Goldmann眼压计与非接触眼压计测量青光眼患者眼压的对比分析

Goldmann眼压计与非接触眼压计测量青光眼患者眼压的对比分析

Goldmann眼压计与非接触眼压计测量青光眼患者眼压的对比分析李红锋;欧阳君;屈晓勇【摘要】• AlM: To compare the two methods for the measurement of glaucoma patients' intraocular pressure ( lOP) between Goldmann applanation tonometer ( GAT) and non-contact tonometer ( NCT) and find the laws of the two methods.•METHODS: The lOP of 108 glaucoma patients ( 206 eyes) were measured by GAT and NCT respectively.•RESULTS: ln 108 glaucoma patients, the average lOP of 206 eyes was 29. 77 ± 10.27mmHg by GAT and 24. 59 ± 8. 58mmHg by NCT. There was signifi cant difference between GAT and NCT (P<0. 01). The higher of lOP, the difference between GAT and NCT was greater.•CONCLUSlON: The measurement results with NCT were lower than that of GAT. The higher of lOP, the difference between GAT and NCT was greater. lt's better to measure lOP by GAT for the glaucoma patients, in order to avoid the misdiagnosis and mistreatment of glaucoma.%目的::对比分析 Goldmann 压平式眼压计( Goldmann applanation tonometer, GAT)和非接触眼压计( non-contact tonometer,NCT)测量青光眼患者眼压的结果,寻找其规律性。

电脑非接触眼压计对比表

电脑非接触眼压计对比表

电脑非接触眼压计对比表产品介绍:1. NIDEK NT-2000利用自动对焦功能进行检测,即使是初次使用也可以轻松掌握操作。

2. NT-2000具有独特的软气流设计:NT-2000在第一次读数后将会自动启动APC功能,根据眼压情况调整喷气量的大小,从而在保证测量准确的同时尽可能让患者感觉更舒适。

3. NT-2000的安全锁装置可避免触及患者眼球。

4. 在检测前,NT-2000会侦测眼睑的位置,避免眼睑睫毛干扰而影响测量精度。

NIDEK眼压计进入中国市场已有十年,目前全国各大医院几乎都有NIDEK眼压计在使用中。

NIDEK眼压计稳定的性能已得到广大医务人员的认可。

功能参数:测量范围:1-60毫米贡柱(1毫米贡柱间隔)APC40.APC60工作距离;11毫米(喷头到角膜顶点)对位模式:一个对位光点和对焦显示内固视灯:绿色LED操纵柑工作范围:前/后36毫米左/右86毫米上/下28毫米对位模式:(对位)(对焦)(喷气)测量范围:1-60毫米贡柱(1毫米贡柱间隔)APC40.APC60工作距离;11毫米(喷头到角膜顶点)对位模式:一个对位光点和对焦显示内固视灯:绿色LED操纵柑工作范围:前/后36毫米左/右86毫米上/下28毫米对位模式:(对位)(对焦)(喷气)(自动) 自动手动自动(手动) 手动手动自动测量范围:1-60毫米贡柱(1毫米贡柱间隔)APC40.APC60工作距离;11毫米(喷头到角膜顶点)对位模式:一个对位光点和对焦显示内固视灯:绿色LED操纵柑工作范围:前/后36毫米左/右86毫米上/下28毫米对位模式: (对位)(对焦)(喷气)(自动) 自动手动自动(手动) 手动手动自动适用范围:测量范围:1-60毫米贡柱(1毫米贡柱间隔)眼科最基本视功能检查项目之一.适用于眼外伤,青光眼,眼底病,白内障,验光,以及原因不明性眼压升高.零配件:标准配件:保险丝打印纸颌托纸颌托纸固定针电源线防尘罩脉冲探测器顶盖。

不同角膜厚度下非接触眼压计与Schiotz眼压计测量眼压的比较

不同角膜厚度下非接触眼压计与Schiotz眼压计测量眼压的比较

不同角膜厚度下非接触眼压计与Schiotz眼压计测量眼压的比较安晓巨;陈开建【摘要】目的:比较不同角膜厚度下非接触眼压计与Schiotz眼压计测量结果的差异.方法:对314例627眼分别用角膜测厚仪测量角膜厚度,非接触眼压计与Schiotz 眼压计测量眼压,根据角膜厚度分为:较薄组(角膜厚度≤530μm)、正常组(530μm<角膜厚度≤570μm)、较厚组(角膜厚度>570μm).采用配对t检验比较不同角膜厚度下两种测量结果的差异,并通过Bland-Altman进行一致性评价.结果:角膜厚度较薄组170眼,正常组301眼,较厚组156眼.两种方法在较薄组、正常组、较厚组三组中的眼压值分别为12.82±2.67,12.84±2.37mmHg;13.67±2.66,13.58±2.41mmHg;15.45±2.91,14. 76±2.39mmHg.两种方法测量结果在角膜厚度正常和偏薄组中差异无统计学意义,偏厚组中差异有统计学意义.结论:在角膜厚度≤570μm的人群中,眼压的检查可在非接触眼压计和Schiotz眼压计间选择一种检查;对于角膜厚度>570μm的人群,需参考两种方法检查结果,进一步检查排除青光眼.【期刊名称】《国际眼科杂志》【年(卷),期】2012(012)007【总页数】3页(P1337-1339)【关键词】非接触眼压计;Schiotz眼压计;眼压;角膜厚度【作者】安晓巨;陈开建【作者单位】400042,中国重庆市,第三军医大学大坪医院野战外科研究所眼科;400042,中国重庆市,第三军医大学大坪医院野战外科研究所眼科【正文语种】中文0 引言眼压是诊断青光眼的重要参考指标之一。

