2D电泳(常见问题与原因)
- 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
- 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
- 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
283
1
Two-dimensional Electrophoresis
Additional trouble shooting guides for 2-D electrophoresis are found in:
.
Berkelman T,Stenstedt T.Handbook:2-D electrophoresis using immobilized pH gradients.Principles &methods.Ame-rsham Biosciences 80-6429-60(1998).
.
Görg A,Weiss W.Two-dimensional electrophoresis with immobilized pH gradients.In Rabilloud T,Ed.Proteome research:Two-dimensional gel electrophoresis and identifica-tion methods.Springer Berlin Heidelberg New York (2000)107±126.
In the following part only a selection of cases are described,which typically occur now and then.
1.1
Isoelectric focusing in IPG strips
Rehydration solution is distributed unevenly within the gel strip.
Some coating in the strip holder or reswelling tray.Wash strip holder and res-welling tray with
detergent,rinse with deionised water.
Uneven pipetting of the rehydration solution.Pipette the solution as a streak.Reswelling tray or IPGphor not levelled.Adjust the level of the res-welling tray or the
IPGphor on the bench.Rehydration liquid is left in the reswelling tray or strip holder.
Rehydration time too
short.
Rehydrate at least for 6hours without and 12hours with sample.A Trouble shooting
Proteomics in Practice:A Laboratory Manual of Proteome Analysis
Reiner Westermeier,Tom Naven
Copyright 2002Wiley-VCH Verlag GmbH
ISBNs:3-527-30354-5(Hardcover);3-527-60017-5(Electronic)
Including images of bad gels.
Liquid volume too high.Follow the recommendati-
ons on the package.
IPG strips improperly sto-red.Always store IPG strips in the freezer,do not leave them on the bench at room temperature for too long time.
Basic part of the gel comes off during rehydration.Surface has been
damaged during removal
of cover film
Always start at the acidic,
pointed side to remove the
cover film.
Voltage too low(8kV not reached)Short strips(7cm and
11cm)are used,8kV is
reached in those strips
only with some samples
under exceptional
conditions.
Nothing to worry about.
Poor quality of urea and/
or thiourea.
Use high quality urea and
thiourea,remove ions with
a mixed bed ion exchan-
ger.
Too much salt in the
sample.
Remove salts by micro-
dialysis or precipitation,
replace PBS for cell wash-
ing with something non-
ionic.
TCA left in the sample
from precipitation.
Add more washing steps.
Bromophenol blue band stops and does not migrate completely into the anode.Too much salt in the
sample.
See above.
Strip starts to burn at a certain position.Too much salt in the
sample.
See above.
Visible brown band develops.Tris-chloride in the
sample and rehydration
solution.
Run the sample in a cup-
loading strip holder.
Cover fluid(paraffin oil) leaks out of the cup loa-ding strip holder during IEF.High protein and salt
load cause water
transport that carries the
oil with it.
Reduce the initial voltage
and prolong the first low-
voltage steps.
284A Trouble shooting