卧式圆形压力蒸汽灭菌器
压力蒸汽灭菌器系列技术参数
压力蒸汽灭菌器系列专业生产厂家:苏州江东精密有限公司蒸汽灭菌器:YX280系列、YM立式系列、YX600等容积不同,功能各异的灭菌器,利用饱和蒸汽,通过高温高压对所需物品进行高温灭菌,广泛适用于医疗、卫生、防疫、环保、高校科研、食品化工等单位消毒灭菌之用。
产品名称型号容积单价主要技术指标与特点手提式不锈钢压力蒸汽灭菌器(防干烧)YX-280A 18L 1210220V 2kW 座式电热工作温度126℃(121℃可选)工作压力0.142mpaYX-280AS 18L 1650220V 2kW 电热定时YX-280A* 24L 2310220V 2kW 座式电热YX-280B 18L 1100220V 2kW 煤电两用YX-280B* 24L 1870220V 2kW 煤电两用不锈钢立式电热蒸汽灭菌器YM30B 30L 7370220V 3kW 自动补水YM30Z 30L 10560220V 3kW 定时数控YM50A 50L 7590220V 3kW人工加水YM50B 50L 10120220V 3kW 自动补水YM50Z 50L 12980220V 3kW 定时数控YM75A 75L 10560220V 3kW 人工加水YM75B 75L 12320220V 3kW 自动补水YM75Z 75L 14520220V 3kW 定时数控YM100A(原YM75AI)100L 16390220V 3.8kW 人工加水YM100Z(原YM75ZI)100L 18920220V 3.8kW 定时数控YM50FG 50L 16280220V 3kW 智能干燥YM50FN 50L 17380220V 3kW 智能内排YM50FGN 50L 21230220V 3kW 干燥内排YM50CM 50L 25960220V 3kW 触摸智能YM75FG 75L 18700220V 3kW 智能干燥YM75FN 75L 19800220V 3kW 智能内排YM75FGN 75L 25080220V 3kW 干燥内排YM75CM 75L 28600220V 3kW 触摸智能卧式圆形压力蒸汽灭菌器YX-600W 300L 39380380V 12kW自动化程序控制(需上门指导使用,另收费1500-3000元)YX-600W- 150L 31680220V 7.5KW自动化程序控制。
医疗机构消毒与灭菌基础知识培训(06、10龙华区)
注意事项: 1 、 所用消毒剂必须有卫生许可批 件且在有效期内。 件且在有效期内。 2、消毒时室内不可有人。 消毒时室内不可有人。 3 、 甲醛因有致癌作用不宜用于空 气消毒。 气消毒。
四、医疗器械的消毒与灭菌
压力蒸汽灭菌法 干热灭菌法 环氧乙烷灭菌法 戊二醛浸泡法 低温气体等离子体灭菌方法
一、压力蒸汽灭菌 压力蒸汽灭菌属于湿热灭菌 的一种,它是通过蒸汽在高过大 气压力情况下,可形成高温,从 而增加杀菌能力和速度,可达到 灭菌的要求,并成为目前使用最 为普遍,效果最为可靠的一种灭 菌方法。
(5)正确的使用和维护 紫外线消毒空气首先应照射 足够的时间和频度,每次照射 30~60min。每天照射不少于2次 30~60min。每天照射不少于2 或每次工作之前照射。
在使用过程中,灯管表面的 灰尘会影响消毒效果,所以应注 意对灯管表面的清洁和环境的条 件,一般两周用酒精棉球擦拭一 次,发现灯管表面有灰尘、油污 时,应随时擦拭;
(1)固定式照射法 最常用的方法是将紫外线杀 菌灯悬挂在室内天花板上,以垂 直向下照射或反向照射方式进行 照射消毒。亦可将紫外线灯固定 在室内墙壁进行横向照射或安装 在过道墙壁形成屏幕式照射。
室内悬吊式紫外线消毒灯安 装数量为平均每立方米不少于 1.5W,如60m3房间需要安装30W 5W,如60m 房间需要安装30W 紫外线灯3 紫外线灯3支,并且要求分布均 匀、吊装高度距离地面1 匀、吊装高度距离地面1.8~ 2.2m,使得人的呼吸带处于有 2m,使得人的呼吸带处于有 效照射范围。
医疗机构常用的消毒 与灭菌方法
海南现代女子医院 叶荣珍 二00六年十月二十三日 00六年十月 化学方法 生物杀菌法
1、物理方法:是利用物理因子杀 灭或清除微生物的方法。
SOP-JZ-016 微生物发酵菌种中心公共文件-卧式圆形压力蒸汽灭菌器使用、维护保养标准操作规程
1.4.5 灭菌锅在灭菌过程中,操作者应经常检查压力表,温度计录仪是否正常,不得擅自
离开岗位。
1.4.6 灭菌终了应对照 121℃压力蒸汽灭菌化学指示卡。指示剂的颜色达到或深于“标准
色”则表示已满足灭菌所需条件,合格可使用,否则无效。
1.5 干燥与冷却
1.5.1 干燥:灭菌时间终了,按各种不同灭菌物品的要求,灭菌室的蒸汽自然冷却或予以
2.注意事项:
2.1 不同类型的物品最好不要同时进行灭菌,如敷料与溶液,橡胶与器械等。以免顾此失
彼,不能获得良好的效果。
SOP-JZ-016-00
卧式圆形压力蒸汽灭菌器使用维护保养标准操作规程
3-3