因此,在行近视激光手术、白内障等手术前,均需要测量眼压。

非接触眼压计与Schiotz眼压计是临床常用的眼压检查方法。

由于非接触眼压计不与角膜接触,避免Schiotz眼压计检查中的麻醉不适、角膜损伤和感染的风险,已广泛用于眼压检查。

Diaton眼压计测量准分子激光原位角膜磨镶术后眼压

Diaton眼压计测量准分子激光原位角膜磨镶术后眼压

Diaton眼压计测量准分子激光原位角膜磨镶术后眼压作者:刘磊,栗静,李新宇,王渝龙,徐曼【摘要】目的:通过与非接触式眼压计(non contact tonometer, NCT)的比较,评价Diaton眼压计(DT)在准分子激光原位角膜磨镶术(laser in situ keratomileusis, LASIK)前后眼压测量中的应用价值。

方法:在LASIK术前及术后1mo,分别使用NCT和DT测量患者眼压,使用SPSS 13.0统计软件对所得结果进行统计学分析。

结果:术后1mo时,NCT眼压测量值较术前下降(5.76±2.31)mmHg,DT眼压测量值则无显著差异(t=0.70,P=0.487);NCT与DT读数测量值差值与术中切削量呈正相关,与术后中央角膜厚度呈负相关。

结论:Diaton测量眼压不受角膜厚度的影响,因此更适用于LASIK术后患者的眼压测量。

【关键词】 Diaton眼压计;非接触式眼压计;眼压测量;角膜磨镶术;眼压Abstract AIM: To evaluate the performance of Diaton tonometer(DT) before and after laser in situ keratomileusis(LASIK) by comparing it with non contact tonometer (NCT). METHODS:NCT and DT were used to measure the intraocular pressure before and 1 month after LASIK, the results were analyzed with SPSS One month postoperatively, NCT readings declined by 5.76±2.31mmHg, DT readings did not differsignificantly(t=0.70,P=0.487); the difference between NCT and DT readings had positive correlation with declined corneal thickness, and had negative correlation with the central corneal thickness postoperatively. CONCLUSION: DT readings were not affected by corneal thickness; therefore, Diaton tonometer may be a better choice for the measurement of intraocular pressure after LASIK.KEYWORDS:Diaton tonometer; non contact tonometer; intraocular pressure measurement; keratomileusis; intraocular pressure0 引言准分子激光原位角膜磨镶术是目前开展最为广泛的角膜屈光手术,它通过切削角膜基质和改变角膜曲率来矫正视力,然而术后角膜厚度的下降导致压平式眼压计测量值较真实眼压偏低[13],如何准确测量LASIK术后眼压是现阶段亟需解决的问题之一。

眼压计及其选购

眼压计及其选购

眼压计及其选购王英丽摘要:眼压计是定量测量眼压的专用仪器,未来会向便携性、非接触性、高精度性方面不断创新发展。

本文具体描述了目前几款常用眼压计的分类、测量原理并对比了各自优缺点。

同时,结合目前眼压计市场情况,通过分析其性能指标及参数,对其中应用最普遍的喷气式眼压计的选购方法进行了说明。

关键词:眼压计;分类;工作原理;选购方法眼睛是以光作为适宜刺激的视觉生物器官,是一种复合的光学系统。

在几何外形上,眼球可以近似认为是一个充满房液的密闭球体,眼球内部的压力叫做眼内压(IOP),简称眼压,是指包括房水、玻璃体、晶状体和血液在内的眼球内容物对于眼球壁所产生的压力。

眼球为了完成其正常视觉功能,必须维持内压高于大气压的一定水平才能保持形状的稳定。

现代眼科学已证实,人的眼压如同血压一样,是24小时波动的,掌握眼压的变化规律对于眼科疾病的诊断具有重要价值,是用于诊断眼科疾病、观察病情、估测预后、评价疗效等的重要手段和指标之一。

对于眼压过高或过低的病人,实施眼压监测是非常重要的。

图1 眼压计结构图眼压计(如图1)是定量测量眼压的专用仪器,它的工作原理是通过角膜形状变化或直接测量角膜血流脉动压力变化,换算获得眼内压。

到目前为止,已经有数十种眼压计相继问世,在众多的眼压计中,大致可分为压陷式眼压计、压平式眼压计两类,压平式眼压计又可分为接触式压平眼压计和非接触式压平眼压计两大类。

