Agilent 6400 Series Triple Quad LCMS Onsite Training
Agilent三重四极杆LCMS维护手册
anty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
三重四极杆气相色谱质谱联用仪结构
三重四极杆气相色谱质谱联用仪结构在科学研究和工业生产中,三重四极杆气相色谱质谱联用仪是一种非常常用的分析仪器。
它结合了气相色谱仪和质谱仪的功能,可以实现复杂样品的高效分离和精确鉴定。
以下是这种仪器的关键组成部分及其作用和原理。
1.质谱仪质谱仪是三重四极杆气相色谱质谱联用仪的核心组成部分,主要作用是将样品离子化并将其分离为不同的成分。
质谱仪的主要部件包括离子源、进样系统、四极杆、检测器和计算机控制系统。
2.真空系统真空系统是质谱仪的重要组成部分,其作用是确保仪器内部的高真空环境。
这有助于减少离子在传输过程中的碰撞,提高离子的飞行速度和路径的准确性,从而提高分析的精确度。
真空系统主要由真空泵、真空阀、真空测量仪表和相关的管道组成。
3.进样系统进样系统的作用是将样品引入质谱仪进行分析。
进样系统通常包括自动进样器、注射器、六通阀等部件,可以实现对样品的准确、快速注入,并对样品进行稀释、汽化等处理,以便于后续的分离和分析。
4.离子源离子源是质谱仪的关键部分之一,它的作用是将样品分子离子化,使其带上电荷。
离子化的样品在电场或磁场的作用下,按照不同的质荷比(m/z)进行分离,最终形成质谱图。
常见的离子源有电子轰击源(EI)、化学电离源(CI)、电喷雾电离源(ESI)等。
5.四极杆四极杆是质谱仪中的重要组成部分,主要作用是对不同质荷比的离子进行筛选和过滤。
四极杆由四根平行金属杆组成,施加特定的射频电压,使离子在四极杆间跳跃并改变方向,从而实现不同m/z离子的分离。
通过调整四极杆的电压,可以选择特定m/z范围的离子通过,并阻止其他离子的通过。
6.检测器检测器的作用是检测并测量通过四极杆筛选后的离子的数量。
检测器通常采用电子倍增管或通道电子倍增管等敏感器件,可以检测到离子的微小电流信号,并将其转化为电信号进行放大和处理。
检测器输出的信号可以进一步用于计算样品的浓度和成分含量。
7.计算机控制系统计算机控制系统是三重四极杆气相色谱质谱联用仪的核心控制系统,它可以实现仪器各个部件的精确控制和数据的实时采集、处理、分析以及结果的输出。
Agilent 6470 Triple Quadrupole LC MS Data Sheet说明书
MRM sensitivity Signal-to-Noise ratio (S/N) ESI Negative
1 pg of chloramphenicol injected on column, S/N > 30,000:1 quantifying on the transition m/z 321 to 152 Noise 1 × RMS
Up to 4,000 dynamic MRM transitions per method
Triggered MRM transitions
Up to 10 MRM transitions (primary and secondary) per analyte for library search and compound confirmation
For More Information
These data represent typical results. For more information on our products and services, visit our Web site at /chem.