2.2 灭菌柜严禁在压力表指示未到“0”时打开柜门,以免发生人员伤亡事故。 2.3 当蒸汽输入灭菌室应徐徐开放冷凝水泄出器前端之阀门,使冷凝水泄出器间断地排出 灭菌室内之空气及冷凝水,确保灭菌效果。 2.4 在每次加热时须检查水位表,确保有足够的贮水量。如水位报警指示灯亮,则立即检 查水位;如水位过低,应停机并将总阀旋钮转至“全排”位置,待灭菌室压力表指针归 零后,补水后再加热。 2.5 要注意检查灭菌器水位表内的导电柱是否在水位玻璃管的正中。歪斜的导电柱不能产 生正常切断电源效果,同时要及时去除导电柱尖端上的水垢,以防电阻增高,造成不能 及时断电的效果。 3.维护: 3.1 检查灭菌器门密闭情况,当垫圈老化和残损而漏气时,应及时更换(一般半年更换一 次)。 3.2 压力表、温度表及安全阀应由设备部相关人员定期(半年/一年)送检。 3.3 应保持设备内外清洁,柜门锁紧密封装置及转动部位每半年加油一次。 3.4 操作人员在完成上述工作后,应及时填写相关记录并签名。
文件编码 Code: SOP-JZ-016-00
湿热灭菌验证报告
文件编号: 版 本 号: 编 制: 审 核: 批 准: 验证时间:
目录
1、湿热灭菌验证方案.................................................................................................................................................1 1.1 概述.................................................................................................................................................................1 1.2 目的.................................................................................................................................................................1 1.3 验证范围.........................................................................................................................................................1 1.4 验证小组及职责.......................................................................................................... 错误!未定义书签。 1.5 验证条件.........................................................................................................................................................1 1.5.1 样品选择............................................................................................................................................ 1 1.5.2 生产设备............................................................................................................................................ 2 1.5.3 产品包装物料.................................................................................................................................... 2 1.5.4 生物指示物........................................................................................................................................ 2 1.5.5 温度传感器........................................................................................................................................ 