现常用的是非接触式压平眼压计,具有操作简便、测量迅速等优点。

1 眼压计的分类及对比1.1 压陷式眼压计压陷式眼压计起步于20世纪70年代,是测量人体眼压的一种专用仪器,Schiotz 眼压计(SIT)是压陷式眼压计的典型代表。

其工作原理是用一定重量的眼压测杆使角膜压成凹陷,在眼压计重量不变的条件下,压陷越深,其眼压越低,其测量值受眼球壁硬度的影响,智能型的压陷式眼压计通过光机电技术结合,可实现眼压的自动化测量。

压陷眼压计的优点是其结构简单、价格便宜、操作简便, 所以至今仍是临床上主要使用的眼压计之一。

非接触式眼压计与回弹式眼压计在眼压测量中的准确度分析

非接触式眼压计与回弹式眼压计在眼压测量中的准确度分析

非接触式眼压计与回弹式眼压计在眼压测量中的准确度分析包宇涵;律鹏;张文芳;吴万民【摘要】目的:分析非接触式眼压计(non contact tonometer,NCT)、ICare回弹式眼压计(ICare rebound tonometer,RBT)在青光眼患者中眼压测量结果的准确性.方法:回顾性病例对照研究.选取青光眼患者113例185眼.用RBT、NCT和Goldmann眼压计(GAT)分别于9:00、16:00两次进行眼压测量并记录为两组数据,分别用秩和检验、国际标准8621指南评估其准确性,Bland-Altman一致性分析两组测量结果的一致性,并以GAT测量值作为标准将数据分为异常眼压组(<10mmHg或>21mmHg)和正常眼压组(10~21mmHg),在不同组间分析其一致性.应用Spearman相关分析眼压计间的相关性.结果:GAT、NCT和RBT三组测量值间存在差异(P<0.01);GAT获得的眼压测量值与RBT和NCT测量值间存在相关性(rs=0.71、0.77,P<0.001).NCT与GAT测量值接近(P=0.92),而RBT与GAT相比,眼压测量值较高(P<0.05).然而NCT、RBT同GAT相比一致性界限范围均较宽,分别为-6.2~6.0、-5.2~7.6mmHg,在正常眼压组一致性界限范围分别为-5.9~5.9、-4.3~7.5mmHg,在非正常眼压组分别为-7.3~6.4、-7.5~5.6mmHg.根据ISO 8612指南进行评价,NCT和RBT在三组中超过95%的一致性限制的异常值分别为3.9%、11.3%、12.2%和26.3%、11.3%、12.2%.结论:NCT和RBT均不能简单代替GAT用于青光眼患者眼压的测量.在青光眼患者中,随着测量值偏离正常眼压范围,NCT和RBT的测量误差也有所增大.【期刊名称】《国际眼科杂志》【年(卷),期】2019(019)008【总页数】5页(P1406-1410)【关键词】青光眼;眼压;非接触式眼压计;回弹式眼压计;Goldmann眼压计【作者】包宇涵;律鹏;张文芳;吴万民【作者单位】730000 中国甘肃省兰州市,兰州大学第二临床医学院;730030 中国甘肃省兰州市,兰州大学第二医院眼科;730030 中国甘肃省兰州市,兰州大学第二医院眼科;730030 中国甘肃省兰州市,兰州大学第二医院眼科;030002 中国山西省太原市,山西省眼科医院眼科【正文语种】中文0引言青光眼是一组以特征性视神经萎缩和视野缺损为共同特征的疾病,病理性眼压升高是其主要危险因素。

三种眼压计计量浅析

三种眼压计计量浅析

技术交流Technical Exchange上渎并量倒試Technica—E三种眼压计计量浅析X■徐聪恩/宁波市计量测试研究院90引言眼球内部的压力叫做眼内压(IOP),简称眼压,眼压是指包括房水、玻璃体、晶状体和血液在内的眼球内容物对于眼球壁所产生的压力。

眼压大小是眼科医生判定眼睛健康的重要标志,是用于诊断眼科疾病、观察病情、估测预后、评价疗效等的重要手段和指标之一。

眼压计是定量测量眼压的专用仪器,在1987年发布和1999年调整实施的《中华人民共和国强制检定的工作计量器具明细目录》中,眼压计都属于强制检定的工作计量器具。

到目前为止,已经有数十种眼压计相继问世,在众多的眼压计中,大致可分为压陷式眼压计、接触式压平眼压计和非接触式眼压计三类。

1压陷式眼压计压陷式眼压计"7(impression tonometers,IT)是测量人体眼压的一种专用仪器,Schiotz眼压计(SIT)是压陷式眼压计的典型代表。

其工作原理是根据一定质量的压针对角膜的压陷深度而间接测量眼压,分度标尺上的每一格相当于1/20mm的角膜压陷深度,根据压陷深度值及压针和砥码的质量,查眼压换算表获得眼压值。

智能型的压陷式眼压计通过光机电技术结合,实现眼压的自动化测量。

压陷式眼压计从20世纪70年代开始起步,上海市计量测试技术研究院根据压陷式眼压计的工作原理和结构特点研制了压陷式眼压计检定装置,起草了JJG574-2004《压陷式眼压计》检定规程,该检定装置由三台测量仪器(专用天平、专用测微计和倾斜仪组成),分别测量眼压计的质量、位移和角度三个参数。