MRM sensitivity Instrument Detection Limit (IDL) 10 fg of reserpine injected on column,
IDL < 4.0 fg
ESI positive
quantifying on the transition m/z 609 to 195
Parameter
Measure
Specification
MRM sensitivity Signal-to-Noise ratio (S/N) ESI positive
安捷伦的质谱仪产品doc安捷伦的质谱仪产品
安捷伦的质谱仪产品doc安捷伦的质谱仪产品新闻媒体背景材料-媒体资料:安捷伦的质谱仪产品概述质谱测量是一种功能强大的化学分析技术。
它能够被用于识别未知物质,量化已知物质并猎取其化学结构方面的信息。
质谱测量化学品的能力专门精细,可达10亿分之一克。
因此,质谱测量在现有和新兴科学分支中差不多上一种专门有用的技术。
质谱仪能确定样本中的分子重量和数量。
通常,它们还能够将分子分割以测量分子片段的重量。
有关分子及分子片段重量和丰度的信息被收集和显示确实是所谓的质谱。
假如指纹一样,质谱能够用来辨识各种物质。
气相色谱法和液相色谱技术可分离复杂的化学品混合物。
气相色谱仪的或液相色谱仪往往会与质谱仪结合在一起,即气相色谱/质谱仪或液相色谱/质谱仪系统。
在使用质谱前采纳色谱仪会让质谱仪能更加简便地确定复杂样本中所有的单一化学成分。
–它包括了单四极杆质谱,离子阱质谱,飞行时刻质谱(TOF),三重串联四极杆质谱(QQQ)和四极杆-飞行时刻串联质谱(Q-TOF)。
长期以来,安捷伦科技一直差不多上气相色谱仪、液相色谱仪和质谱仪的领导型供应商。
其产品以优良的性能、可靠性及易于使用等优点,赢得了良好信誉。
安捷伦要紧有五项要紧配置的质谱仪产品。
–它包括了单四极杆质谱,离子阱质谱,飞行时刻质谱(TOF),三重串联四极杆质谱(QQQ)和四极杆-飞行时刻串联质谱(Q-TOF)。
依照客户对比研究的结果,当前液相色谱/质谱仪的一个要紧缺点是它们需要分步而繁琐的操作。
安捷伦推出了完整的的液-质联用平台,即Agilent 6000系列液-质联用系统。
它包括了单四极杆质谱,离子阱质谱,飞行时刻质谱(TOF),三重串联四极杆质谱(QQQ)和四极杆-飞行时刻串联质谱(Q-TOF)。
- 他们不仅关注匹配或超过竞争对手的性能,还致力于制造一个新的可靠性标准。
应用液相色谱用途广泛,差不多成为大多数化学行业和生命科学行业中的标准配置。
安捷伦液相色谱和液-质联用系统正为各种客户群服务,如药物研制和生产、蛋白质、食品安全、环境、国土安全和石油化工等市场。
Agilent 6400 Series Triple Quad LCMS Onsite Training
Q1 Q2 Q3 检测器
5) 两根四级杆协同工作(MRM)
进入质谱的 离子
不能通过 Q1 的离子
通过 Q1 的离 子(母离子)
母离子经过 Q2 产生 碎片(子离子)
不能通过 Q2 的子离子
通过 Q2 的子离子(被 选择检测的离子)
3. 基本操作 ..................................................................................................................................9 1) 开机 ......................................................................................................................................9 2) 关机 ......................................................................................................................................9 3) Masshunter 采集软件介绍................................................................................................12 4) 调谐 ....................................................................................................................................15 5) LCMS 方法设定(1200 部分) .........................................................................................17 6) LCMS 方法设定(1290 部分) .........................................................................................18 7) LCMS 方法设定(QQQ 部分) .........................................................................................22 8) 运行单个样品 ....................................................................................................................24 9) 运行序列样品 ....................................................................................................................25 10) 定性分析........................................................................................................................29 11) 定量分析........................................................................................................................34