2 1.5.6 装载模式............................................................................................................................................ 2 1.5.7 灭菌参数初步确定............................................................................................................................ 2 1.6 验证方法.........................................................................................................................................................3 1.7 验证内容.........................................................................................................................................................4 1.7.1 安装确认(IQ)................................................................................................................................ 4 1.7.2 运行确认(OQ)................................................................................................................................ 5 1.7.3 性能确认(PQ).............................................................................................................................. 11 1.8 验证依据及标准........................................................................................................................................... 12
卧式方形压力蒸汽灭菌器简介
卧式方形压力蒸汽灭菌器简介
结构特点:
灭菌器内室采用进口304不锈钢板,主板为双层结构;真空系统采用水环式进口真空泵,真空度高,噪音小;采用高性能程序控制。
本机装有安全阀、电磁阀、疏水阀及安全连锁装置结构,整机控制精确、稳定,故障率低,安全可靠。
主要技术参数
灭菌室有效尺寸(mm):500×600×1000
灭菌室有效容积(m3):0.3
设计压力(MPa):0.162
额定工作压力(MPa):0.14
蒸汽发生器容积:42L
内加热式热源电热管6支、单支功率2.5Kw
额定电压::15Kw
外形尺寸(L×W×H):1340×800×1746
包装尺寸(L×W×H):1490×810×1766
净重:600kg
卧式方形压力蒸汽灭菌器特点>>
真空系统采用水环式进口真空泵,真空度高,噪音小
卧式方形压力蒸汽灭菌器应用>>
卧式方形压力蒸汽灭菌器,广泛应用于医院、疾病防控中心等卫生医疗
保健机构,卧式方形压力蒸汽灭菌器的厂家很多,应用广泛,各种产品之间的差别也比较大。
卧式方形压力蒸汽灭菌器。
实验室高压灭菌器仪器分类
实验室高压灭菌器仪器分类
实验室高压灭菌器仪器主要可以分为以下几种类型:
1.辐射灭菌设备:如医用伽玛射线灭菌器。
2.压力蒸汽灭菌设备:包括预真空蒸汽灭菌器、高压蒸汽灭菌器、自动高压蒸汽灭菌器、立式压力蒸汽灭菌器、卧式圆形压力蒸汽灭菌器、卧式矩形压力蒸汽灭菌器、脉动真空压力蒸汽灭菌器、手提式压力蒸汽灭菌器等。
3.气体灭菌设备:如环氧乙烷灭菌器、轻便型自动气体灭菌器等。
4.干热灭菌设备:如干热灭菌器、微波灭菌柜等。
此外,还有高压电离灭菌设备,如手术室用高压电离灭菌设备、病房用高压电离灭菌设备等。
这些设备各有特点,适用于不同的环境和用途。
在选择时,应根据具体的需求和条件进行考虑。
卧式圆形压力蒸汽灭菌器的操作流程
卧式圆形压力蒸汽灭菌器的操作流程1.将压力蒸汽灭菌器放置在水平稳定的台面上。
Place the pressure steam sterilizer on a stable and level surface.2.打开蒸汽灭菌器的盖子,并确保内部干净整洁。
Open the lid of the steam sterilizer and ensure that the interior is clean and tidy.3.将待灭菌的物品放置在蒸汽灭菌器的合适位置上。
Place the items to be sterilized in the appropriate position within the steam sterilizer.4.确保所放置的物品之间相互之间不触碰。
