AOAC990.28 Sulfites in Foods 食品中亚硫酸盐的检测方法

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。

47.3.43

AOAC Official Method990.28

Sulfites in Foods

Optimized Monier–Williams Method

First Action1990

Final Action1994

(Applicable of determination of≥10ppm(µg/g)sulfites in foods. Applicable in presence of other volatile sulfur compounds;not ap-plicable to dried onions,leeks,and cabbage.)

Results of the interlaboratory study supporting the acceptance of the method:

Hominy,9.17ppm(µg/g)sulfites:

s r=1.33;s R=1.42;RSD r=14.5%;RSD R=15.5%

Fruit juice,8.05ppm(µg/g)sulfites:

s r=1.36;s R=1.62;RSD r=16.9%;RSD R=20.1%

Protein(seafood),10.41ppm(µg/g)sulfites:

s r=1.47;s R=2.77;RSD r=14.1%;RSD R=26.6%

A.Principle

Method measures free sulfite plus reproducible portion of bound sulfites,such as carbonyl addition products,in foods.Test portion is heated with refluxing HCl(ca1M)to convert sulfite to SO2.Stream of N2introduced below surface of refluxing solution sweeps SO2 through water-cooled condenser and,via bubbler attached to con-denser,with3%H2O2solution,where SO2is oxidized to H2SO4. Sulfite content is directly related to generated H2SO4,which is deter-mined by titration with standardized NaOH solution.For verifica-tion,sulfate can be determined gravimetrically as BaSO4.

B.Apparatus

(a)Distillation apparatus.—(Note:In this method,back pres-sure inside apparatus is limited to unavoidable pressure due to height of3%H2O2solution above tip of bubbler(F).Keep back pressure as low as possible to avoid loss of SO2through e thin film of stopcock grease on sealing surfaces of all joints except joint between separatory funnel and flask.Clamp together each joint to ensure complete seal throughout analysis.)Assemble apparatus(Figure 990.28A),which includes(1)inlet adapter(A)with hose connector (Kontes183000).Adapter provides means of applying head pres-sure above e of pressure-equalizing dropping funnel is not recommended because condensate,perhaps containing SO2,is deposited in funnel and side arm.(2)Separatory funnel(B),≥100mL capacity.(3)Round-bottom flask(C),1L,with three24/40 tapered joints.(4)Gas inlet tube(D)(Kontes179000)of sufficient length to permit introduction of N2within2.5cm of bottom of flask.

(5)Allihn condenser(E)(Kontes431000-2430),jacket length 300mm.(6)Bubbler(F),fabricated from glass according to dimen-sions in Figure990.28B.(7)Vessel(G),ca2.5cm id and18cm deep.

(b)Buret.—10mL(Kimble Glass,Inc.,No.17124-F)with over-flow tube and hose connections for Ascarite tube or equivalent air-scrubbing apparatus to permit maintenance of CO2-free atmo-sphere over standardized0.010M NaOH.

(c)Chilled water circulator.—Chill condenser with coolant, such as methanol-water(20+40,v/v),maintained at≤15°C.Circu-lating pump,Neslab Coolflow33(Neslab Instruments,Inc.,PO Box 1178,Portsmouth,NH03801,USA),or equivalent,is suitable.C.Reagents

(a)Aqueous hydrochloric acid.—4M.For each analysis,prepare 90mL solution by adding30mL HCl to60mL deionized(18meg-ohm)water.

(b)Methyl red indicator.—Dissolve250mg methyl red in 100mL ethanol.

(c)Standardized titrant.—0.010M NaOH.Certified reagent may be used(Fisher SO-5-284).Standardize solution with reference standard potassium acid phthalate.

(d)Hydrogen peroxide solution.—3%.For each analysis,dilute 3mL ACS reagent grade30%H2O2to30mL with deionized (18megohm)water.Just prior to use,add3drops methyl red indica-tor and titrate with0.010M NaOH to yellow end point.If end point is exceeded,discard solution.

(e)Nitrogen.—High purity,used with regulator to maintain flow of200mL/min.To guard against oxygen in N2gas,use GC-type trap (Oxy-Purge N[Alltech-Applied Science Laboratories,Inc.],or equivalent).

Alternatively,oxygen-scrubbing solution,such as alkaline pyrogallol,in gas-washing bottle(Kimble Glass,Inc.)may be used. Prepare trap as follows:(1)Add4.5g pyrogallol to trap.(2)Purge trap with N2for2–3min.(3)Prepare KOH solution by adding65g

Figure990.28A—Apparatus for optimized Monier-Williams method:A,inlet adapter;B,separatory funnel; C,round-bottom flask;D,gas inlet tube;E,Allihn con-denser;F,bubbler;G,vessel.

相关文档
最新文档