 MV X 线, 由于D20/D10=
收稿日期: 2007- 07- 17 作者简介: 李军( 1975- ) , 男, 陕西周至人, 江苏省苏北人民医院放疗科
物理师兼工程师, 南京大学物理系在读硕士, 主要从事临床 放 疗 物 理 和 放 疗 设 备 维 修 。 Tel: 0514- 87937554, E- mail: liajun.2007@yahoo.com.cn。
满足临床剂量学的要求。
关键词: 高能; X 线; 电子束; 吸收剂量; 测量; 计算
中图分类号: R 814.2
文献标识码: A
文章编号: 1005- 202X( 2008) 02- 0564- 03
Measur ement and Calculation of Absor bed Dose for High Ener gy X- r ay and Electr on Beam
2. 结果
2.1 高能 X 线吸收剂量的测量与计算结果 当机器出束 100 MU 时, M 为剂量仪读数, M 为剂
量仪平均读数,校准误差=[ ( Cx·M- 100) /100] ×100%, 具体在校准深度处的吸收剂量测量与计算结果见表2。 2.2 高能电子束吸收剂量的测量与计算结果
对于不同能量的高能电子束在最大剂量深度 dmax 处校准时的吸收剂量而言, 同样当机器出束 100 MU 时, 校准误差 =[ ( CE·M- 100) /100] ×100%, 具体的测 量与计算结果见表 3。
( 1) 由实际测出的电离量半值深度 DJ50,确定电子 束在水模体表面的平均能量是 ΕS, 再由 ΕS 和最大剂 量深度 dmax 确定 SW, air。
( 2) 确定 ΕS 后, 查表得出电子束在水中的实际射 程 Rp , 再由公式 Ez=ΕS( 1- dmax/Rp) 确定电子束在水模 体中不同深度处的平均能量 Εz。
确定好高能电子束在水中的吸收剂量公式( 2) 中 的各参数值后, 令 CE=N·c (W/e)·( Kat·t km)·SW,ai·r Pu·Pcel× (2.58×10-4)×100,即 Dw= CE·M , CE 可直接求出, 确定 CE 后,当 Dw=100 cGy 时, 剂量仪读数 M=100/ CE。
1.2.2 高能电子束吸收剂量的测量方法
VARIAN Clinac 23EX 直线加速器高能电子束有
4 MeV, 6 MeV, 9 MeV, 12 MeV, 16 MeV, 20 MeV 六
档能量。测量吸收剂量的条件为: 标准源皮距 SSD=
100 cm,照射野 Field 为 15 cm×15 cm 限 光 筒,校 准 深
表 1 电子束各参数的数值
Tab.1 The value of par ameter for electr on beam
E0
Βιβλιοθήκη Baidu
dmax
D50J
ES
RP
EZ
Sw,air
PU
(MeV) (cm) (cm) (MeV) (cm) (MeV)
4 0.6 1.50 3.03 1.52 1.83 1.103 0.952
1. 材料与方法
1.1 材料 美 国 VARIAN Clinac 23 EX 直 线 加 速 器, 德 国
PTW 公司的 UNIDOS E 型剂量仪和 TW30013 0.6cc 指型防水电离室, 30 cm×30 cm×30 cm 四支点可调水 箱, 20 m 长 TNC 三同轴电缆, 并且剂量仪, 电离室及 TNC 电缆均每年送国家标准实验室校准。 1.2 方法 1.2.1 高能 X 线吸收剂量测量方法
面的深度为校准深度 dc, 对 6 MV X 线而言, dc=5 cm;
对 15 MV X 线 而 言 dc=10 cm. 每 次 加 速 器 出 束 100
MU, 重复测量三次, 取其平均值。高能 X 线在水中有
效测量点处的吸收剂量 DW 为:
DW=M·NC·( W/e)·(Kat·t km)·
SW, ai·r Pu·Pcel×2.58×10-4×100
0.58 而知 SW, air=1.199; 而对 15MV X 线, 由于D20/D10=
0.65 而知 SW,air=1.099; Pu 为电离室扰动因子, 其数值经 查图[4]可得出 Pu=1.001;Pcel 为 电 离 室 中 心 电 极 影 响 因
子, 由生产厂家给出 Pcel=0.998.另外由于公式( 1) 两边 的单位的不同, 这里 1R=2.58×10-4 C/kg,而 1J/kg=100
- 564 - 2008 年 3 月 第 25 卷第 2 期
中国医学物理学杂志 Chinese J our nal of Medical Physics
高能 X 线和电子束吸收剂量的测量与计算
Mar.,2008 Vol.25. No.2
李 军, 张西志, 谭 飞 ( 江苏省苏北人民医院 放疗科, 江苏 扬州 225002)
前言