Ensure that the placed items do not touch each other.5.加入适量清水至蒸汽灭菌器的水槽中。
Add an appropriate amount of clean water to the water tank of the steam sterilizer.6.关上蒸汽灭菌器的盖子,并确保盖子已经紧密关闭。
Close the lid of the steam sterilizer and ensure that it is tightly sealed.7.调节蒸汽灭菌器的压力阀和温度控制器至所需的设定数值。
Adjust the pressure valve and temperature controller of the steam sterilizer to the desired setting.8.打开蒸汽灭菌器的电源,启动设备。
Turn on the power of the steam sterilizer to start the device.9.等待蒸汽灭菌器内的水开始产生蒸汽,并加压。
卧式高压灭菌器的使用流程
卧式高压灭菌器的使用流程1. 准备工作在使用卧式高压灭菌器之前,需要进行一些准备工作,以确保其正常使用和灭菌效果。
•清洁灭菌器:首先,需要使用适当的清洁剂对卧式高压灭菌器进行清洁,包括内部和外部。
确保清洁剂不会对灭菌器产生腐蚀作用。
•准备灭菌器配件:检查并确保所有灭菌器配件的完整性和干净程度,如灭菌袋、灭菌器篮子等。
•准备需灭菌的物品:准备好需要进行灭菌的物品,并将其放置在灭菌器篮子中。
注意遵循物品的灭菌要求,不要超过灭菌器的负荷限制。
2. 设置和调试在开始灭菌程序之前,需要对卧式高压灭菌器进行设置和调试,以确保符合所需灭菌参数。
•设置温度和压力:根据灭菌要求,将温度和压力设置到适当的数值。
通常情况下,温度和压力的设置是提前确定好的,可以在灭菌器的控制面板上进行调节。
•调试操作:进行一次试运行以检查卧式高压灭菌器的正常工作状态。
确认灭菌器的门是否正常关闭、排气管道是否通畅等。
3. 进行灭菌在完成准备工作和调试后,可以开始进行卧式高压灭菌器的正式灭菌程序。
以下列点方式描述灭菌步骤:•将灭菌器篮子放置在灭菌器的固定位置,并确保灭菌篮子之间有足够的间隔,以便蒸汽可以顺畅流通。
•打开卧式高压灭菌器的门,将待灭菌的物品放入篮子中,并确保物品之间的间隔合适,以免影响灭菌效果。
•关闭灭菌器的门,并确保门已经完全关闭并且密封。
•在控制面板上选择合适的灭菌程序,包括温度、压力和时间等参数设定。
根据物品的灭菌要求,选择合适的程序。
•按下开始按钮,启动灭菌程序。
卧式高压灭菌器将开始加热、产生蒸汽,并保持一定的温度和压力,以达到灭菌要求。
•等待灭菌程序完成后,灭菌器会自动停止工作,并发出提示音。
此时,可以打开灭菌器的门,取出已灭菌的物品。
4. 灭菌后处理在取出已灭菌的物品之后,需要进行一些处理工作,以保持物品的无菌状态和安全性。
•检查物品:检查已灭菌的物品是否满足要求,如无菌包装完好、物品表面无明显污染等。
如发现问题,应及时进行处理。
SGLASE-RS1D 卧式方形压力蒸汽灭菌器详细说明书
SGLASE-RS1D 卧式方形压力蒸汽灭菌器详细说明
书
一、本说明书用于指导用户正确安装和操作本公司生产制造的卧式矩形灭菌器,在安装和使用本设备之前,必须仔细阅读并理解本说明书,如有疑问,应向本公司专业技术人员咨询。
二、本设备的安装与调试,须由精通本设备的专业人员负责并在其指导下进行。
三、本设备的使用操作者必须经过专门培训,熟悉操作系统,在操作过程中严守操作规程。
每次灭菌完毕,在取出灭菌物后,必须彻底清理灭菌室内杂物,用清水刷洗干净,并用干净棉布或棉纱擦干内壁水渍
四、本设备为一类压力容器,严禁超压使用,有关容器具体数据,详见本设备质量证明书。
五、使用本设备前,应向当地压力容器主管部门申报备案。
本说明书更改权属设计者,本说明书若有更改怨不通知。
六、原理及适用范围
本设备为电加热式卧式矩形压力蒸汽灭菌器,是采用饱和蒸汽进行消毒灭菌,利用蒸汽冷凝时释放大量潜热,使被消毒的物品处于高温和潮湿状态,经过一定时间,破坏所有细菌及芽胞组织,达到消毒灭菌的目的。
被广泛用于医疗、食品、制药等行业以及科研院所,对手术器械、敷料、药品、培养基、食品等作消毒灭菌之
用。
卧式圆形压力蒸汽灭菌器的操作流程
卧式圆形压力蒸汽灭菌器的操作流程英文回答:The operating procedure of a horizontal cylindrical pressure steam sterilizer is as follows:1. Preparing the sterilizer: Ensure that the sterilizer is clean and free from any debris or residue. Check the water level and replenish if necessary. Ensure that the drain valve is closed.2. Loading the items: Place the items