Lenstar及Pentacam与A 超测量白内障患者眼生物参数的比较

合集下载
相关主题
  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。

Lenstar及Pentacam与A 超测量白内障患者眼生物参数的
比较
张静;廉井财;张士胜;于青;周激波
【期刊名称】《国际眼科杂志》
【年(卷),期】2016(16)4
【摘要】目的:比较光学相干生物测量仪( Lenstar )、三维眼前节分析仪
( Pentacam)和A型超声测量仪( A-scan)在白内障患者中测量角膜中央厚度、角膜曲率、前房深度和眼轴长度结果的差异。

方法:分别用3种仪器对158例(158眼)白内障患者进行眼部生物测量,比较角膜中央厚度、角膜曲率、前房深度和眼轴长度的结果,采用重复测量的方差分析及Pearson相关分析,一致性比较采用Bland-Altman统计分析法。

结果:Lenstar 和 Pentacam 测得的角膜中央厚度分别为536.54±27.90μm和541.46±29.85μm,两者差异有统计学意义(t=-5.439;P<0.001);K值分别为43.87±1.45D和43.86±1.44 D,两者差异无明显统计学意义( t=-0.348, P>0.05)。

Pentacam、Lenstar和 A超测量的前房深度分别是2.73±0.38mm、2.71±0.38mm和2.85±0.40mm,三者差异均有统计学意义( F=309.94, P<0.001), Pearson 相关分析显示三者呈正相关( r=0.989 , 0.978, and 0.977;P<0.001),但变异系数较小( CV=3.12%)。

A超和Lenstar测量的眼轴长度分别是24.28±1.70 mm 和24.52±&nbsp;1.73mm,两者差异有统计学意义(t=-19.482, P<0.001, r=0.996;P<0.001)。

Bland-Altman分析显示,对于这几种眼前节参数,三种方法测量的一致性较好。

结论:三种仪器测量的结果尽管有一定差异,但Lenstar 和Pentacam的生物测量可重复性好,操作简便更易用于白内障患者的检查。

%AlM:To compare the central corneal thickness (CCT),
keratometry ( K ) reading, anterior chamber depth ( ACD) , and axial length ( AL) measured with Lenstar and Pentacam with those obtained with the ultrasound ( US ) pachymetry in the cataract patients.METHODS: A total of 158 eyes of 158 patients were examined in this study. The CCT, average K, ACD and AL obtained by Lenstar and/or Pentacam were compared with those obtained from US pachymetry using repeated-measures analysis of variance, Pearson correlation coefficients and Bland-Altman
analyses.RESULTS: The mean CCT obtained using Lenstar and Pentacam were 536.54±27.90μm and 541.46±29.85μm(t=-5.439;P<0. 001 ). The mean Km obtained using Lenstar and Pentacam methods were 43. 87±1. 45D and 43. 86±1. 44 D (t=-0. 348, P>0. 05). The mean ACD measured using the Pentacam, Lenstar, an d US pachymetry were 2. 73 ± 0. 38mm, 2.71±0. 38mm, and 2. 85±0. 40 mm, respectively (F=309.94, P<0.001), and they were positively correlated (r=0. 989, 0. 978, and 0. 977; P<0. 001) and the coefficient of variation was small (3. 12%). The mean AL obtained by US pachymetry and Lenstar were 24. 28 ± 1. 70mm and 24.52±1. 73 mm, respectively (t=-19. 482, P<0. 001, r=0. 996; P<0. 001). The Bland-Altman analysis showed that the three methods were comparable for CCT, Km, ACD and AL. CONCLUSlON: Although there were statistically significant differences, the measurements obtained by the Lenstar and the Pentacam were highly repeatable and the instruments easy to use.
【总页数】6页(P594-599)
【作者】张静;廉井财;张士胜;于青;周激波
【作者单位】200011中国上海市,上海交通大学医学院附属第九人民医院眼
科;200011中国上海市,上海瑞视眼科; 200025 中国上海市,上海交通大学医学
院附属瑞金医院眼科;200011中国上海市,上海瑞视眼科;200011中国上海市,
上海瑞视眼科;200011中国上海市,上海交通大学医学院附属第九人民医院眼科【正文语种】中文
【相关文献】
1.散瞳对Lenstar和A超测量白内障患者眼生物学参数的影响 [J], 罗书科;叶祖科;尹小芳;田思佳;卢彦
2.白内障患者Lenstar与A型超声或角膜曲率计眼部生物学参数测量结果的比较[J], 沈政伟;尹禾;薛林平;姜黎;刘洪;李丽;吴金桃
3.白内障患者Lenstar与A型超声或角膜曲率计眼部生物学参数测量结果的比较[J], 沈政伟;尹禾;薛林平;姜黎;刘洪;李丽;吴金桃;
4.Lenstar 900与接触性A超测量白内障患者眼轴的对比研究 [J], 叶伟明;张春玲;黄永权;师利农;倪梅;廖景丽;梁端儿
5.A型超声、IOLMaster和Pentacam测量白内障患者眼部生物学参数的比较 [J], 刘嫣;方思捷;姚钦科;王静;陆琳娜
因版权原因,仅展示原文概要,查看原文内容请购买。

相关文档
最新文档