安捷伦超高效液相色谱-三重四级杆质谱仪器讲解
液质联用仪本实验室使用的液质联用仪是安捷伦公司6400系列的一款产品,包括超高效液相色谱1290和质谱主机G6460,以及与其配套的计算机和打印机,他们之间通过网络协议通讯,并通过网络交换机连接在一起。
本仪器于2014年年初安装使用,价值两百多万元。
物质只要能溶于液体,均可以被检测。
本实验室主要用于农产品样品的农药残留定性检测,超高效液相色谱1290是整个系统的分离和进样装置,样品在色谱柱中经初步分离,通过接口进入质谱。
质谱以离子源、质量分析器和检测器为核心。
离子源是将分析物中的中性化合物离子化,并将产生的离子在电场的作用下进入离子传输毛细管。
离子传输毛细管是离子的导入通道,它将离子源产生的离子传输进入质谱,同时,隔绝了外部的常压与质谱内部的高真空。
离子通过毛细管后,进入离子光学组件,它包括skimmer1,八极杆以及lens1和lens2,进一步除去了溶剂以及中性分子,也是一个高效的离子传输组件,并聚焦随机运动的离子进入三重四极杆质量分析器。
G6460的质量分析器是三重四级杆,是由三组四极杆空间串联而成,一个质谱就是一个四级杆,所以三重四级杆质谱是空间串联的多级质谱分析,也叫做QQQ质谱。
第一个四级杆根据设定的质合比范围扫描和选择所需的离子。
第二个四级杆,也称碰撞池,用于聚集和传送离子。
在选择离子飞行的途中,引入碰撞气体氮气,第三个四级杆用于分析在碰撞池中产生的碎片离子。
实际上,碰撞池采用了六极杆的设计,拥有更好的聚焦及传输功能,四级杆就被淘汰了,但还沿用三重四级杆的名称。
G6460质谱的检测器包括高能打拿极和电子倍增器,此外,质谱需要在真空环境下工作,它的真空系统由前级真空泵和分子涡轮泵组成,前级真空一般在 1.8-2.5Torr之间,高真空在1.9-2.3*10-5Torr之间。
原理:它以液相色谱作为分离系统,质谱为检测系统。
样品在质谱部分和流动相分离,被离子化后,经质谱的质量分析器将离子碎片按质量数分开,经检测器得到质谱图。
Agilent 1290 6400 LCMS Training(液质培训)
热熔挤出技术制备艾拉莫德固体分散体及体外溶出度的影响因素研究
T614 ) 是治疗类风湿的 艾拉莫德 ( iguratimod, 2 生成, 小分子药物, 可选择性抑制 COX抑制免疫
004 ) 基金项目: 国家重大新药创制科技重大专项( 2009ZX09310作者简介: 陈玉红, 女, 硕士 开发 研究方向: 药物新剂型与新技术 Email: zhangcan@ cpu. edu. cn
[3 ]
电离源( ESI 源 ) ; Agilent 1290 液相色谱系统, 包括 G4220B 型 二 元 输 液 泵, G4226A 型 自 动 进 样 器, G1330B 型柱温箱。 1. 2 药品与试剂 T614 原料药、 艾拉莫德片 ( 江苏先声药业有限 , mA17100701 、 42100803 ) ; PVP / VA 64 公司 批 号 ( Kollidon VA64 ) 、 Kollidon CLSF、 Soluplus ( 德国巴斯夫公司 ) ; HPMC ASLF、 HPMC E6 、 HPCSL ( 信越化学工业株式会社 ) ; Eudragit L10055 ( 赢创 德 固 赛 ) ; 部 分 预 胶 化 淀 粉 PC10 、 MCC302 、 MCC802 ( 旭化成) ; CMSNa、 微粉硅胶 ( 湖州展望 ) ; 硬脂酸镁( 安徽山河 ) ; DMF、 乙腈为色谱纯, 其他均 为分析纯。 1. 3 动物 SpragueDawley 大鼠, ( 260 ± 10) g, 雄性( 上海斯 克莱实验动物有限公司, 合格证号: 2007000537261) 。 2 2. 1 方 法 载体预筛选
。疏水性药物以分子、 无定形或者微
晶状态分散在水溶性载体中, 可以减小药物粒径或 [45 ] 。 改变药物物理状态, 增加润湿性和溶出度 melt extrusion, HME ) 技术在高分 热熔挤出( hot子材料的玻璃化转变温度之上对其加热 、 捏合, 使活 并根据给药目 性成分均匀包被在热塑性聚合物中, 。 HME 的将挤出物塑造成形 具有其他制备 SD 的方 法如溶剂挥发法、 喷雾干燥法、 熔融法等所不具备的 特别是可连续操作, 过程一步完成, 大生产可 优势, 行, 且不使用有机溶剂, 避免了环境污染、 爆炸以及 [6 ] 溶剂残留等隐患 。 应用 HME 技术制备 SD 的报 道日益增多, 成功提高溶解度和生物利用度的有伊 [7 ] [8 ] [9 ] 17 β 曲康唑 、 硝苯地平 、 雌二醇 等。 614 的体 外 溶 出 度 并 改 善 其 流 动 为了提高 THPMC 性, 本 实 验 拟 以 经 典 SD 载 体 PVP / VA 64 、 ASLF、 HPCSL 和新型辅料 Soluplus ( PEG6000 / 乙烯基己内酰胺 / 醋酸乙烯酯共聚物 ) 为载体, 采用 HME 技术, 614 速释固体分散体。 同时用 X 制备 T614 在固体分散体中的物理状 射线粉末衍射分析 T考察影响溶出度和 SD 稳定性的因素以及大鼠 态, 体内的生物利用度。 1 1. 1 材料与仪器
国产注射用头孢西丁钠质量评价