to be sterilized in the sterilization chamber, ensuring that they are properly arranged and spaced to allow for adequate steam penetration. Avoid overloading the chamber to ensure effective sterilization.3. Closing the door: Close the sterilizer door securely and ensure that it is properly sealed. This is important to maintain the pressure and temperature inside the chamberduring sterilization.4. Setting the parameters: Set the desiredsterilization parameters such as temperature, pressure, and time according to the specific requirements of the items being sterilized. These parameters may vary depending on the type of material and the level of sterilization required.5. Starting the sterilization cycle: Once the parameters are set, start the sterilization cycle. The sterilizer will begin to heat up, and the pressure will gradually increase to the desired level. The sterilization time will also start counting down.6. Sterilization phase: During this phase, thesterilizer will maintain the set temperature and pressure for the specified duration. The steam will penetrate the items, killing any microorganisms present and ensuring effective sterilization.7. Cooling phase: After the sterilization phase iscomplete, the sterilizer will enter the cooling phase. The pressure and temperature inside the chamber will gradually decrease. It is important to wait until the pressure reaches a safe level before opening the door.8. Opening the door: Once the pressure has reached a safe level, open the sterilizer door carefully. Use appropriate protective equipment such as gloves to avoid any potential burns from the hot items inside. Remove the sterilized items and handle them appropriately.9. Cleaning and maintenance: After each use, clean the sterilizer chamber and accessories thoroughly to remove any residue or contaminants. Regular maintenance andcalibration of the sterilizer should also be performed to ensure its proper functioning.中文回答:卧式圆形压力蒸汽灭菌器的操作流程如下:1. 准备灭菌器,确保灭菌器清洁无残留物。
卧式灭菌器使用操作规程
标准操作规程Standard Operating Procedure1.目的建立卧式灭菌器使用操作规程,确保其操作规范化。
2.适用范围适用于卧式圆形压力蒸汽灭菌器(电加热)的操作。
3.责任者操作者4.操作规程4.1加水灭菌器必须在操作前将仪表箱上总阀转至全排位置并打开进水阀,加水至水箱高水位标记处,然后将进水阀关闭。
每次灭菌前都应检查贮水量,如有不够,应予补充。