随着技术的不断发展,通过机电结合的方式,在一台检定装置上完成了压陷式眼压计的自动测量。

2接触式压平眼压计接触式压平眼压计13-41(applanation tonometers, AT)工作原理是基于Imhert-Fiek原理,将眼球视为一个理想的表面干燥的弹性薄壁圆球,则内部眼压p、作用力F和角膜被压平面积A具有如下关系:P=£⑴Goldmann眼压计(GAT)是接触式压平眼压计的典型代表,测量时通常与裂隙灯配合使用,用压平眼压计测量头压平一定面积的角膜,根据所需的作用力来测定眼压。

Goldmann眼压测量与眼压计的几点介绍

Goldmann眼压测量与眼压计的几点介绍

眼压是诊断与治疗青光眼的一个必不可少的手段。

眼压测量方法有眼压计法、直接检测法(液体压力计)及指压法。

目前临床上常用的眼压计有Goldmann压平眼压计、Perkins手持眼压计、Tono-Pen眼压计、Proview眼压监测计与非接触式眼压计等。

一、Goldmann压平眼压计1.设计原理Goldmann压平眼压计是国际上用以测量眼压的“金标准”眼压计,它是利用测压头压平角膜来进行间接的眼内压测量。

根据Imbert-Fick原理:Pt(眼内压)=W(压平角膜的外力)/A(压平面积)而推算的。

Goldmann眼压计的直径为3.06mm,当测压头使角膜压平使7.35㎜2的环形面积所需的力即为眼压测量值。

若需1g的力量加在测压头上,达到7.35㎜2环形面积时,眼压为10mmHg(1克重≈1cm3H2o=1000mm3H2o ,Pt=W/A=W/πr2=1000mm3H2o/[3.14×(3.06/2)2]= 1000mm3H2o/ 7.35mm2=135.98mmH2o=135.98/13.6(mmHg)=10mmHg),依此类推。

2.Goldmann压平眼压计的结构(1)测压头:为一表面平滑的透明塑料柱,前端可直接接触角膜,作压平角膜用,压平面直径为3.06mm。

后端固定于测压杠杆末端的金属环内。

测压头内有两个基底相反的三棱镜,故能使与角膜接触处的环形物象移位成为两个半圆环形。

另在测压头前端侧面上有径线刻度,供测量高度散光时,作轴向定位用。

(2)测压装置:为一能前后移动的杠杆,其移动度受内部安装的弹簧控制,弹簧的张弛力可被一测压螺旋调整。

在测压螺旋表面有以克重量为单位的重力刻度,表示弹簧的张力(克重量),范围由0至8g(即相当于0至80mmHg)。

Goldmann眼压计有“悬吊式”(T900型)与“座式” (R900型)两型。

操作时只需捻转测压螺旋,在裂隙灯显微镜观察下,当角膜压平面达3.06mm直径时,所需压力即眼压值。

手持式非接触眼压计测量值准确性的比较

手持式非接触眼压计测量值准确性的比较

手持式非接触眼压计测量值准确性的比较
陈秀琦;彭大伟
【期刊名称】《眼科学报》
【年(卷),期】1995(011)002
【摘要】对手持式非接触眼压计测量眼压的准确性及可靠性进行评价。

分别采用keeler,手持非接触眼压和R900,Goldmann压平眼压计对58例患者(113只眼)的眼内压进行了测量并对两种测量方法的测量结果进行了比较。

采用手持非接触眼压计测得的平均眼压为16.31±5.59mmHg(1mmHg=0.1333kPa),采用Goldmann压平眼压计测得的平均眼压值为17.49±6.13mmHg,经统计学t检
【总页数】3页(P86-88)
【作者】陈秀琦;彭大伟
【作者单位】不详;不详
【正文语种】中文
【中图分类】R770.426.2
【相关文献】
1.非接触眼压计与修氏眼压计测量值的比较及影响因素分析 [J], 王静平;李德姣
2.3 种根测仪定位根管侧穿孔准确性的比较研究 [J], 李月恒;周智;罗俊;李锐;陈娇
3.非接触眼压计与修氏眼压计测值比较 [J], 江滨;曾庆莹
4.3种根测仪测量根管工作长度准确性的比较 [J], 魏雪;李袁媛;张国琴;钟晓波;齐