高能 X 线和电子束进入人体组织后, 通过和人 体组织中的原子相互作用而传递电离辐射的部分或 全部能量,人体组织吸收电离辐射后, 会发生一系列 的物理, 化学, 生物学变化。最后导致组织的生物学损 伤, 即生物效应。生物效应的大小程度正比于组织中 吸收的电离辐射的能量。因此, 确切的了解组织中所 吸收的电离辐射能量,对评估放射治疗的效果和它的 副作用是极其重要的, 他的精确确定时进行放射治疗 的最基本的物理要素[1] , 世界卫生组织( WHO) 和国际 原子能机构( IAEA) 联合组织进行的世界范围的剂量 对比显示,当前放射治疗中约有 15%的患者接受的剂
量不准确[2], 由此可见, 加强吸收剂量测量的规范和测 量的准确性, 不仅有着重要的临床意义, 并且也是医 院放疗物理师的基本任务之一。
目前有多种方法用于吸收剂量的测量, 但在临床 上大多用指型电离室在水模体中测量, 并使用 IAEA 277 报告方法对高能 X 线和电子束进行测量与计算, 使用此方法简便, 易行, 快速而精度很高, 因此对临床 物理师有着积极的意义。
( 3) 确定 Εz 后再由电离室内腔半径 r=3.05 mm,用 内差法及查表确定电离室扰动因子 Pu。
( 4) 通 过 以 上 各 步 骤 确 定 SW, air 和 Pu 的 值 后 即 可 确定公式( 2) 中各个参数的确定值。E0 为电子束初始 能量即标称能量, 再确定公式( 2) 中的各个参数值以 后, 可导出 DW 值, 具体各参数推导值见表 1。
摘要: 目的: 对医用直线加速器高能 X 线和电子束的吸收剂量进行测量与计算, 保证输出剂量的准确性。方法: 以水模
体作为介质, 用 IAEA 277 报告及剂量学原理进行吸收剂量的测量与计算。结果: 在有效测量点的吸收剂量值与标准计
算值之间的误差在 0.17% ̄0.54%之间, 小于国家标准 2.0%的要求。结论: 高能 X 线与电子束的吸收剂量测量结果能够
cGy,因此公式(1)右边乘以 2.58×10-4×100,这样单位统
一为 cGy.将以上各参数代入公式(1),可简化得出: 对 6
MV X 线而言 Dw=0.953 M,再将 Dw 的吸收剂量由百分 深度剂量 PDD (PDD=0.8626), 转换到最大剂量深度
dmax 处 的 吸 收 剂 量 D( dmax) ,这 里 PDD 为 水 深 为 5 cm 处的百分深度剂量 D( dmax) =DW/PDD=0.953 M/0.8626= 1.105 M, 这里令 Cx = NC·( W/e)·(Kat·t km)·SW, ai·r Pu·Pcel× 2.58×10-4×100, 即 Cx=1.105, 即 D( dmax) = Cx·M,所 以 当
6 1.3 2.32 5.41 2.42 2.50 1.084 0.955
Abstr act: objective : To measure and calculate absorbed dose of high energy X-ray and electron beam for medical linear ac- celerator ,assurance the accuracy of the output dose .Methods : Apply water phantom for measured medium ,measure and cal- culate absorbed dose by methods of IAEA 277 report and principle of dosimetry .Result : The errors between the measured absorbed dose and standard absorbed dose in the effective point of measurement were within 0.17%-0.54%.less than 2.0% of national standard .Conclusion : The measurement results of absorbed dose of high energy X-ray and electron beam were satis- fied with clinical dosimetry regulation . Key wor ds: high energy; X- ray; electron beam; absorbed dose; measurement; calculation
D( dmax) =100 cGy 时 , 剂 量 仪 读 数 M=100/1.105=90.50
R,这样可以直观的通过剂量仪读数 M 反映最大剂量
深度处 dmax 的吸收剂量情况, 平时检测时不再推算, 便可知道校准时误差情况.同理, 对于 15 MV X 线, 水
深为 10 cm 处的 PDD=0.7707,Cx =1.214, 剂量仪校准 时读数 M=82.37 R.