4.2加热先将总阀旋钮转至“关闭”,然后将压力控制开关的旋钮旋至“开”处,按物品所需灭菌温度,将压力控制旋纽旋转至相应位置,此时电热指示灯亮,同时其对应的压力指示灯亮,表示加热开始。
4.3堆放将待灭菌物品顺序地、保持适当间隙地放入灭菌室内。
同时将瓶盖旋开少许。
4.4密闭将门关上,按顺时针方向拨动锁紧手柄中心盘转至中方位,使撑挡进入门圈内,然后按顺时针旋转八角转盘直至仪表箱面板上门指示灯亮。
4.5灭菌当夹套内蒸汽压力达到所选择的控制压力时,同时门关闭指示灯亮,即将总阀旋转至“灭菌”位置,同时将冷凝水排出阀旋开适宜位置。
当灭菌室温度达到预选温度值时,开始记时。
注意调整冷凝水排出阀,使灭菌温度保持在预选温度值。
4.6排气当消毒终了时,应将总阀旋钮旋至“慢排”或切断电源使灭菌室压力和温度自然降低,直至灭菌室压力表回降至“0”位,随后将总阀旋钮旋至“全排”档,排去套层和灭菌室内的蒸汽,并打开灭菌室门少许,散发灭菌室内的剩余蒸汽,使灭菌室内壁保持干燥。
4.7物品取出储存当所灭菌物品冷却后,打开灭菌器门,将其取出放于冰箱存放。
(需旋紧瓶盖的先旋紧瓶盖后取出)5.注意事项5.1压力表经常检查是否灵敏,以保安全。
5.2灭菌室内有压力切不可打开灭菌器门,以防事故。
5.3不同类型物品最好不要同时进行灭菌。
5.4每次加热前须检查水位表。
卧式圆形压力蒸汽灭菌器操作规程
卧式圆形压力蒸汽灭菌器操作规程一、设备准备1.检查设备的电源线和插头是否完好无损,并接好电源。
2.检查设备的控制系统、显示屏和操作按钮是否正常工作。
二、设备预热1.打开设备的蒸汽开关,加热锅内的水,等待水温升至设定温度。
2.在设备预热过程中,检查设备的温度计和压力计是否正常显示。
三、操作准备1.检查待灭菌物品是否符合灭菌要求,如密封包装、无潮湿、无明显污染等。
2.检查设备内部的结构件和密封件是否完好无损,如有破损应及时更换。
3.准备好所需的灭菌指示物品和记录表格。
四、灭菌操作1.将待灭菌物品放置在灭菌器的内部篮子中,并确保物品之间有足够的间距。
2.关闭设备的揭门,将锁紧装置锁好,确保密封。
3.设置灭菌时间和温度,可根据不同物品的要求进行调整,一般为121℃、15分钟。
4.启动设备的灭菌程序,设备会开始加热蒸汽,同时控制温度和压力。
5.在灭菌过程中,密切观察设备的温度和压力变化,确保符合设定范围,如有异常应及时停止操作。
6.灭菌完成后,设备会自动停止加热,锁紧装置会解除,灭菌器内的压力逐渐下降。
7.等待设备内压力完全释放后,打开揭门,取出已灭菌物品。
五、灭菌后处理1.将已灭菌物品进行分类整理,如存放在干燥、洁净的区域内。
2.检查已灭菌物品的密封包装是否完好,如有破损应立即重新处理。
3.根据灭菌指示物品的变色情况,确认灭菌效果,并将结果记录在相关表格上。
4.将灭菌器的内部进行清洁,去除残留物和污垢,保持设备的清洁和卫生。
5.定期检查设备的温度计、压力计和密封件的工作状况,如有异常应及时维修或更换。
六、安全注意事项1.操作人员在操作设备时,必须穿戴好防护服和防护手套,以避免烫伤和感染。
2.操作人员要时刻注意设备的工作状态,如发现异常情况应立即停止操作,并报告相关人员处理。
3.操作人员离开设备时,应关闭设备的电源,确保设备的安全。
4.灭菌器内的蒸汽和压力具有一定的危险性,操作人员在打开揭门时要小心操作,避免受伤。
卧式圆形压力蒸汽灭菌器的操作规程
卧式圆形压力蒸汽灭菌器的操作规程使用前的准备工作检查各仪表是否正常,电源是否正确良好,备用阀门应处于关闭状态,工作前必需把水加到高水位线以上,并设定好灭菌温度和灭菌时间,操作规程、步骤1: 堆放灭菌器的装载量不得超出灭菌室内容量的80%。
应尽量将同类物品放在一起灭菌,若必需将不同类物品装放在一起,则以最难实现灭菌物品所需的温度和时间为准物品装放时,上下左右相互间均应间隔确定距离以利蒸汽置换空气,物品包体积不得超出200mmx200mmx250mm难于灭的大包放在上层,较易灭菌的小包放在下层,金属物品放下层,织物包放上层,物品装放不能死靠门和四壁,以防吸入较多的冷凝水,金属包应平放,盘、碟、碗等应处于树立的位置;纤维织物应使折叠的方向与水平面成垂直状态;玻璃瓶等应开口向下或侧放以利蒸汽进入和空气排出,启闭式筛孔容器,应将筛孔的盖打开。
对溶液进行灭菌时,应灌注于硬质而热玻璃瓶中,以不超出其容积的314为宜,用棉龙纱布赛将瓶赛好,并里纱强于瓶颈,防止纱布赛落入瓶内,切勿使用未打孔的像胶或软木赛,最好将玻璃相放置于容积稍大的搪瓷或金属盘内,以防玻璃瓶爆裂时,溶液流失和污染灭菌室金属内壁。
2: 关门将门关上,按顺时针方向拨动锁紧手柄至中方位,使撑挡杆进入门圈内。
然后顺时针方向转动八角手轮,使门和橡胶垫圈密合。
3: 加水必需在操作前关团排水阀,开启进水阀,打开排空阀,使水进入水箱内,加至高水位上5—10m处为上,然后关团进水阀和排空阀。
每次灭菌前都应当检查败水,如有不足应予增补。
4: 加热关紧冷凝掌控阀,将电源开关拨至“开”,电源指示灯亮,水位指示灯亮,开启加热开关,加热指示灯亮,表示水箱已开始加热,产生蒸汽进入夹套。
当水箱水位低于规定时,能自动切断加热电源,水位指示灯熄灭,报警指示亮并发出峰鸣声,这时应关闭电源开关,待夹层蒸汽压力为季时再加水,加水后,方能线续加热。