5.国产手持式压平眼压计与非接触眼压计测量结果比较 [J], 俞振飞
因版权原因,仅展示原文概要,查看原文内容请购买。

动态轮廓眼压计与非接触眼压计的比较

动态轮廓眼压计与非接触眼压计的比较

Z HUA—i. w t州 D ^^ i n b m pl
T d M o y. eu g In 3 0 0 . n4 0 3 I l
叼 o i o i l r, i d a ol e Hu.o g( 矗 -r, s t o i lC lg - ah ̄ , 们  ̄ H pa  ̄ Me c e r . 刀
&i n e ad  ̄e
【 A
Байду номын сангаас
】 O jcv T opr te i r c fn ou r r sr m au db O cn c t o t be i t e oc a f e eo i m el e u es yNN— ot t o  ̄ e m eh d e n t aps e a n r
【Cr w s h g e 砌 ' a ih rt l w i h me r d b C h c 删 e y N T.T e if e c fc nr l O l hc n s 1 t eme s r me t o t e - h n u n eo e t mCI t i k e s O a u e n s fi mo u l aC 1 1h n
关, 相关系数 r:O 3 8 .4 2(P < .1 。D T测量 值与 c r无 相关 性 , OO ) c c 相关 系数 r:0 0 3 , >0 0 ) . 65 P .5 。结论 N 和 D T测 量眼压值有 差异 , C cr C D T测量值 大于 N , cr N T的影响大于 D T cr c 对 C C。
【 关键词】 眼内压; 中央角膜厚度; 动态轮廓眼压计; 非接触眼压计
[ 临床眼科杂志 .0 7 1 :1 ] 2 0 .5 5 9
C‘ p rs n o  ̄ a cc no ra d n n c 吡 t 岫 ee 脚 a io fd ml o tu n o o ∞ tr LULi A I e-W Ne 一 J - Xi-u- J - ny i

三种眼压计与Pentacam眼压校正系统的应用比较

三种眼压计与Pentacam眼压校正系统的应用比较

三种眼压计与Pentacam眼压校正系统的应用比较赵剑;孟觉天【摘要】目的:比较非接触式眼压计(NCT)、Goldmann压平式眼压计(GAT)、动态轮廓眼压计(DCT)和Pentacam三维眼前节分析系统四种方法检查眼压的差异,并研究它们与中央角膜厚度(CCT)和中央角膜曲率(CCV)的关系.方法:对拟行LASIK 手术的143例286眼患者用Pentacam系统测量CCT和CCV后,分别用NCT,GAT和DCT测量眼压,并用Pentacam系统的Ehlers,Shah,Dresden,Orssengo/Pye和Kohlhaas 5种眼压校正方法对NCT眼压值进行校正,将测量结果进行方差分析、相关性及线性回归分析.结果:DCT值最高,DCT与其他方法之间均有显著性差异(P<0.01); GAT与DCT,Kohlhaas,Shah 之间有差异(P<0.05),与其他方法之间无差异; 各组眼压值之间有较好的相关性,GAT与DCT的相关性最强(r=0.702); NCT受CCT影响最大,GAT受CCV影响最大,DCT,Dresden,Orssengo/Pye,Shah与CCT和CCV均不存在相关性(P>0.05).结论:对CCT和CCV正常的健康人群进行青光眼筛查时建议使用压平式眼压计,对可疑的患者,尤其是CCT和CCV偏离正常的患者,宜进一步行DCT检查或者用Pentacam系统进行眼压校正.【期刊名称】《国际眼科杂志》【年(卷),期】2011(011)010【总页数】4页(P1726-1729)【关键词】动态轮廓眼压计;压平式眼压计;Pentacam分析系统;对比研究【作者】赵剑;孟觉天【作者单位】221003,中国江苏省徐州市,徐州医学院附属第三医院眼科;221002,中国江苏省徐州市第一人民医院眼科,徐州市眼病防治研究所【正文语种】中文0 引言眼压测量是诊断和筛查青光眼最重要的依据。

常见眼压计对比分析

常见眼压计对比分析

常见眼压计对比分析眼压计是用来测量眼压的仪器,通常用于诊断青光眼等眼部疾病。

随着科技的进步,眼压计也不断发展,不同类型的眼压计有着各自的特点和优势。

下面将对常见的眼压计进行对比分析。

1.电子眼压计电子眼压计使用电子技术和传感器来测量眼压,通常具有数字显示屏和内置计算功能。

它们具有测量准确、速度快、使用便捷等优点。

同时,电子眼压计可以记录测量结果,便于医生进行分析和追踪观察。

然而,电子眼压计的价格较高,并且需要电池供电。

2.气压眼压计气压眼压计是一种常见的眼压测量仪器,它使用气体力量来测量眼内压力。

气压眼压计通常具有一个软尖的头部,当它与眼睛接触时会释放气体。

根据眼压的弹性变化,仪器可以测量出眼压值。

气压眼压计价格相对较低,但测量结果可能受到外界因素的影响,如正确的仪器使用和患者的配合度。

3.应变传感眼压计应变传感眼压计是一种常用的眼压测量仪器,它使用微小的变形传感器来测量眼压。

这些传感器通常与角膜接触,并测量角膜的形状变化来间接测量眼压。

应变传感眼压计具有比气压眼压计更准确的测量结果,但它们需要专业人员来进行操作,并且价格相对较高。

4.穿透眼压计穿透眼压计是一种新型眼压测量仪器,它利用光学技术来测量眼内压力。

这些仪器通常通过测量眼底血管的扩张和收缩来间接测量眼压。

穿透眼压计具有无需接触眼球、测量快速、无痛无创等优点,但它们的价格较高,并且测量结果可能受到患者眼部情况的影响。

综合来看,不同类型的眼压计各有优缺点。

对于患者来说,选择合适的眼压计应根据医生的建议和个人的情况来决定。

对于医生来说,他们需要综合考虑测量准确性、使用便捷性、价格等因素来选择适合的眼压计。

无论使用何种类型的眼压计,都需要专业人员进行操作和解读测量结果,以确保准确的诊断和治疗。

总结起来,不同类型的眼压计各有优劣,电子眼压计准确性高但价格昂贵;气压眼压计价格低廉但结果可能受到外界因素影响;应变传感眼压计准确但需要专业人员操作;穿透眼压计无创且快速但价格高且受到个体差异影响。