VARIAN Clinac 23EX 直线加速器高能 X 线有 6 MV 和 15 MV 两种。测量吸收剂量的条件为:标准源
中国医学物理学杂志 第 25 卷 第 2 期 2008 年 3 月
- 565 -
皮距 SSD=100 cm,水模体表面照射野 Field=10 cm×10
cm, 电离室的有效测量点在射线束轴上距水模体表
度 dc 取各能量电子束的最大剂量深度 dmax,同理, 高能
电子束在水中有效测量点处的吸收剂 DW 仍为:
Dw=M·Nc·(W/e)·(Kat·t km)·SW, ai·r
Pu·Pcel×2.58×10- 4×100
( 2)
公式( 2) 中除 SW, air 和 Pu 数值不同于公式(1)外, 其它均 相同, 下面就推导 SW, air 和 Pu 做简单描述, 具体推导过 程如下:
( 1)
对于公式( 1) 而言, Dw 的单位为 cGy;M 为标准剂 量仪读数, 单位为伦琴( R) ; Nc 为剂量仪校准因子, 每 年 剂 量 仪 送 检 一 次 , 由 国 家 标 准 实 验 室 给 出 , Nc=
1.000; W/e 为在空气中形成每对离子 (其电荷为 1 个
电子的电荷)所消耗的平均能量, 为一常数 W/e=33.97
J/C;Katt 为在校准电离室时, 电离室及平衡帽对校准辐 射 ( 一般为 60Co 的 ! 射线) 的吸收和散射的修正; Km
为电离室室壁及平衡帽的材料对校准辐射空气等效
不充分而引起的修正, 这里 Kat·t Km 数值[3]由剂量仪厂 家 给 出 Kat·t Km=0.973;Sw,air 为 校 准 深 度 水 对 空 气 的 平 均阻止本领比, 其数值由辐射质剂量比 D20/D10 或组 织模体比 TPR1020 导出, 对 6 MV X 线, 由于D20/D10=
收稿日期: 2007- 07- 17 作者简介: 李军( 1975- ) , 男, 陕西周至人, 江苏省苏北人民医院放疗科
物理师兼工程师, 南京大学物理系在读硕士, 主要从事临床 放 疗 物 理 和 放 疗 设 备 维 修 。 Tel: 0514- 87937554, E- mail: liajun.2007@yahoo.com.cn。
满足临床剂量学的要求。
关键词: 高能; X 线; 电子束; 吸收剂量; 测量; 计算
中图分类号: R 814.2
文献标识码: A
文章编号: 1005- 202X( 2008) 02- 0564- 03
Measur ement and Calculation of Absor bed Dose for High Ener gy X- r ay and Electr on Beam
2. 结果
2.1 高能 X 线吸收剂量的测量与计算结果 当机器出束 100 MU 时, M 为剂量仪读数, M 为剂
量仪平均读数,校准误差=[ ( Cx·M- 100) /100] ×100%, 具体在校准深度处的吸收剂量测量与计算结果见表2。 2.2 高能电子束吸收剂量的测量与计算结果
对于不同能量的高能电子束在最大剂量深度 dmax 处校准时的吸收剂量而言, 同样当机器出束 100 MU 时, 校准误差 =[ ( CE·M- 100) /100] ×100%, 具体的测 量与计算结果见表 3。
( 1) 由实际测出的电离量半值深度 DJ50,确定电子 束在水模体表面的平均能量是 ΕS, 再由 ΕS 和最大剂 量深度 dmax 确定 SW, air。
( 2) 确定 ΕS 后, 查表得出电子束在水中的实际射 程 Rp , 再由公式 Ez=ΕS( 1- dmax/Rp) 确定电子束在水模 体中不同深度处的平均能量 Εz。
确定好高能电子束在水中的吸收剂量公式( 2) 中 的各参数值后, 令 CE=N·c (W/e)·( Kat·t km)·SW,ai·r Pu·Pcel× (2.58×10-4)×100,即 Dw= CE·M , CE 可直接求出, 确定 CE 后,当 Dw=100 cGy 时, 剂量仪读数 M=100/ CE。
1.2.2 高能电子束吸收剂量的测量方法
VARIAN Clinac 23EX 直线加速器高能电子束有
4 MeV, 6 MeV, 9 MeV, 12 MeV, 16 MeV, 20 MeV 六
档能量。测量吸收剂量的条件为: 标准源皮距 SSD=
100 cm,照射野 Field 为 15 cm×15 cm 限 光 筒,校 准 深
表 1 电子束各参数的数值
Tab.1 The value of par ameter for electr on beam
E0
Βιβλιοθήκη Baidu
dmax
D50J
ES
RP
EZ
Sw,air
PU
(MeV) (cm) (cm) (MeV) (cm) (MeV)