当夹套压力超出夹套压力表设定上限值时,加热电源会自动切断,加热指示灯随即熄灭,以后电流时通时断,自动使压力保持稳定,当夫套压力超出压力掌控器的设定压力时,加热指示灯熄灭,加热停止。
卧式蒸汽灭菌器操作方法
卧式蒸汽灭菌器操作方法
卧式蒸汽灭菌器是一种常见的灭菌设备,用于对医疗器械、制药品等进行灭菌处理。
以下是卧式蒸汽灭菌器的操作方法:
1. 填充水箱:打开灭菌器的水箱盖,倒入适量清水,确保水箱中有足够的水。
2. 放置物品:将待灭菌的物品放置在灭菌室内,注意排列整齐,确保蒸汽能够充分接触到每个物品表面。
3. 设置灭菌程序:根据物品的特性和灭菌要求,在控制面板上选择合适的灭菌程序和参数,如温度、压力和时间等。
通常,灭菌程序会根据物品的不同而有所区别。
4. 启动灭菌器:按下启动按钮,蒸汽灭菌器开始工作。
设备会自动加热水箱内的水,产生蒸汽,并将蒸汽输入到灭菌室中。
5. 灭菌过程中的监控:在整个灭菌过程中,可以通过控制面板上的显示屏实时监控灭菌的进度和状态。
确保温度、压力等参数在设定范围内。
6. 灭菌完成提示:当灭菌程序完成后,灭菌器会发出声音提示或显示灯亮起。
此时,需要关闭蒸汽灭菌器,并等待一段时间,待灭菌室内的压力和温度下降后再打开设备门。
7. 取出物品:小心打开设备门,使用器械或工具将灭菌后的物品取出。
注意操作过程中的个人安全和卫生要求。
8. 清洁和维护:使用后,务必对蒸汽灭菌器进行清洁和维护工作。
清理蒸汽灭菌器内的水箱、灭菌室和排放管道,确保设备的卫生和功能正常。
以上是卧式蒸汽灭菌器的基本操作方法,根据不同品牌和型号,具体的操作步骤可能会有所不同,建议在使用前仔细阅读设备的使用说明书,并按照厂家提供的操作指导进行操作。
病原微生物实验室生物安全手册
目录一、总则二、实验室生物安全管理体系三、生物因子生物危害评估四、实验室人员准入制度五、人员培训考核制度六、人员健康监护制度七、生物安全检查制度八、实验室人员生物安全行为规范及内务管理制度九、实验室菌种安全保管制度十、生物安全实验室资料档案管理制度十一、实验室废弃物管理制度十二、实验室消毒隔离制度十三、实验室生物危险标识使用规定十四、事件、伤害、事故和职业性疾病报告制度十五、实验室应急处置预案十六、生物安全标准操作规程(一)生物洁净安全柜标准操作规程(二)净化工作台操作规程(三)手提式压力蒸汽灭菌器操作规程(四)离心机操作规程(五)实验室操作人员洗手、消毒操作规程十七、其他必要的管理性和技术性文件(一)设施/设备监测,检测和维护制度(二)防火安全制度(三)实验室用电安全制度(四)微生物实验室安全保卫制度一、总则1、目的为确保实验室全体人员熟悉生物安全法律、法规,建立生物安全意识,保证相关工作人员掌握开展工作必需的生物安全知识和技术,避免实验室感染,防止实验室事故,特制定此手册。
2、依据《病原微生物实验室生物安全管理条例》、《医疗废物管理条例》、《实验室生物安全通用要求》、《实验室生物安全手册》、《生物安全实验室建筑技术规范》3、适用范围适用于进入微生物检查实验室所有工作人员。
4、修订国家及卫生部门涉及生物安全的法律法规发生修订更改时,本手册应相应作出修订。
二、实验室生物安全管理体系1、单位成立生物安全委员会,全面负责公司生物安全工作。
责任人:单位法人2、微生物检验科实验室成立生物安全三级管理组织。
一级生物安全管理:微生物检验科科长(1)由微生物检验科科长全面负责科内生物安全管理文件的制定。
(2)督促、检查各个实验室遵守、落实生物安全情况。
(3)设立生物安全监督员一名。
(4)组织全科定期、不定期的学习生物安全的相关法律、法规。
二级生物安全管理:各个实验室负责人(1)负责本实验室生物安全工作。
(2)全面落实生物安全管理有关本组的相关制度。
GMP认证全套文件资料16蒸汽灭菌器操作SOP
(8)如不需要连续灭菌时,将电源或热源开关断开,随后将总阀拨至“全排”排去套层内的蒸汽,并将门开启少许,散发灭菌室内的剩余蒸汽,使灭菌室内壁经常保持干燥。
3、填写记录。
(2)将门关上,按顺时针方向拨动锁紧手柄至中方位,使撑档进入门圈内,旋动转盘,使门和垫圈密合。
(3)接通电源,将电源开关按到“开”,电源指示灯亮,若水箱内水位低于规定时,加热灯不亮,电热器不加热,同时低水位指示灯亮,蜂鸣器报警,此时应关闭电源后,加水至规定水位。
(4)如为直接蒸汽加热型设备,先将总阀拨到“关闭”,再将蒸汽减压阀拨至所需工作压力档,然后,开动蒸汽控制阀使外来蒸汽进入灭菌器夹套,在加热时要适当套内排出,确保达到相应的温度。
蒸汽灭菌器操作SOP
第1页共2页
文件类型
设备SOP文件
文件编码
SOP-EQ-1016-00
执行日期
执行部门
设备工程部、质量管理部
起 草 人:
起草日期:
审 核 人:
审核日期:
批 准 人:
批准日期:
修 订 号 批准日期 执行日期
变更原因及目的:
目 的:建立蒸汽灭菌器使用的标准操作程序。
范 围:WY21/22600YXQ卧式圆形压力蒸汽灭菌器的使用操作过程。
责任者:操作人员、设备管理人员。
程 序:
1、准备工作
(1)开启进水阀,使水进入蒸汽发生器内,加至离水位表顶端5─10mm处为止,然后关闭进水阀,每次灭菌前都应检查贮水,如有不够,须补充。
(2)将待灭菌的物品予以妥善包扎,包装不宜过大,一般以200×200×100mm为佳。