非接触眼压计与修氏眼压计测值比较

非接触眼压计与修氏眼压计测值比较

非接触眼压计与修氏眼压计测值比较
江滨;曾庆莹
【期刊名称】《眼科新进展》
【年(卷),期】1998(018)001
【总页数】2页(P41-42)
【作者】江滨;曾庆莹
【作者单位】广东省深圳南山人民医院眼科;广东省深圳南山人民医院眼科
【正文语种】中文
【中图分类】R770.2
【相关文献】
1.非接触眼压计与修氏眼压计测量值的比较及影响因素分析 [J], 王静平;李德姣
2.非接触眼压计与修氏眼压计测量结果的比较 [J], 李德胜;朱淑琴;葛春英;席小凤
SIK术后修氏和非接触眼压计测量结果的比较 [J], 余野;王勤美;许琛琛
4.修氏眼压计与非接触眼压计测量结果的比较 [J], 潘秀云;刘大川
5.非接触式眼压计与修氏眼压计检测正常人眼压值比较 [J], 许永春;许永根;姜燕因版权原因,仅展示原文概要,查看原文内容请购买。

非接触式眼压计与压陷眼压计测量眼压的比较

非接触式眼压计与压陷眼压计测量眼压的比较

非接触式眼压计与压陷眼压计测量眼压的比较
杜敏
【期刊名称】《临床医药实践》
【年(卷),期】2007(016)008
【摘要】非接触式眼压计(NCT)因其独特的优越性而在临床上广泛应用,但其测量眼压的准确性尚需进一步证实.目前基层医院多用压陷眼压计测量眼压.本文就NCT 和压陷眼压计测量眼压值进行比较,以探求NCT测量眼压的精准程度,为临床上评估NCT测量眼压值提供依据.……
【总页数】2页(P761-762)
【作者】杜敏
【作者单位】晋中市第二人民医院,山西,太谷,030800
【正文语种】中文
【中图分类】R77
【相关文献】
1.非接触眼压计与Schiotz压陷式眼压计测量眼压结果比较
2.Diaton眼压计与非接触式眼压计测量眼压的比较
3.非接触式眼压计与Goldmann压平眼压计测量眼压的比较
4.非接触式眼压计与Goldmann压平眼压计测量眼压的比较
5.非接触式眼压计与压陷式眼压计临床应用观察
因版权原因,仅展示原文概要,查看原文内容请购买。

三种眼压计测量高眼压儿童眼压及其与中央角膜厚度的关系

三种眼压计测量高眼压儿童眼压及其与中央角膜厚度的关系

三种眼压计测量高眼压儿童眼压及其与中央角膜厚度的关系刘俐娜;陈海波;徐力;史贻玉【期刊名称】《国际眼科杂志》【年(卷),期】2018(18)10【摘要】目的:探讨经Topcon眼压计(non-contact tonometry,NCT)初筛高眼压的屈光不正儿童的中央角膜厚度(central corneal thickness,CCT)分布,以及NCT、Goldmann压平眼压计(Goldmann applanation tonometry,GAT)与OCULUS Corvis ST角膜生物力学分析仪(CST)所测眼压值的差异,并分析眼压和CCT的相关性.方法:选择经NCT测量单眼或双眼眼压高于21 mmHg的非青光眼屈光不正儿童39例78眼,用CST和GAT再次进行眼压测量,两种仪器的测量顺序随机,同时用CST测量CCT并根据CCT进行眼压校正.根据CCT分成正常组和较厚组,分析经NCT初筛高眼压的屈光不正儿童的总体CCT分布情况,探讨三种仪器所测眼压值的不同及其与CCT的相关性.结果:正常CCT组,CST低于GAT和NCT的眼压测量值,差异有统计学意义(F=5.12,P=0.01);较厚组,三种眼压测量方式的眼压测量值比较,差异均有统计学意义(F=15.72,P<0.001).NCT和GAT眼压值与CCT呈明显正相关(rNCT=0.298,PNCT=0.04;rGAT=0.408,PGAT=0.01);CST校正眼压值与CCT无明显相关性(rCST=0.062,PCST=0.593).结论:NCT初筛高眼压的屈光不正儿童的CCT偏厚,CST校正眼压低于NCT和Goldmann眼压,NCT和GAT所测眼压与CCT呈正相关;对于CCT偏厚的儿童患者,CST校正眼压比NCT和GAT更接近于真实眼压值.【总页数】3页(P1890-1892)【作者】刘俐娜;陈海波;徐力;史贻玉【作者单位】570311 中国海南省海口市,中山大学中山眼科中心海南眼科医院海南省眼科医院海南省眼科学重点实验室;570311 中国海南省海口市,中山大学中山眼科中心海南眼科医院海南省眼科医院海南省眼科学重点实验室;570311 中国海南省海口市,中山大学中山眼科中心海南眼科医院海南省眼科医院海南省眼科学重点实验室;570311 中国海南省海口市,中山大学中山眼科中心海南眼科医院海南省眼科医院海南省眼科学重点实验室【正文语种】中文【相关文献】1.TonoLab眼压计在遗传性小鼠高眼压模型中眼压测量的可重复性 [J], 刘丽芳;黄楚开;张铭志;2.中央角膜厚度对3种眼压计眼压测量值的影响 [J], 周和政;张绍阳3.非接触眼压计回弹式眼压计及Goldmann三种眼压仪测量眼压的对比研究 [J], 张莉;李双农;范永4.CorvisST与Pentacam及Topcon眼压计测量中央角膜厚度、眼压结果的一致性观察 [J], 韩晓彤;赵海霞;关文英;王召格;张凤云;黄武多;孙怡5.中央角膜厚度校正前后三种眼压计眼压测量值的一致性分析 [J], 张绍阳;周和政;江文珊;张文强;陈云辉;袁敏因版权原因,仅展示原文概要,查看原文内容请购买。