国产注射用头孢西丁钠质量评价肖慧;洪建文;彭洁;武建卓;陈冬妮;陆江赢;罗嘉琳【摘要】Objective To evaluate the suitability of statutory standard and the quality of cefoxitin sodium for injection of different domestic manufacturers.Methods According to the general requirements of national assessment programs,the standard methods combined with exploratory researches were used to evaluate the quality of cefoxitin sodium for injection.Results According to the statutory standards for the examination of the 227 batches of samples,the unqualified rate was 6.6%.The main unqualified items for the solution was the clarity,the color,and bacterial endotoxin.The results showed that the content of the individual impurities was correlated with the color of the solution,and the amount of the polymer was correlated with the color of the solution.It is necessary for the current standard to include the related substances II (polymer) inspection items,to revise the related substances limit,the color of the solution and content limit,and the water determination method.Conclusion At present the overall quality of cefoxitin sodium for injection is not satisfactory.Starting from the integrity of the test project,the rationality and validity of the test method,current standards should be further improved to ensure quality control.%目的评价国内不同厂家生产的注射用头孢西丁钠的质量现状及存在问题.方法按照国家评价性抽验计划总体要求,采用法定检验方法结合探索性研究进行样品检验,统计分析检验结果.结果按法定标准检验抽取的227批次样品,不合格率为6.6%,主要不合格项目为溶液的澄清度与颜色,不合格项目涉及安全性项目细菌内毒素.探索性研究表明个别杂质的含量、聚合物的量与溶液的颜色有一定的相关性;建议现行标准增加有关物质II(聚合物)检查项,修订有关物质限度,修订溶液颜色和含量限度,修订水分测定方法.结论目前国内的注射用头孢西丁钠的总体质量一般,存在不合格产品集中企业;从检验项目的完整性、限度制定的合理性和检测方法的有效性出发,现行标准应进一步提高,以保证该药物的有效和安全.【期刊名称】《中国抗生素杂志》【年(卷),期】2017(042)006【总页数】6页(P470-475)【关键词】注射用头孢西丁钠;杂质谱;有关物质;评价性抽验;质量分析【作者】肖慧;洪建文;彭洁;武建卓;陈冬妮;陆江赢;罗嘉琳【作者单位】广东省药品检验所,广州510180;广东省药品检验所,广州510180;广东省药品检验所,广州510180;广东省药品检验所,广州510180;广东省药品检验所,广州510180;广东省药品检验所,广州510180;广东省药品检验所,广州510180【正文语种】中文【中图分类】R918.1头孢西丁为头霉素类抗生素,属第二代头孢菌素,1974年由美国默沙东公司开发上市。
三重四级杆液相质谱联用仪参数
三重串联四极杆液质联用仪品牌:WA TERS1、设备名称:液相色谱串联四极杆质谱联用仪2、主要用途:用于多种农、兽药残留、生物毒素残留、食品添加剂的检测,动物及动植物源食品的残留监控工作,特别是对检测结果为阳性的样品,可进行结果确证,并且可以对预期要开展的检测项目进行检测技术储备,以加强对出口食品、农产品安全卫生质量的监管和控制。
3、工件条件:工作电压:230V±10%温度:18-250C湿度:350-85%4、技术指标:1)超高效液相色谱系统技术要求1.1二元或四元泵*1.1.1压力:0-15,000 psi1.1.2流量:0.001-2.000ml/min, 以0.001ml/min 为增量1.1.3精度:≤0.075%RSD1.1.4流速准确度:±1.0%1.1.5延迟体积:<120μL1.1.6溶剂数量:4种1.1.7设定范围:室温上5 – 60 °C,1°C 步距1.1.8温度控制精度:±0.1oC温度1.2自动进样器1.2.1精度:< 0.5%1.2.2样品交叉污染度:<0.005%1.2.3进样准确度:±1 μL1.2.4进样体积:0.1 ~50μl1.2.5进样线性度: > 0.9991.3在线真空脱气机1.4二极管阵列检测器1.4.1波长范围:190-800nm1.4.2二极管个数:1024个1.4.3光学分辨率:1.2nm1.4.4狭缝宽度:1,2,4,8,16nm1.4.5实时信号:同时输出8个实时信号1.4.6波长精度:1nm1.4.7基线噪音:±1.0 x 10-5 AU,在254nm和750nm(1mL/min甲醇)1.4.8基线漂移: 2 x 10-3AU/hr , 在254nm(1mL/min甲醇)1.4.9数据采集率:〉80Hz2)质谱部分指标*2.1质量范围m/z: MS1及MS2:30-3,000amu2.2分辨率:?2.5M2.3质量数稳定性:≤0.1 amu /8H2.4电喷雾灵敏度指标(相同绝对量,不同浓度下的灵敏度):1pg利血平,m/z 609-195,信噪比≥500:12.5ESI负离子:2.5pg氯霉素,m/z321-152,信噪比≥100:12.6正、负离子采集切换速率≤20 ms2.7MRM扫描时间≤5 ms,一个采集通道可MRM定量数≥502.8扫描速率:?5000amu/秒2.9动态范围:?4X1062.10质量精度:≤0.1 amu2.11离子源:大气压离子源是双正交设计,而且离子源和质谱间有隔断阀,锥孔有N2保护气,可容忍不挥发性的缓冲盐。