医疗机构常用的消毒与灭菌方法
⑵应尽量将同类物品放在一 起灭菌,若将不同类物品装放在 一起,则以最难达到灭菌物品所 需要的温度和时间为准。
⑶物品装放时,上下左右相 互间均应间隔一定距离以利蒸汽 置换空气;
⑷难于灭菌的大包放在上层 ,较易灭菌的小包放在下层;金 属物品放在下层,织物包放在上 层; ⑸启闭式筛孔容器,应将筛 孔的盖打开。
2、包装: ⑴包装材料应允许物品内部 空气的排出和蒸汽的透入。市售 的普通铝饭盒与搪瓷盆,不得用 于装放待灭菌的物品,应用自动 启闭式或带通气孔的器具装放; 常用的有储物槽(盒);
(2)布包装层数不少于两层, 用下排气式压力蒸汽灭菌器的物 品包,体积不得超过 30cm×30cm×25cm ; 用于预真空和脉动真空压力 蒸汽灭菌器的物品包,体积不得 超过30cm×30cm×50cm。 金属包的重量不超过7公斤, 敷料包不超过5公斤;
不能用于凡士林等油类和粉 剂的灭菌。预真空式压力蒸汽灭 菌器因所抽负压较强,可使瓶内 液体溢出,因此不能用于瓶装液 体的灭菌。
(三)灭菌前物品的准备
1 、清洗:灭菌前应将物品彻 底清洗干净,物品洗涤后,应干 燥并及时包装。清洗不彻底是灭 菌失败的因素之一。特别是接触 脓血便的物品污染严重,不容易 清洗,有机物残留影响了灭菌效 果。对于污染的医疗器械处理程 序应该是先消毒、再清洗。
医院的无菌器材中心供应室 往往将之作为灭菌的首选方法。 主要用于对耐热、耐湿物品的处 理。因此,正确的掌握压力蒸汽灭
菌器的使用方法及注意事项等知识, 对保证灭菌效果,减少医院性感染 是非常必要的。
(一)压力蒸汽Biblioteka 菌的分类根据排冷空气的方式和程度不同, 分为下排气式和预真空式两大类。手 提式、立式、卧式压力蒸汽灭菌器均 属于下排气式压力蒸汽灭菌器。
卧式矩形压力蒸汽灭菌器的操作流程
卧式矩形压力蒸汽灭菌器的操作流程下载温馨提示:该文档是我店铺精心编制而成,希望大家下载以后,能够帮助大家解决实际的问题。
文档下载后可定制随意修改,请根据实际需要进行相应的调整和使用,谢谢!并且,本店铺为大家提供各种各样类型的实用资料,如教育随笔、日记赏析、句子摘抄、古诗大全、经典美文、话题作文、工作总结、词语解析、文案摘录、其他资料等等,如想了解不同资料格式和写法,敬请关注!Download tips: This document is carefully compiled by theeditor. I hope that after you download them,they can help yousolve practical problems. The document can be customized andmodified after downloading,please adjust and use it according toactual needs, thank you!In addition, our shop provides you with various types ofpractical materials,such as educational essays, diaryappreciation,sentence excerpts,ancient poems,classic articles,topic composition,work summary,word parsing,copy excerpts,other materials and so on,want to know different data formats andwriting methods,please pay attention!卧式矩形压力蒸汽灭菌器是一种常用的实验室设备,用于杀灭细菌、病毒和其他微生物,以确保实验材料的无菌状态。
- 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
- 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
- 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
卧式圆形压力蒸汽灭菌器产品简介及性能特点一、产品简介:
该系列产品是利饱和蒸汽对物品进行消毒灭菌。
其结构合理、灭菌速度快、自带蒸发器、不需外接蒸汽。
适用于医院、制药、防疫、食品、科研等单位,对手术器械、器皿、药品、敷料和棉织物等进行消毒灭菌。
二、性能特点:
1.灭菌器灭菌室内壳采用优质304不锈钢板材制作,与灭菌物品接触的各处采用抛光处理,光滑平整、无死角、易清洗,避免了传统不光滑而引起的残留细菌繁殖。
2.门装有安全联锁装置,该装置可保证门未关紧,蒸汽不能进入灭菌室,灭菌室内压力大于0.027Mpa时门打不开,而且具有与上述动作同步的报警功能。
3.自带蒸汽发生器,不需外接蒸汽。
4.主要控制配件选项用进口件,使用寿命长、安全可靠。
5.灭菌器灭菌室内达到工作压力后,灭菌室内各测试点的温差不超过2℃。
卧式圆形压力蒸汽灭菌器主要技术参数
卧式圆形压力蒸汽灭菌器主要配件型号、规格、产地
注:使用寿命系根据正常使用频率进行测算的。