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。

重复测量值
r
p
右眼 17.49±8.17
17.69±8.11 0.986 0.287
1
56
左眼 16.98±5.69
16.96±5.63 0.969 0.925
右眼 18.82±8.80
18.82±8.76 0.990 1.000
2
52
左眼 17.54±5.82
17.48±5.99 0.979 0.733
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
Icare回弹式眼压计临床应用
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
虽然ICare眼压计在眼压测量中优势明显。受角膜状 况影响较小,可进行卧位测量,大多数儿童也能很 好配合。但是眼压测量的稳定性及准确性就成为了 临床工作者关注的焦点。 下面我们引用一篇2011年中华实验眼科杂志上的论 文《iCare回弹式眼压计测量的重复性及与 Goldmann压平式眼压计测量值间的一致性研究》 对ICare眼压计测量的稳定性及准确性做进一步论述。
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
对比结果
• 3.不同眼压区间iCareRBT与GAT测量值差 值分布范围的统计
不同眼压范围内,iCareRBT与GAT测量值差值的均值均<1mmHg,表 明2种测量方法的一致性较好。在高眼压段标准差增大,测量的一致性有 所下降
压平式眼压计
• 压平式眼压计是通过外力将角膜压平来测 量眼压。压平式眼压计测量眼压时,使角 膜凸面稍稍变平而不下陷,眼球容积变化 很小,因此不受眼球壁硬度影响,可认为 P0=Pt
①变力压平眼压计:固定压平面积,看压平该面积所需力的 大小判断眼压高低。如Goldmann眼压计、非接触眼压计、 Perkins眼压计、Ton-Pen眼压计和气动式压平眼压计。 ②恒力压平眼压计:力量一定,根据压平面积判断。使用较 少,如Maklakoff压平眼压计
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
对比分组
考虑到不同眼压计对眼压的影响,按顺序依 次进行NCT,RBT和GAT的测量。按GAT测量 将患者分为低眼压组(≤21mmHg)和高眼 压组(>21mmHg),对比分析各种眼压计 的测量结果。用超声测厚仪测量角膜厚度 CCT。
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
测量不准的常见误差来源以下几个方面:
1.患者由于紧张挤眼或频繁眨眼 2.喷气刺激导致的泪液分泌增多,读数偏高 3.患者配合不佳,眼的注视方向不对,则需 较大力量使角膜压平。 4.角膜表面不光滑,反射不佳影响读数 5.角膜混浊、高度散光和固视不佳
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
眼压计种类
• 压陷式眼压计 如Schiøtz眼压计
• 压平式眼压计 如Goldmann眼压计、非接触式眼压计
• 回弹式眼压计 如Icare回弹式眼压计
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
压陷式眼压计
• 利用眼压计砝码的重量使角膜压成凹陷, 在压的重量不变的条件下,压陷越深其眼 压越低。由于砝码重量将角膜压出凹陷, 常受到眼球壁硬度的影响,也对眼球容积 的影响较大。
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
Goldmann压平眼压计
• 国际公认的标准眼压计 “金标准”
固定压平面积,直径选定为3.06mm, 面积为7.354mm,由于压平面积很小, 眼球容积改变也很小,仅0.56mm3, 使原始眼压升高3%,影响很小, 因此不受眼球壁硬度影响。 从理论上讲Pt=P0
左眼 17.22±4.82 17.17±4.42 -0.06±2.24 0.887 0.881
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
对比结果
• 2.每位操作者对患者重复测量的iCareRBT 测量值
两位检查者两次测量的眼压值间差异均无统计学意义
(P>0.05)
检查者 例数 眼别 首次测量值
• 眼压计测量
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
眼压计测量
• 原理:利用力的平衡原理,以不同类型的眼压计 作用于眼球壁上,根据所用的重量压平的角膜面 积或压陷的深度测量出的眼球壁的张力,得到的 是间接的眼压。
lgPt=lgP0+0.0215V 式中Pt为眼压计测得值,P0为真正的眼内值,V为 眼球容积和改变量(被压平下得体积),0.0215为 巩膜硬度(弹性)系数。
