Body Components
Body
Houses Suture, Needles and Lever Mechanisms
ProGlide Overview April 2007 LT2925874
12 © 2007 Abbott Laboratories
Suture Trimmer
Thumb Knob
5 © 2007 Abbott Laboratories
Heat-Formed Knot
Heat forming process creates permanent shape of knot in suture
ProGlide Overview April 2007 LT2925874
6 © 2007 Abbott Laboratories
Monofilament Suture
• Non-absorbable 3-0 polypropylene suture (Prolene) • Suture of choice by vascular surgeons due to:
– Minimized inflammatory response – High-knot tensile strength – Reduced perception of infection
ProGlide Overview April 2007 LT2925874
14 © 2007 Abbott Laboratories
Suture Trimmer
Allows suture to be placed coaxial to Suture Trimmer
ProGlide Overview April 2007 LT2925874
2 © 2007 Abbott Laboratories
Sheath
Hemostasis Valve Restricts the blood flow through the sheath while the sheath is in the artery.
Guide Wire Exit Port Directs the guide wire out of the sheath through the GW port
ProGlide Overview April 2007 LT2925874
8 © 2007 Abbott Laboratories
Body Components
Provides visual confirmation that device is correctly positioned in the vessel
`
Proglide
March 19,2010 Beijing
ProGlide Overview April 2007 LT2925874
1 © 2007 Abbott Laboratories
ProGlide Components
ProGlide Overview April 2007 LT2925874
© 2007 Abbott Laboratories
Suture Gate
15
Deployment Steps
ProGlide Overview April 2007 LT2925874
16 © 2007 Abbott Laboratories
Device Pre-Deployment Review
Opens Suture Gate when slid backwards
ProGlide Overview April 2007 LT2925874
13 © 2007 Abbott Laboratories
Suture Trimmer
Suture Trimmer Lever
Trims/Cuts suture above arteriotomy
ProGlide Overview April 2007 LT2925874
4 © 2007 Abbott Laboratories
Proximal Guide
Auto-Tie monofilament knot is housed in the distal Proximal Guide
ProGlide Overview April 2007 LT2925874
QuickCut™
Allows for sterile cutting of suture limbs
10
Body Components
Lever
Opens and Closes Foot
ProGlide Overview April 2007 LT2925874
11 © 2007 Abbott Laboratories
Marker Lumen
ProGlide Overview April 2007 LT2925874
9 © 2007 Abbott Laboratories
Body Components
ProGlide Overview April 2007 LT2925874
© 2007 Abbott Laboratories
Tapered Sheath 6F compatible
ProGlide Overview April 2007 LT2925874
3 © 2007 Abbott Laboratories
Distal Guide and Foot
Distal Guide • Houses the foot, suture bearing and O-ring • Designed to provide easy device insertion Foot • Provides tactile resistance against the arterial wall
Post implant
ProGlide Overview April 2007 LT2925874
© 2007 Abbott Laboratories
Knot at 28 Days
7
Body Components
HaBaidu Nhomakorabeadle
Lever Marker Lumen
Collar
Plunger Body QuickCut™