Agilent 6400系列 Triple Quad LCMS实验室准备通知书2
Agilent 6420/30/60/90 Triple Quad LC/MS实验室准备通知书尊敬的客户,由衷感谢您选用安捷伦 6420/6430/6460/6490 Triple Quad LC/MS,本文档将协助您进行实验室准备以确保安装顺利开展。
在您的实验室场地准备完毕后请填写最后一页的实验室条件确认表并回传至:8008201182杜晴(长江以南地区) E-mail: qing_du@余慧(长江以北地区) E-mail: hui_yu@任何关于场地准备方面的问题,请拨打8008203278 咨询。
请注意:在安捷伦没有收到完整的场地确认表之前,不会正式考虑您的安装要求。
以下章节供您在准备实验室时参考,请仔细阅读参数条件,这些参数对您的仪器能否运转正常有着至关重要的影响。
请尤其留意红色字体的标注。
一、空间尺寸、重量、电源等要求实验台高×宽×长:约80 ×80 ×300 cm 高度不超过85cm,离墙距离0.5米以上(放置油泵)见图一(a)(b),以便于日后维护,并保证仪器操作人员可以随意进入至仪器后方位置。
承重大于300公斤。
仪器附近墙上有一个直径4cm左右的孔(见图二),便于排放废气,如果此孔距离仪器大于3m,请用户自备废气管。
电源要求:1、单相交流220V,(+5% ~ -10%),50-60Hz,有良好接地。
(零地线电压不大于0.5V,非常重要!!!)见图三。
务必请专业电工测量后,将电压值填入实验室条件确认表。
整个LC-QQQ的额定功率不超过4.0 千瓦(kW)(6490的仪器为5.0 千瓦(kW))2、若现场电压不稳或经常停电,需配置功率大于6 k VA的不间断电源(UPS),UPS供电时间依您的需要而定,一般不短于45分钟,我们推荐山特、或者APC品牌的工频UPS。
(注意在购买UPS时,须说明最好购买工频机,工频机适合精密仪器及检测设备使用)。
四季青中原儿茶酸在beagle犬体内的药动学研究
四季青中原儿茶酸在beagle犬体内的药动学研究程强;羿国娟;周杨杨;贺英;孙文霞;张敬尧【摘要】目的建立LC-MS/MS法测定血浆中原儿茶酸的浓度,并对beagle犬灌胃四季青水煎液后原儿茶酸的药动学进行研究.方法 Beagle犬血浆样品以酮洛芬为内标,用乙腈沉淀蛋白,采用LC-MS/MS法进行分离和测定.色谱柱为C18(4.6mm×100mm,2.7μm);流动相为80%乙腈-20%水(含0.1%甲酸),流速为0.2mL/min,柱温为30℃.质谱条件:电喷雾离子源(ESI),负离子模式,多反应离子监测(MRM),检测离子为原儿茶酸m/z 153.0→109.0,内标酮洛芬m/z 253.1→209.1.6只beagle犬单次灌胃四季青水煎液50mL(0.2g/mL)后,于给药前及给药后0.083、0.167、0.333、0.5、0.667、0.833、1、1.5、2、3、4和6h采血,以LC-MS/MS法测定原儿茶酸在beagle犬体内的血浆浓度,并用非房室模型计算药动学参数.结果原儿茶酸浓度在0.05~2.00μg/mL范围内线性关系良好,定量下限为0.05μg/mL,批内和批间精密度RSD均小于8.6%,准确度为92.3%~102.0%.Beagle犬分别单次灌胃1g/kg(生药量)四季青水煎液后,其主要药动学参数Cmax为(1.53±0.33)μg/mL,Tmax为(0.42±0.09)h,t1/2为(0.80±0.16)h,MRT0~t为(1.04±0.14)h,MRT0~∞为(1.32±0.26)h,AUC0~t为(1.90±0.35)mg/mL)·h,AUC0-∞为(2.02±0.33)(μg/mL)·h.结论本实验中建立的LC-MS/MS法专属性强、快速灵敏,可以用于犬血浆中原儿茶酸的测定.Beagle犬单次灌胃四季青水煎液后,血浆中原儿茶酸能快速吸收,达峰速度较快,在体内能滞留一定时间,有利于四季青药效作用的发挥.【期刊名称】《中国抗生素杂志》【年(卷),期】2018(043)011【总页数】6页(P1407-1412)【关键词】四季青;水煎液;Beagle犬;原儿茶酸;药动学;LC-MS/MS【作者】程强;羿国娟;周杨杨;贺英;孙文霞;张敬尧【作者单位】抗生素研究与再评价四川省重点实验室,四川抗菌素工业研究所,成都大学,成都610052;抗生素研究与再评价四川省重点实验室,四川抗菌素工业研究所,成都大学,成都610052;抗生素研究与再评价四川省重点实验室,四川抗菌素工业研究所,成都大学,成都610052;抗生素研究与再评价四川省重点实验室,四川抗菌素工业研究所,成都大学,成都610052;抗生素研究与再评价四川省重点实验室,四川抗菌素工业研究所,成都大学,成都610052;四川省食品药品检验检测院,成都610097【正文语种】中文【中图分类】R9四季青是冬青科植物冬青(Ilex Chinensis sims)的干燥叶,中国药典2015版一部收载品种,功能与主治为清热解毒、消肿祛瘀,用于治疗肺热咳嗽,咽喉肿痛,痢疾,肋痛,热淋,外用治疗烧烫伤,皮肤溃疡[1-2]。
安捷伦-液质联用技术(LCMS)及其应用