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
对比结果
• 高、低眼压组眼压测量眼压测量值比较
高眼压组RBT和NCT与GAT的平均偏离值较大
组别
GAT≤21mmHg(n=48) GAT>21mmHg(n=14)t值
p值
RBT-GAT
0.46±2.34
-3.07±3.83
-3.273 0.005
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
Schiøtz眼压计
• 根据所用的重量使角膜压陷的深度来间接测得眼压。压陷 角膜的程度可通过指针在刻度尺上显示。砝码分别为5.5g、 7.5g、10g和15g,一定重量的砝码加压在角膜上,压陷 越多,读数越高,所测眼压越低。然后通过查表得出眼压 值。
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
• 优点
操作简单,携带方便
不用麻醉,患者几乎没 有感觉
一次性针头头避免交叉 感染
可以兼顾坐位及卧位测 量
适用于异常角膜测量
可测量范围大
• 缺点
耗材偏贵
对于配合不佳患者误差 较大
非正常范围眼压测量偏 高
易受角膜粘滞度影响
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
非接触眼压计 临床最常用的眼压计
通过气流来压平角膜恒定面积, 测量压头不与眼球接触。气体 压平角膜至设计直径时,角膜 作为”反光镜“把协调光线反射 入接收器的光量最多。根据气 体压平角膜的时间与眼压成比 例原理,处理器将压平时间转 换成眼压值,以数字形式显示 出来。
对比结果
• 3种眼压测量值比较
3种眼压计测量值两两之间比较无统计学意义
组别
眼数/只
平均差值
t值
(mmHg,χ±s)
RBT-GAT
62
RBT-NCT
62
NCT-GAT
62
-0.34±3.09 0.65±2.98 -0.98±4.02
0.862 1.707 1.928
p值
0.392 0.093 0.059
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
• 优点 测量“金标准”
• 缺点 操作繁琐 对配合欠佳者及角膜情 况异常者误差较大 较难用于儿童测量 仅能用于坐位测量
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
测量不当也会导致误差,误差主要来源于以下几个方面:
1.检查者对眼球施加压力或患者在测量时挤眼,所测眼压偏 高,告诉患者尽量放松。 2.检查时最好嘱患者双眼注视5~6米处,尽量减少调节影响。 3.眼睑在水平位上方<15度,所测眼压偏高,应尽量将上下 睑固定在上下眶缘,不要挤压眼球。如果睑裂过宽也会使测 得眼压偏高。 4.荧光素对泪液膜染色不够充分,所测得眼压可能偏低,可 能由于染色时间太长,荧光素半环太窄,应在染色后1分钟 内测量。泪液过多,荧光素半环太宽,眼压可能偏高。 5.角膜白斑,不规则或水肿使荧光素半环观测不清。 6.角膜散光度数>3D,顺规散光时眼压偏低,逆规散光时眼 压偏高,每3D散光约产生1mmHg偏差。 7.角膜中央厚度。
NCT-GAT
-0.17±3.14
-3.79±5.41
-2.389 0.030
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
讨论
本研究果显示 ①3种眼压计测量值两两之间比较差异无统计学意义,说明 RBT,GAT和NCT的眼压测量值之间具有良好的一致性。 ② 高 眼 压 组 R BT 和 N C T 的 测 量 值 偏 高 , 分 别 比 G AT 偏 高 (3.07±3.83)mmHg和(3.79±5.41)mmHg。 ③RBT、GAT、NCT测量值随CCT增厚而增大目前多数研究 认为RBT与GAT一样也受CCT影响。但通过多因素分析发现, RBT测量值仅与角膜黏滞性及角膜阻力因子相关,与CCT无 关。
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
对比结果
• 1.不同操作者iCareRBT测量值的重复性分 析
2位检查者间右眼和左眼眼压测量值的比较差异无统计学意
义(P>0.05)
眼别
Icare RBT测量值
r
p
检查者1
检查者2
差值
右眼 17.20±4.60 17.80±5.74 -0.60±2.16 0.937 0.109
常用眼压计对比分析
三六三医院 特检技师 徐斌
国家公立三级医院西南医科大学附属医院华西医院区域联盟法:将一个套管和针头直接插入前房,另一端 与液体压力计连接,直接测量眼压。用这种方法测量的 眼压很精准,但临床上并不适用,仅用于动物实验。
• 指测法:又称指诊眼压测量法。这种方法适用于临床上 不方便测量的患者的眼压的粗略估计,若能熟练掌握有 一定实用价值。
国家公立三级医院西南医科大学附属医院华西医院区域联盟中心医院
• 优点 价廉 耐用 易操作 较精准 测量范围广 [最高可达100mmHg以 上]