4 0.6 1.50 3.03 1.52 1.83 1.103 0.952
1. 材料与方法
1.1 材料 美 国 VARIAN Clinac 23 EX 直 线 加 速 器, 德 国
PTW 公司的 UNIDOS E 型剂量仪和 TW30013 0.6cc 指型防水电离室, 30 cm×30 cm×30 cm 四支点可调水 箱, 20 m 长 TNC 三同轴电缆, 并且剂量仪, 电离室及 TNC 电缆均每年送国家标准实验室校准。 1.2 方法 1.2.1 高能 X 线吸收剂量测量方法
面的深度为校准深度 dc, 对 6 MV X 线而言, dc=5 cm;
对 15 MV X 线 而 言 dc=10 cm. 每 次 加 速 器 出 束 100
MU, 重复测量三次, 取其平均值。高能 X 线在水中有
效测量点处的吸收剂量 DW 为:
DW=M·NC·( W/e)·(Kat·t km)·
SW, ai·r Pu·Pcel×2.58×10-4×100
0.58 而知 SW, air=1.199; 而对 15MV X 线, 由于D20/D10=
0.65 而知 SW,air=1.099; Pu 为电离室扰动因子, 其数值经 查图[4]可得出 Pu=1.001;Pcel 为 电 离 室 中 心 电 极 影 响 因
子, 由生产厂家给出 Pcel=0.998.另外由于公式( 1) 两边 的单位的不同, 这里 1R=2.58×10-4 C/kg,而 1J/kg=100
- 564 - 2008 年 3 月 第 25 卷第 2 期
中国医学物理学杂志 Chinese J our nal of Medical Physics
高能 X 线和电子束吸收剂量的测量与计算
Mar.,2008 Vol.25. No.2
李 军, 张西志, 谭 飞 ( 江苏省苏北人民医院 放疗科, 江苏 扬州 225002)
前言
高能 X 线和电子束进入人体组织后, 通过和人 体组织中的原子相互作用而传递电离辐射的部分或 全部能量,人体组织吸收电离辐射后, 会发生一系列 的物理, 化学, 生物学变化。最后导致组织的生物学损 伤, 即生物效应。生物效应的大小程度正比于组织中 吸收的电离辐射的能量。因此, 确切的了解组织中所 吸收的电离辐射能量,对评估放射治疗的效果和它的 副作用是极其重要的, 他的精确确定时进行放射治疗 的最基本的物理要素[1] , 世界卫生组织( WHO) 和国际 原子能机构( IAEA) 联合组织进行的世界范围的剂量 对比显示,当前放射治疗中约有 15%的患者接受的剂
量不准确[2], 由此可见, 加强吸收剂量测量的规范和测 量的准确性, 不仅有着重要的临床意义, 并且也是医 院放疗物理师的基本任务之一。
目前有多种方法用于吸收剂量的测量, 但在临床 上大多用指型电离室在水模体中测量, 并使用 IAEA 277 报告方法对高能 X 线和电子束进行测量与计算, 使用此方法简便, 易行, 快速而精度很高, 因此对临床 物理师有着积极的意义。
( 3) 确定 Εz 后再由电离室内腔半径 r=3.05 mm,用 内差法及查表确定电离室扰动因子 Pu。
( 4) 通 过 以 上 各 步 骤 确 定 SW, air 和 Pu 的 值 后 即 可 确定公式( 2) 中各个参数的确定值。E0 为电子束初始 能量即标称能量, 再确定公式( 2) 中的各个参数值以 后, 可导出 DW 值, 具体各参数推导值见表 1。
摘要: 目的: 对医用直线加速器高能 X 线和电子束的吸收剂量进行测量与计算, 保证输出剂量的准确性。方法: 以水模
体作为介质, 用 IAEA 277 报告及剂量学原理进行吸收剂量的测量与计算。结果: 在有效测量点的吸收剂量值与标准计
算值之间的误差在 0.17% ̄0.54%之间, 小于国家标准 2.0%的要求。结论: 高能 X 线与电子束的吸收剂量测量结果能够
cGy,因此公式(1)右边乘以 2.58×10-4×100,这样单位统
一为 cGy.将以上各参数代入公式(1),可简化得出: 对 6
MV X 线而言 Dw=0.953 M,再将 Dw 的吸收剂量由百分 深度剂量 PDD (PDD=0.8626), 转换到最大剂量深度
dmax 处 的 吸 收 剂 量 D( dmax) ,这 里 PDD 为 水 深 为 5 cm 处的百分深度剂量 D( dmax) =DW/PDD=0.953 M/0.8626= 1.105 M, 这里令 Cx = NC·( W/e)·(Kat·t km)·SW, ai·r Pu·Pcel× 2.58×10-4×100, 即 Cx=1.105, 即 D( dmax) = Cx·M,所 以 当
6 1.3 2.32 5.41 2.42 2.50 1.084 0.955