NIST MS library search mode: Automatic library search
NIST MS library file No. : 1 NIST MS library file name : douglib
RetTime Sig MW Library Match Formula
by Li Ping
4
LC/MS联用技术的相对适用范围
by Li Ping
5
大气压电喷雾质谱应用范围
100,000
大气压电喷雾源
10,000
大气压化学源
1000
气相色谱/质谱
分子量
非极性
化合物极性
by Li Ping
极性化合物
6
电喷雾(ESI)的电离过程
带电液滴
+ +
+ -
+ -+
-+
--
+
+ +
502
684.65
m/z--> Abundance
8000 6000 4000 2000
350
400
450
500
碰撞电压= 175 volts
467
439
411
495
550
600
650
700
684.65
loss of C12 fatty acid loss of C14 fatty acid
loss of C16 fatty acid
+
溶剂蒸发
分析物离子
溶剂离子束
盐/离子对 中性化和物
+
Rayleigh Limit
安捷伦LCMS操作
118
922
6460
302 1034
200
400
600 800 1000 1200 1400 m/z
200 400 600 800 1000 1200 1400 m/z
6490 with iFunnel vs 6460 in positive ion mode
1 pg Reserpine (7X) x103 10
6460 6430 6410
2006
2007
2008
2009
2010
New 6490 Triple installation Spec: S/N > 10,000:1
so the 6490 should be able to detect 1 fg ?
X103
1 pg reserpine on-column on column S/N > 22,605:1
iFunnel Technology
Agilent Jet Stream Hexabore capillary Dual-stage ion funnel
Collision Cell – hexapole
Tapered structure for increased ion acceptance Reduction of background noise
3
Agilent Jet Stream Ion Generation
Gas Dynamics View
Enhanced efficiency nebulizer Super-heated sheath gas S h d h h Nozzle voltage g Heated drying gas The supe eated s eat gas e super-heated sheath collimates the nebulizer spray and creates a dramatically “brighter source”
光电比色计(说明书上)
Spectroquant® Photometer NOVA60/60A
便携式多参数分光光度计
操作说明书中欣大厦 40 楼 T/F:021-32224788/62479680
Merck Chemicals (Shanghai) Co.,Ltd
3.1 取样 ..........................................................................................................................................................................9 3.2 初始测试 ..................................................................................................................................................................9 3.3 稀释 ..........................................................................................................................................................................9 3.4 过滤 ........................................................................................................................................................................10 3.5 均质 ........................................................................................................................................................................10 3.6 消解 ........................................................................................................................................................................10
三重四级杆质谱主流产品大盘点!
三重四级杆质谱主流产品大盘点!展开全文三重四极杆质谱作为售价数倍于普通GCMS或LCMS的高端产品,已经大量进入了我国市场,在科研、化工、环境、食品、地质、农业、临床等领域均有广泛的应用。
毋庸置疑,三重四极杆是定量的不二之选,在定量方面有绝对优势,但是在高分辨、多级串级质谱等定性方面优势较弱。
很多国外的实验室都采用装备数台QQQ并配备一台离子阱或者QTOF仪器,以弥补定性能力的不足。
BCEIA 2017)召开期间,仪器信息网邀约三重四级杆质谱主流厂商介绍自家三重四级杆质谱仪产品的技术特点和应用案例,并请各厂商预测了未来一段时间内三重四级杆质谱仪的市场热点及潜力。
目前,三重四级杆质谱仪的市场主要由赛默飞、安捷伦、沃特世、岛津、Sciex等大的跨国公司所占据。
产品形态主要有三个类型:LC-MS/MS、GC-MS/MS、ICP-TQ MS。
各家三重四极杆质谱产品的技术特点(按回稿时间先后排序)赛默飞赛默飞2017年发布两款新型三重四极杆质谱仪TSQ Altis and TSQ Quantis,特别是对客户最为关心的仪器性能指标RRS(Robustness, Reliability, Sensitivity)有很大突破(更高的灵敏度,可靠性和耐用性),并集成了多种新型硬件结构革新,提升仪器性能。
在食品安全、制药、临床研究、法医毒理及环境保护等领域为客户提供极致体验。
由于采用了新型双曲面分段四极杆,显著增强离子传输效率和一致性,同时为客户提供H-SRM分析方法,极大提高对大分子及小分子各化合物目标定量的选择性,提供最佳性能和最高灵敏度。
另外赛默飞新型三重四极杆质谱仪通过精密设计确保主动离子管控+(AIM+)的卓越性能,AIM+集成了双曲面分段四极杆和增强型RF和DC电子元件,进一步优化离子管控精度、可靠性、速度和重现性。
为下一代目标定量提供优质的平台质量保证。
面对未来定量的挑战,赛默飞在提供高品质质谱产品的同时,为客户提供从样品前处理到数据分析管理的一站式服务,更有优质的客户服务及技术支持为客户检测工作提供强大的支撑。
安捷伦科技 Agilent 6100 系列四极杆 LC MS 系统快速入门指南说明书
Agilent 6100 系列四极杆 LC/MS 系统快速入门指南从哪里获得更多信息1配置图4化学工作站视图6基本操作20操作提示27故障排除提示29本手册适用于 OpenLAB CDS 化学工作站版本软件(C.01.03 版及更高版本,直到被替代)和 Agilent 6100 系列 LC/MS 仪器。
从哪里获得更多信息联机帮助使用联机帮助可深入了解超越本快速入门指南所述范围的信息。
联机帮助可通过下列方式显示:• 单击工具栏上的帮助按钮。
• 从帮助菜单选择帮助主题。
• 在大多数对话框中单击帮助按钮可以显示特定任务的帮助。
用户指南除了本快速入门指南以外,系统还附带下列指南,这些指南也是 PDF格式,可在安装盘上找到。
• Agilent 6100 系列四极杆 LC/MS 系统入门指南• Agilent 6100 系列四极杆 LC/MS 系统概念指南• Agilent 6100 系列四极杆 LC/MS 系统安装指南• Agilent 6100 系列四极杆 LC/MS 系统维护指南C.01.03 的新增功能• 支持 32 位和 64 位 Windows 7(Service Pack 1)。
• 在 G6120B 上支持快速扫描。
• 在 G6120B 上支持快速极性切换(需要硬件升级)。
• 在 G6130B 上支持使用安捷伦喷射流技术的 ESI(需要硬件升级)。
C.01.02 的新增功能• 支持 32 位和 64 位 Windows 7。
• 在 OpenLAB CDS 化学工作站版本上支持 6100 系列 LC/MS。
B.04.03 的新增功能• 支持 32 位 Windows 7。
B.04.02 SP1 的新增功能• 支持新的 G6120B、G6130B 和 G6150B 单四极杆 LC/MS 系统。
• 在 6150 型号上支持安捷伦喷射流技术。
• 支持 32 位 Microsoft Windows XP Professional(Service Pack 3)。
Agilent MassHunter LC MS 用户手册与工具说明书
/cs/library/usermanuals/public/G333590228_QQQ_QuickStart.pdf
Triple Quad LC/MS Upgrade Guide
Triple Quad LC/MS System Installation Guide
/cs/library/usermanuals/public/Data Acquisition for TQ LCMS eFamiliarization.zip
Gate Valve Quick Reference Poster
http://wபைடு நூலகம்/cs/library/quickreference/public/G257 1-90028_GateValve_QuickRef.pdf
IM-QTOF System Installation Guide
/cs/library/installationguide/G3335-90194_IMQTOF_Installation_B.pdf
TOF/Q-TOF/IM-QTOF Upgrade Guide
http://www//cs/library/usermanuals/public/6000 Series LCMS Hardware eFamiliarization.rar
/cs/library/installationguide/G333590223_LCMS_Offline_Install.pdf
G6470A to G6495B Upgrade Kit Installation Guide
/cs/library/installationguide/G257190022_TQ_UpgradeKit_Installation.pdf
- 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
- 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
- 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
1.QQQ 基本原理
1) 离子的产生(ESI)
离子源将进入其中的流动相变成雾状,并且使用高温气体反吹使其迅速蒸发以释放出其 中的离子;离子在电场的作用下,最终进入毛细管入口。 注意:离子本身就存在于溶液中(APCI 源不同),电喷雾源的作用仅仅是提取出溶液中已 经存在的离子。我们可以使用各种促进离子化的流动相,比如正离子模式下加一定量的甲 酸,负离子模式下避免使用酸性流动相等手段增加溶液中的离子。 注意:其中雾化气压力(Nebulizer);干燥气流量(Gas flow);干燥气温度(Gas temp); 和毛细管电压(Capillary)的设置对于离子是否能最大化的进入质谱仪起到关键的作用。 所以在必要的时候,我们需要优化这些参数。
3. 基本操作 ..................................................................................................................................9 1) 开机 ......................................................................................................................................9 2) 关机 ......................................................................................................................................9 3) Masshunter 采集软件介绍................................................................................................12 4) 调谐 ....................................................................................................................................15 5) LCMS 方法设定(1200 部分) .........................................................................................17 6) LCMS 方法设定(1290 部分) .........................................................................................18 7) LCMS 方法设定(QQQ 部分) .........................................................................................22 8) 运行单个样品 ....................................................................................................................24 9) 运行序列样品 ....................................................................................................................25 10) 定性分析........................................................................................................................29 11) 定量分析........................................................................................................................34
安捷伦 6400 系列 液相色谱-三重串联四级杆质谱联用仪
现场培训手册
பைடு நூலகம்
版本
© 安捷伦科技(中国)有限公司, 2010 2010 年 6 月 Rev. 6.0 本手册适用于安捷伦 MassHunter 工作站软件 B.02.01 及以上版本。
声明
本书内容如有改变,恕不另行通知。安捷伦科技公司对本材料,及由此引起的任何商务和特 种用途不承担责任。安捷伦科技公司对本手册中可能有的错误或与装置、性能及材料使用有 关内容而带来的意外伤害和问题不负任何责任。如果安捷伦与用户对本书中的警告术语有不 同的书面协议,这些术语与本书中的警告术语冲突,则以协议中的警告术语为准。
2. 硬件概述 ..................................................................................................................................6 1) LC-QQQ 的典型配置 ............................................................................................................6 2) 1290 中各模块典型摆放及介绍 .........................................................................................6 3) 1200 中各模块典型摆放及介绍 .........................................................................................7 4) QQQ 的离子源的种类介绍 .................................................................................................7 5) 仪器序列号的位置 ..............................................................................................................8
Q1 Q2 Q3 检测器
5) 两根四级杆协同工作(MRM)
进入质谱的 离子
不能通过 Q1 的离子
通过 Q1 的离 子(母离子)
母离子经过 Q2 产生 碎片(子离子)
不能通过 Q2 的子离子
通过 Q2 的子离子(被 选择检测的离子)
APCI 以及 Jet Stream 的原理,请见附录
2) 离子如何进入质谱
Fragmentor 电压
离子源
毛细管
八级杆组件
四级杆
离子进入四级杆前必须要经过毛细管,八级杆组件。其中毛细管的作用是导入离子,在 毛细管末端施加一个电压(Fragmentor 电压),把离子推进八级杆组件中。而八级杆组件的 作用大致为导向随机运动的离子,并且让其以一定速度进入四级杆。
联系方式
售后服务中心电话:800 820 3278 手机用户:400 820 3278 (手机用户需支付本地电话费) 免费传真:800 820 1172 电子信箱:lsca-china_800@ 中文网站:
目录
1. QQQ 基本原理 .........................................................................................................................3 1) 离子的产生(ESI) .............................................................................................................3 2) 离子如何进入质谱 ..............................................................................................................4 3) 四级杆的作用(Q1,Q3) .........................................................................................................4 4) 碰撞池的作用(Q2) ...............................................................................................................4 5) 两根四级杆协同工作(MRM) .........................................................................................5 6) 检测器 ..................................................................................................................................5
4. 日常维护 ................................................................................................................................45 1) 液相的日常维护(1200,及 1290) ...............................................................................45 2) QQQ 的日常维护 ...............................................................................................................47